510(k) · Cleared
Recalled · Class II · 2012Advantage Windows Volume Rendering Option
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Medical Systems, Inc. · JAK System, X-Ray, Tomography, Computed
Cleared on Sep 23, 1997 as substantially equivalent to K923077 (GE Medical Systems, 1993) and 2 other predicates after an 89-day review. Its family tree reaches back 1 generation to 1987. 1 of the 520 devices in its family, counting itself, has a Class I recall. 516 later devices descend from it. It has been recalled once.
- Decided
- Sep 23, 1997
- Review
- 89 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 1
- Predicates
- 3
Its longest line of descent runs back 1 generation to K871859 (1987). 3 ancestors in all. 516 later devices descend from it, 2 citing it directly. 1 of the 520 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backK871859 CT9800 Quick 3D #B7919DAGeneral Electric Co. · 1987
- This deviceK972399 Advantage Windows Volume Rendering OptionGe Medical Systems, Inc. · 1997
- Descendant · cites it directly · 469 of its ownK041521 Volume Viewer PlusGeneral Electric Medical Systems · 2004 · Recalled
- Descendant · 2 generations on · 217 of its ownK061696 Iqqa-Liver SoftwareEdda Technology, Inc. · 2006
- Descendant · 2 generations on · 204 of its ownK063324 Pet VcarGE Healthcare · 2007 · Recalled
- Descendant · 3 generations on · 200 of its ownK091373 Syngo TruedSiemens Medical Solutions USA, Inc. · 2009
- Descendant · 3 generations on · 195 of its ownK071310 Syngo CT Oncology Software PackageSiemens AG · 2007
Showing 5 of 516 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K250443 MAGNETOM Avanto Fit; MAGNETOM Skyra Fit…8 generations on · 2025
Most influential descendants
Showing 5 of 516 · See all- K041521Volume Viewer PlusGeneral Electric Medical Systems · 2004 · cites it directly469descendants
- K061696Iqqa-Liver SoftwareEdda Technology, Inc. · 2006 · 2 generations on217descendants
- K063324Pet VcarGE Healthcare · 2007 · 2 generations on204descendants
- K091373Syngo TruedSiemens Medical Solutions USA, Inc. · 2009 · 3 generations on200descendants
- K071310Syngo CT Oncology Software PackageSiemens AG · 2007 · 3 generations on195descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K871859 | CT9800 Quick 3D #B7919DA | General Electric Co. | 1987 | — |
| 0 (this device) | K972399 | Advantage Windows Volume Rendering Option | Ge Medical Systems, Inc. | 1997 | — |
| +1 | K041521 | Volume Viewer Plus | General Electric Medical Systems | 2004 | 2 |
| +2 | K061696 | Iqqa-Liver Software | Edda Technology, Inc. | 2006 | — |
| +2 | K063324 | Pet Vcar | GE Healthcare | 2007 | 1 |
| +3 | K091373 | Syngo Trued | Siemens Medical Solutions USA, Inc. | 2009 | — |
| +3 | K071310 | Syngo CT Oncology Software Package | Siemens AG | 2007 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1995 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 1 | 0 |
| 2004 | 1 | 0 |
| 2005 | 2 | 0 |
| 2006 | 8 | 0 |
| 2007 | 5 | 0 |
| 2008 | 5 | 0 |
| 2009 | 7 | 0 |
| 2010 | 7 | 0 |
| 2011 | 13 | 0 |
| 2012 | 17 | 0 |
| 2013 | 17 | 0 |
| 2014 | 20 | 0 |
| 2015 | 20 | 0 |
| 2016 | 24 | 0 |
| 2017 | 21 | 0 |
| 2018 | 31 | 0 |
| 2019 | 26 | 0 |
| 2020 | 40 | 0 |
| 2021 | 33 | 0 |
| 2022 | 37 | 0 |
| 2023 | 36 | 0 |
| 2024 | 43 | 0 |
| 2025 | 56 | 1 |
| 2026 | 46 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K923077 | Advantage Windows 3D Option and Dentascan OptionSummary page 1…Road ° New Berlin, W1 53151 Marketed Devices: The Advantage Windows Valume Rendering Option is substantially equivalent to the currently marketed Advantage Windows 3D (K923077), Advantage Windows 3D with Navigator Option (K954355), and Advantage CT (9800 Quick) 3D (KB871859) Indications for Use: Advanta… | 1993 |
| * | K954355 | Advantage Windows 3D with Navigator OptionSummary page 1…vantage Windows Valume Rendering Option is substantially equivalent to the currently marketed Advantage Windows 3D (K923077), Advantage Windows 3D with Navigator Option (K954355), and Advantage CT (9800 Quick) 3D (KB871859) Indications for Use: Advantage Windows Volurne Rendering Option is intended to pro… | 1995 |
| * | K871859 | CT9800 Quick 3D #B7919DASummary page 1…stantially equivalent to the currently marketed Advantage Windows 3D (K923077), Advantage Windows 3D with Navigator Option (K954355), and Advantage CT (9800 Quick) 3D (KB871859) Indications for Use: Advantage Windows Volurne Rendering Option is intended to provide fast, easy volume visualization of three… | 1987 |
What descends from it? (516)
See all 5162 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 25, 2012 | Class II | Z-0901-2013 | Software design | Incorrect views and measurements possible. GE Healthcare became aware of multiple potential safety issues involving the Advantage workstation, Volume Viewer Plus and Volume Analysis. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 8 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls
What does this lineage say?
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
The Advantage Windows Volume Rendering Option is intended to provide volume visualization of three dimensional structures imaged with computed tomography (CT), or magnetic resonance (MR). It reconstructs and displays translucent 3D models to be used for diagnosis confidence or planning surgical treatment. The images are reconstructed and displayed on the CT/MR Advantage Windows Diagnostic Workstation (K913770).
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K972399 · printed Nov 4, 2026
Cite this page
510(k) Report. “K972399: Advantage Windows Volume Rendering Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K972399
Report an extraction error