Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System

K191027510(k) TraditionalCleared Mar 18, 2020335-day reviewClass II21 CFR 868.5160Recalls 11 (8 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Maquet Critical Care AB · BSZ Gas-Machine, Anesthesia

Cleared on Mar 18, 2020 as substantially equivalent to K160665 (Maquet Critical Care AB, 2017) after a 335-day review. Its family tree reaches back 12 generations to 1994. 4 of the 39 devices in its family, counting itself, have a Class I recall. It has been recalled 11 times, including 8 Class I recalls.

Decided
Mar 18, 2020
Review
335 days
Class
Class II
Product code
BSZ
Recalls
11 (8 Class I)
Predicates
1

Its longest line of descent runs back 12 generations to K930017 (1994). 38 ancestors in all, the oldest K811102 (1981). No later device cites it as a predicate yet. 4 of the 39 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2008: 1 other ancestor2004: 2 other ancestors2002: 1 other ancestor2000: 2 other ancestors1999: 1 other ancestor1997: 1 other ancestor1991: 1 other ancestor1989: 2 other ancestors1981: 1 other ancestor1996: 3 other ancestors1995: 2 other ancestors1993: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestor1985: 2 other ancestors1984: 2 other ancestors1998: 2 other ancestorsReference device: K170872 · Aisys CS2 · Datex-Ohmeda, Inc. · 2017 · recalledK930017 · PB 7200 Series Ventilatory System W/Keyboard · Puritan Bennett Corp. · 1994K023366 · Datex-Ohmeda 7900 Ventilator Enhancements · Datex-Ohmeda, Inc. · 2003 · Class I recall×2: 2 generations cleared close together, K040743 (2004), K042154 (2004) K061609 · GE Datex-Ohmeda Aisys · Datex-Ohmeda, Inc. · 2006 · recalledK071142 · GE Datex-Ohmeda Avance Anesthesia System · Datex-Ohmeda · 2007K081844 · GE Datex-Ohmeda Avance · Datex-Ohmeda, Inc. · 2008 · recalledK090233 · GE Datex-Ohmeda Aisys · Datex-Ohmeda, Inc. · 2009 · recalledK102182 · Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30 · Maquet Critical Care AB · 2011K112114 · Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40 · Maquet Critical Care AB · 2012 · Class I recallK133958 · Flow-I Anesthesia System C20, C30 & C40 · Maquet Critical Care AB · 2014 · Class I recallK160665 · FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i Anesthesia System C40 · Maquet Critical Care AB · 201719801990200020102020
  1. Root · 12 generations backK930017 PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp. · 1994
  2. 8 generations between
  3. 3 generations backK112114 Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012 · Recalled (Class I)
  4. 2 generations backK133958 Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · Recalled (Class I)
  5. 1 generation backK160665 FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i…Maquet Critical Care AB · 2017
  6. This deviceK191027 Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e…Maquet Critical Care AB · 2020
Show all 12 generations
  1. Root · 12 generations backK930017 PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp. · 1994
  2. 11 generations backK023366 Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · Recalled (Class I)
  3. 10 generations backK040743 Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc. · 2004
  4. 9 generations backK042154 GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc. · 2004 · Recalled
  5. 8 generations backK061609 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2006 · Recalled
  6. 7 generations backK071142 GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda · 2007
  7. 6 generations backK081844 GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · Recalled
  8. 5 generations backK090233 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · Recalled
  9. 4 generations backK102182 Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30Maquet Critical Care AB · 2011
  10. 3 generations backK112114 Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012 · Recalled (Class I)
  11. 2 generations backK133958 Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · Recalled (Class I)
  12. 1 generation backK160665 FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i…Maquet Critical Care AB · 2017
  13. This deviceK191027 Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e…Maquet Critical Care AB · 2020

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K930017PB 7200 Series Ventilatory System W/KeyboardPuritan Bennett Corp.1994—
−11K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc.2003Class I
−10K040743Modification to Datex-Ohmeda S/5 Avance Anesthesia SystemDatex-Ohmeda, Inc.2004—
−9K042154GE Datex-Ohmeda Aisys CarestationDatex-Ohmeda, Inc.20043
−8K061609GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20061
−7K071142GE Datex-Ohmeda Avance Anesthesia SystemDatex-Ohmeda2007—
−6K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc.20084
−5K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20092
−4K102182Maquet Flow-I C20, Maquet Flow-I C30, Model 66 77 200, 66 77 30Maquet Critical Care AB2011—
−3K112114Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB2012Class I
−2K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB2014Class I
−1K160665FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i Anesthesia…Maquet Critical Care AB2017—
0 (this device)K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198420
198520
198610
198710
198920
199110
199310
199520
199630
199710
199820
199910
200020
200210
200420
200810
FIG. 1 — Lineage of K191027, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160665FLOW-i Anesthesia System C20, FLOW-i Anesthesia System C30, FLOW-i Anesthesia System C40
Summary page 5
…Regulation Number 21 CFR 868.5160 Predicate Device Identification Maquet FLOW-i Anesthesia System K160665 Reference Device Identification GE Aisys CS2 K170872…
Maquet Critical Care AB2017—
†K170872Aisys CS2Datex-Ohmeda, Inc.20174

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 24, 2026Class IIZ-2860-2026Under Investigation by firmEvents observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Jan 15, 2025Class IZ-1074-2025Device DesignSevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which…
May 1, 2024Class IZ-1877-2024Under Investigation by firmSevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which…
Dec 22, 2022Class IZ-0959-2023Software Design ChangeDue to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
Dec 22, 2022Class IZ-0961-2023Software Design ChangeDue to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
Dec 22, 2022Class IZ-0960-2023Software Design ChangeDue to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
Nov 7, 2022Class IIZ-0422-2023Component design/selectionThe power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could…
May 2, 2022Class IZ-1279-2022Component design/selectionPotential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
May 2, 2022Class IZ-1278-2022Component design/selectionPotential for crack to form on the on/off switch of the suction unit, which may result in an unreasonable risk of harm to the patient. In the worst case, the switch can break, preventing the suction unit from being activated.
Dec 9, 2021Class IZ-0468-2022Under Investigation by firmA potential chemical degradation of Sevoflurane by Lewis acids (metal oxides and metal halides) has been discovered that may result in inhalation and/or skin exposure of hydrogen fluoride.
Jan 29, 2021Class IIZ-1181-2021Device DesignMaquet Critical Care AB received complaints where the FiCO2 value was not zero or as close to zero as expected and/or experienced System Check Out failures when docking the absorber to the Flow i Anesthesia machine.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K112114Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 8 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 1.8 years on median.
  • 14 of the 38 ancestors have a recall of any class; 3 have a Class I recall.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

The indication for Flow-i/Flow-c/Flow-e Anesthesia System is administering inhalation Anesthesia while controlling the entire ventilation of patients with no ability to breathe, as well as in supporting patients with a limited ability to breathe. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment, by healthcare professionals trained in inhalation Anesthesia administration.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191027: Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K191027

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.