510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Ziehm Vision 2

K073346510(k) TraditionalCleared Mar 21, 2008113-day reviewClass II21 CFR 892.1750Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ziehm Imaging, Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Mar 21, 2008 as substantially equivalent to K011292 (Ziehm International, Inc., 2001) and 2 other predicates after a 113-day review. Its family tree reaches back 3 generations to 2000. 1 of the 36 devices in its family, counting itself, has a Class I recall. 28 later devices descend from it. It has been recalled 5 times.

Decided
Mar 21, 2008
Review
113 days
Class
Class II
Product code
JAK
Recalls
5
Predicates
3

Its longest line of descent runs back 3 generations to K003266 (2000). 7 ancestors in all. 28 later devices descend from it, 4 citing it directly. 1 of the 36 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2001: 3 other ancestors, 1 with a Class I recall12016: 1 descendant cleared2015: 1 descendant cleared2009: 2 descendants cleared2013: 1 descendant cleared2020: 3 descendants cleared2019: 2 descendants cleared2023: 3 descendants cleared2021: 3 descendants cleared2025: 6 descendants cleared2024: 2 descendants cleared2026: 3 descendants cleared2022: 1 descendant clearedK003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 2000K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K061310 · Ziehm Vario 3D Digital Mobile Imaging System · Ziehm Imaging, Inc. · 2006200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors3 generations back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 4 descendants2 generations forward: 6 descendants3 generations forward: 7 descendants4 generations forward: 7 descendants5 generations forward: 3 descendants6 generations forward: 1 descendant K061310 · Ziehm Vario 3D Digital Mobile Imaging System · Ziehm Imaging, Inc. · 2006K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 20000+5
  1. Root · 3 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 2 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 1 generation backK061310 Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006
  4. This deviceK073346 Ziehm Vision 2Ziehm Imaging, Inc. · 2008
  5. Descendant · cites it directly · 17 of its ownK083545 Ziehm Vision RFDZiehm Imaging, Inc. · 2009 · Recalled
  6. Descendant · 2 generations on · 16 of its ownK132904 Ziehm Vision RFDZiehm Imaging GmbH · 2013 · Recalled
  7. Descendant · cites it directly · 10 of its ownK142740 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2015 · Recalled
  8. Descendant · cites it directly · 9 of its ownK092438 Ziehm SoloZiehm Imaging, Inc. · 2009 · Recalled
  9. Descendant · cites it directly · 8 of its ownK161976 Ziehm Solo FDZiehm Imaging GmbH · 2016 · Recalled

Showing 5 of 28 descendants. See all descendants

Descendants cleared per year

2009: 22013: 12015: 12016: 12019: 22020: 32021: 32022: 12023: 32024: 22025: 62026: 3
2009high 6 (2025) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 28 · See all
  • K083545Ziehm Vision RFDZiehm Imaging, Inc. · 2009 · cites it directly17descendants
  • K132904Ziehm Vision RFDZiehm Imaging GmbH · 2013 · 2 generations on16descendants
  • K142740Ziehm Vision RFD 3DZiehm Imaging GmbH · 2015 · cites it directly10descendants
  • K092438Ziehm SoloZiehm Imaging, Inc. · 2009 · cites it directly9descendants
  • K161976Ziehm Solo FDZiehm Imaging GmbH · 2016 · cites it directly8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−2K032280Siremobil Iso-C 3DSiemens AG2003—
−1K061310Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc.2006—
0 (this device)K073346Ziehm Vision 2Ziehm Imaging, Inc.2008—
+1K083545Ziehm Vision RFDZiehm Imaging, Inc.20093
+2K132904Ziehm Vision RFDZiehm Imaging GmbH20131
+1K142740Ziehm Vision RFD 3DZiehm Imaging GmbH20151
+1K092438Ziehm SoloZiehm Imaging, Inc.20094
+1K161976Ziehm Solo FDZiehm Imaging GmbH20161

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200131
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200920
201310
201510
201610
201920
202030
202130
202210
202330
202420
202560
202630
FIG. 1 — Lineage of K073346, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011292Ziehm Vision
Summary page 2
…m .2 510(k) Summary Zichm Imaging, Inc. Ziehm Yision * existing Ziehm Vision C-arm devices, which are legally marketed devices (Ziehm Vision application K011292, Ziehm Vision FD application K 061534, and Ziehm Vario 3D application K061310). Substantial Equivalence: The ZIEHM VISION2 prod…
Ziehm International, Inc.2001—
*K061534Ziehm Vision FD Digital Mobile Imaging System
Summary page 2
…ichm Imaging, Inc. Ziehm Yision * existing Ziehm Vision C-arm devices, which are legally marketed devices (Ziehm Vision application K011292, Ziehm Vision FD application K 061534, and Ziehm Vario 3D application K061310). Substantial Equivalence: The ZIEHM VISION2 product family of mobile c-arm products do…
Ziehm Imaging, Inc.2006—
*K061310Ziehm Vario 3D Digital Mobile Imaging System
Summary page 2
…ng Ziehm Vision C-arm devices, which are legally marketed devices (Ziehm Vision application K011292, Ziehm Vision FD application K 061534, and Ziehm Vario 3D application K061310). Substantial Equivalence: The ZIEHM VISION2 product family of mobile c-arm products does not raise new questions of safety or…
Ziehm Imaging, Inc.2006—

Also cleared against K011292: 2 other devices →

What descends from it? (28)

See all 28

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083545Ziehm Vision RFDZiehm Imaging, Inc.200917
K142740Ziehm Vision RFD 3DZiehm Imaging GmbH201510
K092438Ziehm SoloZiehm Imaging, Inc.20099
K161976Ziehm Solo FDZiehm Imaging GmbH20168

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2024Class IIZ-0575-2025Radiation Control for Health and Safety ActZiehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when…
Apr 15, 2016Class IIZ-2644-2016Device DesignMalfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.
Nov 21, 2013Class IIZ-0564-2014Device DesignEngineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
Nov 21, 2013Class IIZ-0565-2014Device DesignEngineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.
Apr 29, 2011Not classifiedZ-1430-2012Component design/selectionOn 04/29/2011, Ziehm Imaging, Inc., Orlando, FL recalled the double foot switch, type KF 2 1S/1S-MED-AP, a component used with Ziehm Imaging GmbH Mobile C-arm due to the potential of unintentional continued activation of fluoroscopic…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 3 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 2.8 years on median.
  • 2 of the 7 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K083545 (Ziehm Imaging, Inc., 2009) has gone furthest, with 17 descendants of its own.
  • 2 other devices also cite its predicate K011292.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The ZIEHM VISION’ product Family is intended to provide pulsed and continuous fluoroscopic imaging of patients during diagnostic, interventional and surgical procedures. It is intended for use in visualizing complex anatomical structures and procedures such as, vascular, cardiac, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurological, critical care, and emergency room procedures. Radiographic film examinations can be made with an…

Read the full statement (393 words)

The ZIEHM VISION’ product Family is intended to provide pulsed and continuous fluoroscopic imaging of patients during diagnostic, interventional and surgical procedures. It is intended for use in visualizing complex anatomical structures and procedures such as, vascular, cardiac, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurological, critical care, and emergency room procedures. Radiographic film examinations can be made with an accessory cassette device when attached to the Image Intensifier. At the discretion of a physician the device may be used for other imaging applications.

ZIEHM VISION with option VARIO is intended to provide pulsed and continuous fluoroscopic imaging of patients during diagnostic, interventional and surgical procedures. It is intended for use in visualizing complex anatomical structures and procedures such as, vascular, cardiac, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurological, critical care, and emergency room procedures. and whenever the clinician benefits from variable Iso-centric positioning. Radiographic film examinations can be made with an accessory cassette device when attached to the Image Intensifier. At the discretion of a physician the device may be used for other imaging applications.

ZIEHM VISION® with option FD is intended to provide pulsed and continuous fluoroscopic imaging of patients during diagnostic, interventional and surgical procedures. It is intended for use in visualizing complex anatomical structures and procedures such as, vascular, cardiac, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurological, critical care, and emergency room procedures and where higher accuracy in image geometry is required in and around high magnetic fields. At the discretion of a physician the device may be used for other imaging applications.

ZIEHM VISION® with option 3D is intended to provide pulsed and continuous fluoroscopic imaging of patients during diagnostic, interventional and surgical procedures. It is intended for use in visualizing complex anatomical structures and procedures such as, vascular, cardiac, angiographic, cholangiography, endoscopic, urologic, orthopedic, neurological, critical care, and emergency room procedures and Whenever the clinician benefits from variable Iso-centric positioning and/or intraoperatively generated 3D imaging of high contrast objects (bones and joints), complex anatomical structures such as cervical, thoracic, and lumber regions of the spine and joint fractures of the upper and lower extremities, and where digital image and c-arm positioning data is required for computer aided surgery procedures. Radiographic film examinations can be made with an accessory cassette device when attached to the Image Intensifier. At the discretion of a physician the device may be used for other imaging applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K073346: Ziehm Vision 2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K073346

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026