Drake Willock System & Altra Touch 1000 Dialysis Delivery System

K970446510(k) TraditionalCleared Jun 2, 1997117-day reviewClass II21 CFR 876.5860Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Althin Medical AB an Affiliate of Baxter Intl · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Jun 2, 1997 as substantially equivalent to K910215 (Cd Medical, Inc., 1991) and 5 other predicates after a 117-day review. Its family tree reaches back 2 generations to 1991. 11 later devices descend from it. It has been recalled 5 times.

Decided
Jun 2, 1997
Review
117 days
Class
Class II
Product code
KDI
Recalls
5
Predicates
6

Its longest line of descent runs back 2 generations to K910215 (1991). 7 ancestors in all, the oldest K873279 (1987). 11 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor1987: 1 other ancestor2002: 2 descendants cleared1999: 1 descendant cleared2007: 2 descendants cleared2004: 1 descendant cleared2009: 1 descendant cleared2005: 1 descendant cleared2008: 1 descendant cleared2022: 2 descendants clearedReference device: K935958 · Crit-Line Blood Chamber with E-Beam · In-Line Diagnostics Corp. · 1995 · recalledK910215 · Drake Willock System 1000 Dialysate Delivery Syst · Cd Medical, Inc. · 1991 · recalledK954987 · Altratouch 1000 Hemodialysis Delivery System · Althin Medical AB an Affiliate of Baxter Intl · 1996 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors2 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 3 descendants5 generations forward: 1 descendant K954987 · Altratouch 1000 Hemodialysis Delivery System · Althin Medical AB an Affiliate of Baxter Intl · 1996 · recalledK910215 · Drake Willock System 1000 Dialysate Delivery Syst · Cd Medical, Inc. · 1991 · recalled0+5
  1. Root · 2 generations backK910215 Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc. · 1991 · Recalled
  2. 1 generation backK954987 Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1996 · Recalled
  3. This deviceK970446 Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1997
  4. Descendant · cites it directly · 9 of its ownK993064 Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products · 1999 · Recalled
  5. Descendant · 2 generations on · 8 of its ownK021615 Baxter Accura System, Model 5M5660Baxter Healthcare Corp · 2002 · Recalled
  6. Descendant · 3 generations on · 4 of its ownK041005 Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products · 2004 · Recalled
  7. Descendant · 4 generations on · 1 of its ownK080519 Prismaflex M150 SetGambro Renal Products, Inc. · 2008 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK070320 Edwards Aquarius Hemofiltration System, Model Abm 01Edwards Lifesciences Services GmbH · 2007 · Recalled

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

1999: 12002: 22004: 12005: 12007: 22008: 12009: 12022: 2
1999high 2 (2002) · low 0 (2000)2026

Most influential descendants

Showing 5 of 11 · See all
  • K993064Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products · 1999 · cites it directly9descendants
  • K021615Baxter Accura System, Model 5M5660Baxter Healthcare Corp · 2002 · 2 generations on8descendants
  • K041005Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products · 2004 · 3 generations on4descendants
  • K080519Prismaflex M150 SetGambro Renal Products, Inc. · 2008 · 4 generations on1descendant
  • K070320Edwards Aquarius Hemofiltration System, Model Abm 01Edwards Lifesciences Services GmbH · 2007 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K910215Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc.19915
−1K954987Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl19961
0 (this device)K970446Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl1997—
+1K993064Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products19991
+2K021615Baxter Accura System, Model 5M5660Baxter Healthcare Corp20023
+3K041005Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products20047
+4K080519Prismaflex M150 SetGambro Renal Products, Inc.20083
+2K070320Edwards Aquarius Hemofiltration System, Model Abm 01Edwards Lifesciences Services GmbH20071

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199310
199510
199610
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200220
200410
200510
200720
200810
200910
202220
FIG. 1 — Lineage of K970446, 1991–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K910215Drake Willock System 1000 Dialysate Delivery Syst
Summary page 1
…on: The Hematocrit and Blood Volume (HCT/BV) Monitor feature is a modification to the predicate System 1000®/AltraTouch™ 1000 dialysis delivery system [510(k) numbers * K910215 System 1000® Dialysis Delivery System, K954987 AltraTouch™ 1000 machine and K955384 SND option.] that integrates the stand-alone…
Cd Medical, Inc.19915
*K954987Altratouch 1000 Hemodialysis Delivery System
Summary page 1
…onitor feature is a modification to the predicate System 1000®/AltraTouch™ 1000 dialysis delivery system [510(k) numbers * K910215 System 1000® Dialysis Delivery System, K954987 AltraTouch™ 1000 machine and K955384 SND option.] that integrates the stand-alone Crit-Line™ II Instrument for measuring hematoc…
Althin Medical AB an Affiliate of Baxter Intl19961
*K955384Drake Willock System 1000 Dialysate Delivery System W/Single Needle Single Lumen Option
Summary page 1
…e predicate System 1000®/AltraTouch™ 1000 dialysis delivery system [510(k) numbers * K910215 System 1000® Dialysis Delivery System, K954987 AltraTouch™ 1000 machine and K955384 SND option.] that integrates the stand-alone Crit-Line™ II Instrument for measuring hematocrit (Hct), percent blood volume change…
Althin Medical AB an Affiliate of Baxter Intl1997—
*K924167Crit-Line
Summary page 1
…nd-alone Crit-Line™ II Instrument for measuring hematocrit (Hct), percent blood volume change (ABV), oxygen saturation (O, Sat) and access recirculation [510(k) numbers K924167 Crit-Line I with Blood Chamber, K935958 Blood Chamber with E-Beam, K950942 Crit-Line II with Hematocrit Alarm feature and K95387…
Noninvasive Medical Technology Corp.1993—
*K950942Crit-Line Hematocrit Alert
Summary page 1
…lood volume change (ABV), oxygen saturation (O, Sat) and access recirculation [510(k) numbers K924167 Crit-Line I with Blood Chamber, K935958 Blood Chamber with E-Beam, K950942 Crit-Line II with Hematocrit Alarm feature and K953875 Crit-Line II with Access Recirculation feature.] into the dialysis deliv…
In-Line Diagnostics Corp.1995—
*K953875Crit-Line Hematocrit Alert (Modification)
Summary page 1
…d access recirculation [510(k) numbers K924167 Crit-Line I with Blood Chamber, K935958 Blood Chamber with E-Beam, K950942 Crit-Line II with Hematocrit Alarm feature and K953875 Crit-Line II with Access Recirculation feature.] into the dialysis delivery system. The primary reason for making the modificat…
In-Line Diagnostics Corp.1996—
†K935958Crit-Line Blood Chamber with E-BeamIn-Line Diagnostics Corp.19951

Also cleared against K910215: 3 other devices →

What descends from it? (11)

See all 11

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K993064Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products19999
K013562Aurora System 1000 Series Home Self-Care Single Patient Delivery System; Models- Sys…Baxter Healthcare Corp20020

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 19, 2011Not classifiedZ-0173-2012Employee errorThe Tina System was released to the customer with OOS values for the Pressure Displays Accuracy and Leakage Rate Test.
Jun 21, 2006Not classifiedZ-1253-06OtherThe UF Removal Regulators were assembled incorrectly. The mis-assembly results in a failure to successfully calibrate the instrument and begin patient therapy.
Jun 14, 2005Not classifiedZ-1394-05OtherOperator's Manuals addenda are being issued to provide correct information concerning the correct syringe size to use for heparin infusion, the use of chemical disinfectants in the System 1000, TMP Advise, and Bicarbonate Value…
Dec 14, 2004Not classifiedZ-0367-05OtherAir bubbles have been observed in the tubing sets past the air detector of the instrument with no alarms. Air bubbles in the circuit can cause an air embolism.
Feb 20, 2004Not classifiedZ-0583-04OtherThese devices have the potential to overheat in the absence of a Ground Fault Circuit Interrupter (GFCI). This heater system failure could result in a fire.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 5.3 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • Of its descendants, K993064 (Gambro Renal Care Products, 1999) has gone furthest, with 9 descendants of its own.
  • 3 other devices also cite its predicate K910215.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The intended use of the device is to provide hemodialysis treatments in the acute and chronic setting, including high flux hemodialysis. The device is intended to be used by trained operators when prescribed by a physician. The device is intended to be use in conjunction with a hollow fiber or parallel plate dialyzer. The intended use of this device is to monitor Hematocrit, Percent Blood Volume Change, Oxygen Saturation value, and Access Recirculation.…

Read the full statement (153 words)

The intended use of the device is to provide hemodialysis treatments in the acute and chronic setting, including high flux hemodialysis.

The device is intended to be used by trained operators when prescribed by a physician.

The device is intended to be use in conjunction with a hollow fiber or parallel plate dialyzer.

The intended use of this device is to monitor Hematocrit, Percent Blood Volume Change, Oxygen Saturation value, and Access Recirculation.

The device is intended to be used as an integrated option of the System 1000° Dialysis Delivery System.

The device is intended to be used by trained operators when prescribed by a physician.

The device is intended to be used in conjunction with the Crit-Line disposable Blood Chamber. The Blood Chamber is a non reusable device.

To provide the medical professional with a means of monitoring Hematocrit, Percent Blood Volume Change, Oxygen Saturation value, and Access Recirculation during hemodialysis treatment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970446: Drake Willock System & Altra Touch 1000 Dialysis Delivery System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K970446

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026