Hotline Fluid Warmer

K911383510(k) TraditionalCleared Oct 17, 1991203-day reviewClass II21 CFR 880.5725Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Level 1 Technologies, Inc. · LGZ Warmer, Thermal, Infusion Fluid

Cleared on Oct 17, 1991 after a 203-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 18 devices in its family, counting itself, have a Class I recall. 17 later devices descend from it. It has been recalled 11 times.

Decided
Oct 17, 1991
Review
203 days
Class
Class II
Product code
LGZ
Recalls
11
Predicates
0

17 later devices descend from it, 2 citing it directly. 2 of the 18 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2001: 1 descendant cleared1999: 1 descendant cleared2011: 1 descendant cleared2006: 2 descendants cleared, 1 with a Class I recall12012: 3 descendants cleared2024: 1 descendant cleared2016: 1 descendant cleared2023: 1 descendant cleared2019: 2 descendants cleared2013: 1 descendant cleared, 1 with a Class I recall12017: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant cleared1990200020102020
  1. This deviceK911383 Hotline Fluid WarmerLevel 1 Technologies, Inc. · 1991
  2. Descendant · cites it directly · 16 of its ownK984640 Thermal AngelEstill Medical Technologies, Inc. · 1999 · Recalled
  3. Descendant · cites it directly · 13 of its ownK012031 Modification to Thermal AngelEstill Medical Technologies, Inc. · 2001 · Recalled
  4. Descendant · 2 generations on · 12 of its ownK060537 Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model…Enginivity, LLC · 2006 · Recalled (Class I)
  5. Descendant · 3 generations on · 10 of its ownK112902 Enflow IV Fluid WarmerVital Signs, Inc. · 2012
  6. Descendant · 4 generations on · 1 of its ownK232107 °M Warmer SystemMequ A/S · 2024

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

1999: 12001: 12006: 22011: 12012: 32013: 12016: 12017: 12019: 22023: 12024: 12025: 12026: 1
1999high 3 (2012) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 17 · See all
  • K984640Thermal AngelEstill Medical Technologies, Inc. · 1999 · cites it directly16descendants
  • K012031Modification to Thermal AngelEstill Medical Technologies, Inc. · 2001 · cites it directly13descendants
  • K060537Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model 200 and Eflow AC…Enginivity, LLC · 2006 · 2 generations on12descendants
  • K112902Enflow IV Fluid WarmerVital Signs, Inc. · 2012 · 3 generations on10descendants
  • K232107°M Warmer SystemMequ A/S · 2024 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K911383Hotline Fluid WarmerLevel 1 Technologies, Inc.1991—
+1K984640Thermal AngelEstill Medical Technologies, Inc.19991
+1K012031Modification to Thermal AngelEstill Medical Technologies, Inc.20011
+2K060537Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model 200 and…Enginivity, LLC2006Class I
+3K112902Enflow IV Fluid WarmerVital Signs, Inc.2012—
+4K232107°M Warmer SystemMequ A/S2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200110
200621
201110
201230
201311
201610
201710
201920
202310
202410
202510
202610
FIG. 1 — Lineage of K911383, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LGZ device

  1. #1 K260988 °M Warmer System with °M Stationscore 12.7 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (17)

See all 17

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K984640Thermal AngelEstill Medical Technologies, Inc.199916
K012031Modification to Thermal AngelEstill Medical Technologies, Inc.200113

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 24, 2026Class IIZ-2759-2026Process controlICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
Mar 17, 2014Class IIZ-1444-2014Nonconforming Material/ComponentInjection port of the Set may leak blood/ IV fluids from the septum of the injection port or the septum may become dislodged
Mar 17, 2014Class IIZ-1443-2014Nonconforming Material/ComponentInjection port of the Set may leak blood/ IV fluids from the septum of the injection port or the septum may become dislodged.
Jun 27, 2012Class IIZ-1979-2012Device DesignHOTLINE 3 Blood and Fluid Warmer( Reorder No. HL-390 ) disposable is not correctly installed , the microswitch may become damaged and unusable
Nov 21, 2011Not classifiedZ-0424-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Nov 21, 2011Not classifiedZ-0425-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Nov 21, 2011Not classifiedZ-0423-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Nov 21, 2011Not classifiedZ-0427-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Nov 21, 2011Not classifiedZ-0422-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Nov 21, 2011Not classifiedZ-0428-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets
Nov 21, 2011Not classifiedZ-0426-2012Device DesignDisconnected/ loose Luer connectors at the patient end of the triple lumen tubing of the HOTLlNE Disposable Administration Sets and Level 1 D/DI-60HL Normothermic IV Administration Sets

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K060537Eflow IV Fluid Warmer, Model 100, Eflow Disposable Cartridge, Model 200 and Eflow AC Power Supply, Model 120Enginivity, LLC · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 12 Class I recalls
  • K130867Enflow IV Fluid Warmer Insulated StrapVital Signs, Inc., A GE Healthcare Company · 2013 · 5 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K984640 (Estill Medical Technologies, Inc., 1999) has gone furthest, with 16 descendants of its own.

How does its category compare?

All LGZ devices

LGZ Warmer, Thermal, Infusion Fluid: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LGZ

2in 2025

2007: 22008: 42009: 22010: 22011: 32012: 72013: 52015: 22016: 22017: 22018: 12019: 32022: 32023: 12024: 12025: 22026 (partial year): 2
2007high 7 (2012) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LGZ

546in 2022

2008: 99 days median2011: 77 days median2012: 162 days median2013: 143 days median2019: 209 days median2022: 546 days median
2007high 546 (2022) · low 77 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LGZ

1in 2016

2009: 12010: 12011: 82015: 12016: 12026 (partial year): 1
2007high 8 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LGZ

376in 2025

2007: 112008: 102009: 122010: 432011: 542012: 522013: 162014: 232015: 342016: 152017: 582018: 942019: 3402020: 1,2212021: 8852022: 1,5482023: 1,0332024: 8622025: 376
2007high 1,548 (2022) · low 10 (2008)2025

MAUDE reports naming product code LGZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—011
2008499010
20092—112
20102—143
2011377854
20127162052
20135143016
20140—023
20152—134
20162—115
20172—058
20181—094
201932090340
20200—01,221
20210—0885
2022354601,548
20231—01,033
20241—0862
20252—0376
20262—1—
Cite this page

510(k) Report. “K911383: Hotline Fluid Warmer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K911383

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026