510(k) Report

510(k) · Cleared

Recalled · 2026

Orthoscan VERSA Mini C-Arm

K243452510(k) TraditionalCleared Jan 14, 202568-day reviewClass II21 CFR 892.1650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ziehm-Orthoscan, Inc. · OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Cleared on Jan 14, 2025 as substantially equivalent to K213113 (Orthoscan, Inc., 2021) and 1 other predicate after a 68-day review. Its family tree reaches back 8 generations to 2000. 1 of the 17 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled once.

Decided
Jan 14, 2025
Review
68 days
Class
Class II
Product code
OXO
Recalls
1
Predicates
2

Its longest line of descent runs back 8 generations to K003266 (2000). 14 ancestors in all. 2 later devices descend from it, 1 citing it directly. 1 of the 17 devices in this family, counting this one, has a Class I recall.

  1. Root · 8 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 7 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 6 generations backK061310 Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006
  4. 5 generations backK073346 Ziehm Vision 2Ziehm Imaging, Inc. · 2008 · Recalled
  5. 4 generations backK092438 Ziehm SoloZiehm Imaging, Inc. · 2009 · Recalled
  6. 3 generations backK161976 Ziehm Solo FDZiehm Imaging GmbH · 2016 · Recalled
  7. 2 generations backK183220 OrthoScan TAU Mini C-ArmOrthoscan, Inc. · 2019
  8. 1 generation backK213113 Orthoscan Tau Mini C-ArmOrthoscan, Inc. · 2021
  9. This deviceK243452 Orthoscan VERSA Mini C-ArmZiehm-Orthoscan, Inc. · 2025
  10. Descendant · cites it directly · 1 of its ownK250587 Orthoscan TAU Mini C-ArmZiehm-Orthoscan, Inc. · 2025
  11. Descendant · 2 generations on · 0 of its ownK252579 Orthoscan TAU MVP Mini C-Arm SystemZiehm-Orthoscan, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K250587Orthoscan TAU Mini C-ArmZiehm-Orthoscan, Inc. · 2025 · cites it directly1descendant
  • K252579Orthoscan TAU MVP Mini C-Arm SystemZiehm-Orthoscan, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−7K032280Siremobil Iso-C 3DSiemens AG2003—
−6K061310Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc.2006—
−5K073346Ziehm Vision 2Ziehm Imaging, Inc.20085
−4K092438Ziehm SoloZiehm Imaging, Inc.20094
−3K161976Ziehm Solo FDZiehm Imaging GmbH20161
−2K183220OrthoScan TAU Mini C-ArmOrthoscan, Inc.2019—
−1K213113Orthoscan Tau Mini C-ArmOrthoscan, Inc.2021—
0 (this device)K243452Orthoscan VERSA Mini C-ArmZiehm-Orthoscan, Inc.2025—
+1K250587Orthoscan TAU Mini C-ArmZiehm-Orthoscan, Inc.2025—
+2K252579Orthoscan TAU MVP Mini C-Arm SystemZiehm-Orthoscan, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200131
200610
201210
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
202610
FIG. 1 — Lineage of K243452, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OXO device

  1. #1 K253296 Digital C-ARM with 3Dscore 15.1 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K213113Orthoscan Tau Mini C-Arm
Summary page 5
…ication Names: Image-intensified fluoroscopic x-ray system Product Code: OXO, JAA, MQB Primary Predicate: Orthoscan TAU Mini C-Arm 510(k) Number: K213113 Classification(s): 21CFR 892.1650 Device Class: Class II Classification Names: Image-intensified fluoroscopic x-…
Orthoscan, Inc.2021—
*K113708Orthoscan Mobile DI Mini C-Arm
Summary page 5
…es: Image-intensified fluoroscopic x-ray system Product Codes: OXO, JAA, MQB Secondary Predicate: Orthoscan Mobile DI Mini C-Arm 510(k) Number: (K113708) Classification(s): 21CFR 892.1650 Device Class: Class II Classification Names: Image-intensified fluoroscopic x…
Orthoscan, Inc.20121

Also cleared against K213113: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250587Orthoscan TAU Mini C-ArmZiehm-Orthoscan, Inc.20251

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 3, 2026Not classifiedZ-0286-2027Process controlThe accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 7 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 25 years and passes through 6 companies; a new generation was cleared every 2.8 years on median.
  • 7 of the 14 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K250587 (Ziehm-Orthoscan, Inc., 2025) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K213113.

How does its category compare?

All OXO devices

OXO Image-Intensified Fluoroscopic X-Ray System, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · OXO

4in 2025

2010: 12012: 42013: 22014: 22015: 12016: 32017: 42018: 12019: 32021: 22024: 22025: 42026 (partial year): 2
2007high 4 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OXO

97in 2025

2012: 48 days median2016: 128 days median2017: 57 days median2019: 213 days median2025: 97 days median
2007high 213 (2019) · low 48 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OXO

4in 2024

2012: 42013: 32014: 42016: 32017: 42018: 112024: 42026 (partial year): 1
2007high 11 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OXO

118in 2025

2012: 52013: 412014: 892015: 972016: 3,8202017: 1,6972018: 1,6472019: 1,2972020: 8162021: 5542022: 2952023: 2752024: 1922025: 118
2007high 3,820 (2016) · low 0 (2007)2025

MAUDE reports naming product code OXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
201244845
20132—341
20142—489
20151—097
2016312833,820
201745741,697
20181—111,647
2019321301,297
20200—0816
20212—0554
20220—0295
20230—0275
20242—4192
20254970118
20262—1—

What is it indicated for?

The Orthoscan VERSA Mini C-arm X-ray system is designed to provide physicians with general fluoroscopic visualization, using pulsed or continuous fluoroscopy, of a patient including but not limited to, diagnostic, surgical, and critical emergency care procedures for patients of all ages including pediatric populations when imaging limbs/extremities, shoulders, at locations including but not limited to, hospitals, ambulatory surgery, emergency, traumatology, orthopedic, critical care, or physician office environments.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K243452: Orthoscan VERSA Mini C-Arm.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K243452

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026