510(k) · Cleared
Recalled · Class II · 2026Gambro Prismaflex and Gambro Prismaflex M60 & M100 Sets
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Gambro Renal Products · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Cleared on Oct 7, 2004 as substantially equivalent to K021615 (Baxter Healthcare Corp, 2002) and 3 other predicates after a 171-day review. Its family tree reaches back 5 generations to 1991. 1 of the 30 devices in its family, counting itself, has a Class I recall. 4 later devices descend from it. It has been recalled 7 times.
- Decided
- Oct 7, 2004
- Review
- 171 days
- Class
- Class II
- Product code
- KDI
- Recalls
- 7
- Predicates
- 4
Its longest line of descent runs back 5 generations to K910215 (1991). 25 ancestors in all, the oldest K850569 (1985). 4 later devices descend from it, 3 citing it directly. 1 of the 30 devices in this family, counting this one, has a Class I recall.
- Root · 5 generations backK910215 Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc. · 1991 · Recalled
- 4 generations backK954987 Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1996 · Recalled
- 3 generations backK970446 Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1997 · Recalled
- 2 generations backK993064 Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products · 1999 · Recalled
- 1 generation backK021615 Baxter Accura System, Model 5M5660Baxter Healthcare Corp · 2002 · Recalled
- This deviceK041005 Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products · 2004
- Descendant · cites it directly · 1 of its ownK080519 Prismaflex M150 SetGambro Renal Products, Inc. · 2008 · Recalled
- Descendant · cites it directly · 1 of its ownK042938 Gambro Prismaflex HF1000 and HF1400 SetsGambro Renal Products · 2005 · Recalled
- Descendant · 2 generations on · 0 of its ownK220281 multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC · 2022
- Descendant · cites it directly · 0 of its ownK212216 Prismaflex ST60 Set, Prismaflex ST100 Set, Prismaflex ST150 setBaxter Healthcare Corporation · 2022 · Recalled
Class I recalls in this family
- K963440 Dialog Dialysis Machinedirect predicate · 1998
Most influential descendants
Showing 4 of 4 · See all- K080519Prismaflex M150 SetGambro Renal Products, Inc. · 2008 · cites it directly1descendant
- K042938Gambro Prismaflex HF1000 and HF1400 SetsGambro Renal Products · 2005 · cites it directly1descendant
- K220281multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC · 2022 · 2 generations on0descendants
- K212216Prismaflex ST60 Set, Prismaflex ST100 Set, Prismaflex ST150 setBaxter Healthcare Corporation · 2022 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K910215 | Drake Willock System 1000 Dialysate Delivery Syst | Cd Medical, Inc. | 1991 | 5 |
| −4 | K954987 | Altratouch 1000 Hemodialysis Delivery System | Althin Medical AB an Affiliate of Baxter Intl | 1996 | 1 |
| −3 | K970446 | Drake Willock System & Altra Touch 1000 Dialysis Delivery System | Althin Medical AB an Affiliate of Baxter Intl | 1997 | 5 |
| −2 | K993064 | Prisma CFM (Continuous Fluid Management System) | Gambro Renal Care Products | 1999 | 1 |
| −1 | K021615 | Baxter Accura System, Model 5M5660 | Baxter Healthcare Corp | 2002 | 3 |
| 0 (this device) | K041005 | Gambro Prismaflex and Gambro Prismaflex M60 & M100 Sets | Gambro Renal Products | 2004 | — |
| +1 | K080519 | Prismaflex M150 Set | Gambro Renal Products, Inc. | 2008 | 3 |
| +1 | K042938 | Gambro Prismaflex HF1000 and HF1400 Sets | Gambro Renal Products | 2005 | 7 |
| +2 | K220281 | multiFiltratePRO System | Fresenius Medical Care Renal Therapies Group, LLC | 2022 | — |
| +1 | K212216 | Prismaflex ST60 Set, Prismaflex ST100 Set, Prismaflex ST150 set | Baxter Healthcare Corporation | 2022 | 3 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1985 | 2 | 0 |
| 1987 | 1 | 0 |
| 1988 | 1 | 0 |
| 1991 | 2 | 0 |
| 1993 | 1 | 0 |
| 1994 | 1 | 0 |
| 1995 | 1 | 0 |
| 1996 | 1 | 0 |
| 1997 | 3 | 0 |
| 1998 | 4 | 1 |
| 1999 | 1 | 0 |
| 2002 | 1 | 0 |
| 2003 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
| 2008 | 1 | 0 |
| 2022 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KDI device
- #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K021615 | Baxter Accura System, Model 5M5660Summary page 4…terology - Urology (GU) CLASSIFICATION: Class II per 21 CFR 876.5860 PREDICATE DEVICES: Baxter Accura System K021615 B.Braun Diapact CRRT K963440 Gambro… | 2002 |
| * | K963440 | Dialog Dialysis MachineSummary page 4…6.5860 PREDICATE DEVICES: Baxter Accura System K021615 B.Braun Diapact CRRT K963440 Gambro Prisma System K010805 Gambro… | 1998 |
| * | K010805 | Prisma SystemSummary page 4…K021615 B.Braun Diapact CRRT K963440 Gambro Prisma System K010805 Gambro Prisma M60/M100 Sets K032431 DEVICE DESCRIPTION: The Prismaflex Syste… | 2002 |
| * | K032431 | Prisma M60/M100 SetsSummary page 4…K963440 Gambro Prisma System K010805 Gambro Prisma M60/M100 Sets K032431 DEVICE DESCRIPTION: The Prismaflex System is an acute renal failure treatment device for removal of waste products, restorat… | 2003 |
What descends from it? (4)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 6, 2026 | Class II | Z-1227-2026 | Component design/selection | There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. |
| Sep 14, 2022 | Class II | Z-0081-2023 | Error in labeling | Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions. |
| Jul 22, 2021 | Class II | Z-2378-2021 | Vendor change control | Products labeled as sterile were distributed, but may not have been sterilized. " Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter. Data was manipulated to appear to be conforming |
| Jul 18, 2014 | Class II | Z-2202-2014 | Device Design | Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood. |
| Jul 18, 2014 | Class II | Z-2201-2014 | Device Design | Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood. |
| Jan 4, 2007 | Not classified | Z-0695-2007 | Device Design | Under certain conditions, * an excessive amount of anticoagulant may be infused into the patient. (* interruption of a self-test by an alarm and obstruction of the access lines) |
| May 16, 2006 | Not classified | Z-0032-2007 | Software design | A potential risk of infusing air or infusion fluid to the patient during dialysis. Also, a possibility of an incorrect scale reading of the amount of patient fluid removal. |
The 5 latest of 7 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K963440Dialog Dialysis MachineB. Braun Medical AG · 1998 · Direct predicate · ◆ in the 1998 ancestors bar of FIG. 16 Class I recalls
What does this lineage say?
- The line of descent spans 13 years and passes through 5 companies; a new generation was cleared every 2.3 years on median.
- 11 of the 25 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K080519 (Gambro Renal Products, Inc., 2008) has gone furthest, with 1 descendant of its own.
- 2 other devices also cite its predicate K021615.
How does its category compare?
All KDI devicesKDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KDI
9in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KDI
199in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KDI
2in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · KDI
2,786in 2025
MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 11 | 91 | 4 | 916 |
| 2008 | 6 | 148 | 0 | 1,042 |
| 2009 | 4 | 236 | 3 | 522 |
| 2010 | 9 | 186 | 1 | 823 |
| 2011 | 5 | 85 | 4 | 2,270 |
| 2012 | 6 | 153 | 1 | 1,765 |
| 2013 | 11 | 247 | 2 | 2,360 |
| 2014 | 8 | 164 | 3 | 6,154 |
| 2015 | 5 | 69 | 5 | 4,182 |
| 2016 | 5 | 230 | 4 | 766 |
| 2017 | 8 | 188 | 1 | 963 |
| 2018 | 2 | — | 1 | 1,716 |
| 2019 | 4 | 263 | 6 | 1,587 |
| 2020 | 9 | 266 | 0 | 1,633 |
| 2021 | 2 | — | 2 | 1,944 |
| 2022 | 7 | 259 | 17 | 2,180 |
| 2023 | 7 | 157 | 8 (8) | 2,692 |
| 2024 | 7 | 235 | 4 (2) | 2,441 |
| 2025 | 9 | 199 | 2 | 2,786 |
| 2026 | 6 | 154 | 2 | — |
What is it indicated for?
The Prismaflex is indicated for the following use:
continuous solute and/or fluid removal in patients with acute renal failure or fluid overload
All treatments administered by the Prismaflex must be prescribed by a physician.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K041005 · printed Nov 4, 2026
Cite this page
510(k) Report. “K041005: Gambro Prismaflex and Gambro Prismaflex M60 & M100 Sets.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K041005
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026