510(k) Report

510(k) · Cleared

Recalled · 2006

Drake Willock System 1000 Dialysate Delivery Syst

K910215510(k) TraditionalCleared Apr 9, 199182-day reviewClass II21 CFR 876.5820Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cd Medical, Inc. · FKP System, Dialysate Delivery, Single Patient

Cleared on Apr 9, 1991 after an 82-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 14 later devices descend from it. It has been recalled 5 times.

Decided
Apr 9, 1991
Review
82 days
Class
Class II
Product code
FKP
Recalls
5
Predicates
0

14 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 1996: 1 descendant cleared2002: 2 descendants cleared1997: 2 descendants cleared1999: 1 descendant cleared2007: 2 descendants cleared2004: 1 descendant cleared2009: 1 descendant cleared2005: 1 descendant cleared2008: 1 descendant cleared2022: 2 descendants cleared1990200020102020
  1. This deviceK910215 Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc. · 1991
  2. Descendant · cites it directly · 13 of its ownK954987 Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1996 · Recalled
  3. Descendant · cites it directly · 11 of its ownK970446 Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1997 · Recalled
  4. Descendant · 2 generations on · 9 of its ownK993064 Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products · 1999 · Recalled
  5. Descendant · 3 generations on · 8 of its ownK021615 Baxter Accura System, Model 5M5660Baxter Healthcare Corp · 2002 · Recalled
  6. Descendant · 4 generations on · 4 of its ownK041005 Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products · 2004 · Recalled

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

1996: 11997: 21999: 12002: 22004: 12005: 12007: 22008: 12009: 12022: 2
1996high 2 (1997) · low 0 (1998)2026

Most influential descendants

Showing 5 of 14 · See all
  • K954987Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1996 · cites it directly13descendants
  • K970446Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1997 · cites it directly11descendants
  • K993064Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products · 1999 · 2 generations on9descendants
  • K021615Baxter Accura System, Model 5M5660Baxter Healthcare Corp · 2002 · 3 generations on8descendants
  • K041005Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products · 2004 · 4 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K910215Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc.1991—
+1K954987Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl19961
+1K970446Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl19975
+2K993064Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products19991
+3K021615Baxter Accura System, Model 5M5660Baxter Healthcare Corp20023
+4K041005Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products20047

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199610
199720
199910
200220
200410
200510
200720
200810
200910
202220
FIG. 1 — Lineage of K910215, 1991–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FKP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (14)

See all 14

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K954987Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl199613
K970446Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl199711
K964922Drake Willock System 1000 Dialysate Delivery System Altratouch 1000 Dialysate Delivery…Althin Medical AB an Affiliate of Baxter Intl19971
K013562Aurora System 1000 Series Home Self-Care Single Patient Delivery System; Models- Sys…Baxter Healthcare Corp20020

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 21, 2006Not classifiedZ-1253-06OtherThe UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
Feb 24, 2006Not classifiedZ-0766-06OtherThe older style 'Through Hardware' Dialysate Monitoring Manifolds and 'Noryl' Ultra Filtration Removal Regulators may become cracked due to mechanical and thermal stress. If either of these parts become cracked during patient treatment…
Nov 18, 2005Not classifiedZ-0347-06OtherPotential for fluid or air to be passed through the venous line clamp if the tubing is not centered in the clamp and extends beyond the edge of the clamp's pinch zone. This could result in an air emboli condition.
May 28, 2003Not classifiedZ-0395-05OtherThe air detector may not detect air bubbles consistently at the selected limit.
Mar 20, 2003Not classifiedZ-0833-03OtherThe dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K954987 (Althin Medical AB an Affiliate of Baxter Intl, 1996) has gone furthest, with 13 descendants of its own.

How does its category compare?

All FKP devices

FKP System, Dialysate Delivery, Single Patient: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FKP

1in 2018

2013: 22018: 1
2007high 2 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FKP

Too few FKP 510(k)s in any year since 2007 to give a median.

Recalls per year · FKP

1in 2010

2008: 42009: 22010: 1
2007high 4 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FKP

4in 2021

2007: 82008: 262009: 222010: 92011: 42013: 22014: 12015: 12018: 52019: 12021: 4
2007high 26 (2008) · low 0 (2012)2025

MAUDE reports naming product code FKP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—08
20080—426
20090—222
20100—19
20110—04
20120—00
20132—02
20140—01
20150—01
20160—00
20170—00
20181—05
20190—01
20200—00
20210—04
20220—00
20230—00
20240—00
20250—00
20260—0—
Cite this page

510(k) Report. “K910215: Drake Willock System 1000 Dialysate Delivery Syst.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K910215

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026