510(k) · Cleared
Recalled · Class II · 2012Volume Viewer Plus
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
General Electric Medical Systems · JAK System, X-Ray, Tomography, Computed
Cleared on Jun 22, 2004 as substantially equivalent to K972399 (Ge Medical Systems, Inc., 1997) and 1 other predicate after a 14-day review. Its family tree reaches back 2 generations to 1987. 1 of the 475 devices in its family, counting itself, has a Class I recall. 469 later devices descend from it. It has been recalled 2 times.
- Decided
- Jun 22, 2004
- Review
- 14 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 2
- Predicates
- 2
Its longest line of descent runs back 2 generations to K871859 (1987). 5 ancestors in all. 469 later devices descend from it, 17 citing it directly. 1 of the 475 devices in this family, counting this one, has a Class I recall.
- Root · 2 generations backK871859 CT9800 Quick 3D #B7919DAGeneral Electric Co. · 1987
- 1 generation backK972399 Advantage Windows Volume Rendering OptionGe Medical Systems, Inc. · 1997 · Recalled
- This deviceK041521 Volume Viewer PlusGeneral Electric Medical Systems · 2004
- Descendant · cites it directly · 217 of its ownK061696 Iqqa-Liver SoftwareEdda Technology, Inc. · 2006
- Descendant · cites it directly · 204 of its ownK063324 Pet VcarGE Healthcare · 2007 · Recalled
- Descendant · 2 generations on · 200 of its ownK091373 Syngo TruedSiemens Medical Solutions USA, Inc. · 2009
- Descendant · 2 generations on · 195 of its ownK071310 Syngo CT Oncology Software PackageSiemens AG · 2007
- Descendant · 3 generations on · 117 of its ownK083524 Syngo Dual Energy with Extended FunationallySiemens Medical Solutions USA, Inc. · 2009
Showing 5 of 469 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K250443 MAGNETOM Avanto Fit; MAGNETOM Skyra Fit…7 generations on · 2025
Most influential descendants
Showing 5 of 469 · See all- K061696Iqqa-Liver SoftwareEdda Technology, Inc. · 2006 · cites it directly217descendants
- K063324Pet VcarGE Healthcare · 2007 · cites it directly204descendants
- K091373Syngo TruedSiemens Medical Solutions USA, Inc. · 2009 · 2 generations on200descendants
- K071310Syngo CT Oncology Software PackageSiemens AG · 2007 · 2 generations on195descendants
- K083524Syngo Dual Energy with Extended FunationallySiemens Medical Solutions USA, Inc. · 2009 · 3 generations on117descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K871859 | CT9800 Quick 3D #B7919DA | General Electric Co. | 1987 | — |
| −1 | K972399 | Advantage Windows Volume Rendering Option | Ge Medical Systems, Inc. | 1997 | 1 |
| 0 (this device) | K041521 | Volume Viewer Plus | General Electric Medical Systems | 2004 | — |
| +1 | K061696 | Iqqa-Liver Software | Edda Technology, Inc. | 2006 | — |
| +1 | K063324 | Pet Vcar | GE Healthcare | 2007 | 1 |
| +2 | K091373 | Syngo Trued | Siemens Medical Solutions USA, Inc. | 2009 | — |
| +2 | K071310 | Syngo CT Oncology Software Package | Siemens AG | 2007 | — |
| +3 | K083524 | Syngo Dual Energy with Extended Funationally | Siemens Medical Solutions USA, Inc. | 2009 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1995 | 1 | 0 |
| 2001 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 2 | 0 |
| 2006 | 5 | 0 |
| 2007 | 5 | 0 |
| 2008 | 5 | 0 |
| 2009 | 5 | 0 |
| 2010 | 7 | 0 |
| 2011 | 11 | 0 |
| 2012 | 15 | 0 |
| 2013 | 14 | 0 |
| 2014 | 19 | 0 |
| 2015 | 19 | 0 |
| 2016 | 20 | 0 |
| 2017 | 20 | 0 |
| 2018 | 30 | 0 |
| 2019 | 25 | 0 |
| 2020 | 38 | 0 |
| 2021 | 32 | 0 |
| 2022 | 34 | 0 |
| 2023 | 35 | 0 |
| 2024 | 40 | 0 |
| 2025 | 47 | 1 |
| 2026 | 41 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K972399 | Advantage Windows Volume Rendering OptionSummary page 1…Advantage Windows Volume Rendering Option * Manufacturer: General Electric Medical Systems, Buc, France 510(k)#: K972399 * Model: CT Colonography / Navigator2 * Manufacturer: General Electric Medical Sys… | 1997 |
| * | K012313 | CT Colonography/NAVIGATOR2Summary page 1…del: CT Colonography / Navigator2 * Manufacturer: General Electric Medical Systems, Buc, France 510(k) #: K012313 Device Description: Volume Viewer Plus is a software package to be used on the GE Advantage Workstation, the GE Centricity PAC… | 2001 |
What descends from it? (469)
See all 46917 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K061696 | Iqqa-Liver Software | 2006 | 217 |
| K063324 | Pet Vcar | 2007 | 204 |
| K052995 | Advantage Workstation 4.3. | 2005 | 113 |
| K092639 | Innova Vision, Model S18751VN, Innova Trackvision, Model S18751VT, Innova Epvision… | 2009 | 65 |
| K070248 | Velocityais | 2007 | 65 |
| K081076 | Velocityais, Velocity AI, Velocityvx, Velocityrt, V-Response | 2008 | 63 |
| K110573 | Ready View | 2011 | 38 |
| K052345 | Advantage Sim MD | 2005 | 21 |
| K131498 | Iqqa-Liver Multimodality Software | 2013 | 16 |
| K141745 | Iqqa-Bodyimaging Software | 2014 | 6 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 25, 2012 | Class II | Z-0901-2013 | Software design | Incorrect views and measurements possible. GE Healthcare became aware of multiple potential safety issues involving the Advantage workstation, Volume Viewer Plus and Volume Analysis. |
| May 1, 2008 | Not classified | Z-2312-2008 | Software design | GE Healthcare has recently became aware of a possible mismatch between the label of the tracked vessel and the underlying image associated with the cardio- vascular applications of the Advantage workstation that may impact patient safety. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 7 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls
What does this lineage say?
- The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 10.1 years on median.
- 2 of the 5 ancestors have a recall of any class.
- Of its descendants, K061696 (Edda Technology, Inc., 2006) has gone furthest, with 217 descendants of its own.
- 1 other device also cites its predicate K972399.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
Volume Viewer Plus is a medical diagnostic software that allows the processing, review, analysis and communication of 3D reconstructed images and their relationship to originally acquired images from CT, MR, X-Ray Angio and PET Scanning devices. The combination of acquired images, reconstructed images, annotations and measurements performed by the clinician are intended to provide to the referring physician clinically relevant information for diagnosis, surgery and treatment planning.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K041521 · printed Nov 6, 2026
Cite this page
510(k) Report. “K041521: Volume Viewer Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041521
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026