510(k) Report

510(k) · Cleared

Cleared

Advantage Windows 3D with Navigator Option

K954355510(k) TraditionalCleared Nov 14, 199557-day reviewClass II21 CFR 892.1000

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

General Electric Co. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Nov 14, 1995 after a 57-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 of the 759 devices in its family, counting itself, has a Class I recall. 758 later devices descend from it.

Decided
Nov 14, 1995
Review
57 days
Class
Class II
Product code
LNH
Recalls
None
Predicates
0

758 later devices descend from it, 10 citing it directly. 1 of the 759 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 1999: 3 descendants cleared1997: 2 descendants cleared2004: 7 descendants cleared2003: 8 descendants cleared2001: 2 descendants cleared2006: 19 descendants cleared2002: 3 descendants cleared1998: 1 descendant cleared2013: 24 descendants cleared2012: 28 descendants cleared2005: 6 descendants cleared2009: 13 descendants cleared2008: 16 descendants cleared2010: 13 descendants cleared2014: 27 descendants cleared2016: 29 descendants cleared2025: 71 descendants cleared, 1 with a Class I recall12011: 22 descendants cleared2017: 32 descendants cleared2007: 7 descendants cleared2021: 47 descendants cleared2018: 50 descendants cleared2024: 59 descendants cleared2015: 25 descendants cleared2022: 52 descendants cleared2020: 53 descendants cleared2019: 38 descendants cleared2023: 46 descendants cleared2026: 55 descendants cleared1995200020052010201520202025
  1. This deviceK954355 Advantage Windows 3D with Navigator OptionGeneral Electric Co. · 1995
  2. Descendant · cites it directly · 580 of its ownK012313 CT Colonography/NAVIGATOR2Ge Medical Systems, Inc. · 2001 · Recalled
  3. Descendant · cites it directly · 516 of its ownK972399 Advantage Windows Volume Rendering OptionGe Medical Systems, Inc. · 1997 · Recalled
  4. Descendant · 2 generations on · 469 of its ownK041521 Volume Viewer PlusGeneral Electric Medical Systems · 2004 · Recalled
  5. Descendant · cites it directly · 298 of its ownK974715 Advantage 3D XRGE Medical Systems · 1998
  6. Descendant · 2 generations on · 245 of its ownK984634 3D-AngioSiemens Medical Corp. · 1999

Showing 5 of 758 descendants. See all descendants

Descendants cleared per year

1997: 21998: 11999: 32001: 22002: 32003: 82004: 72005: 62006: 192007: 72008: 162009: 132010: 132011: 222012: 282013: 242014: 272015: 252016: 292017: 322018: 502019: 382020: 532021: 472022: 522023: 462024: 592025: 712026: 55
1997high 71 (2025) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 758 · See all
  • K012313CT Colonography/NAVIGATOR2Ge Medical Systems, Inc. · 2001 · cites it directly580descendants
  • K972399Advantage Windows Volume Rendering OptionGe Medical Systems, Inc. · 1997 · cites it directly516descendants
  • K041521Volume Viewer PlusGeneral Electric Medical Systems · 2004 · 2 generations on469descendants
  • K974715Advantage 3D XRGE Medical Systems · 1998 · cites it directly298descendants
  • K9846343D-AngioSiemens Medical Corp. · 1999 · 2 generations on245descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K954355Advantage Windows 3D with Navigator OptionGeneral Electric Co.1995—
+1K012313CT Colonography/NAVIGATOR2Ge Medical Systems, Inc.20011
+1K972399Advantage Windows Volume Rendering OptionGe Medical Systems, Inc.19971
+2K041521Volume Viewer PlusGeneral Electric Medical Systems20042
+1K974715Advantage 3D XRGE Medical Systems1998—
+2K9846343D-AngioSiemens Medical Corp.1999—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199720
199810
199930
200120
200230
200380
200470
200560
2006190
200770
2008160
2009130
2010130
2011220
2012280
2013240
2014270
2015250
2016290
2017320
2018500
2019380
2020530
2021470
2022520
2023460
2024590
2025711
2026550
FIG. 1 — Lineage of K954355, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (758)

See all 758

10 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K012313CT Colonography/NAVIGATOR2Ge Medical Systems, Inc.2001580
K972399Advantage Windows Volume Rendering OptionGe Medical Systems, Inc.1997516
K974715Advantage 3D XRGE Medical Systems1998298
K032483Viatronix V3D Explorer, Revision 1.2Viatronix, Inc.200376
K993792Smart Vessel Analysis (Smart VA) OptionGe Medical Systems, Inc.199960
K962010Epi-Scope Option for Q Series CT SystemsPhilips Medical Systems (Cleveland), Inc.19979
K020658Viatronix V3D ColonViatronix, Inc.20024
K040126Viatronix V3D Colon, Revision 1.3Viatronix, Inc.20042
K061532Gentle Colon (Revision 2.0)Rendoscopy, Inc.20060
K022789Viatronix V3D ExplorerViatronix, Inc.20020

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 9 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K012313 (Ge Medical Systems, Inc., 2001) has gone furthest, with 580 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—
Cite this page

510(k) Report. “K954355: Advantage Windows 3D with Navigator Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K954355

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026