510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Sterrad 50 & Sterrad 100S Sterilizers

K023290510(k) TraditionalCleared Apr 3, 2003183-day reviewClass II21 CFR 880.6860Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advanced Sterilization Products · FLF Sterilizer, Ethylene-Oxide Gas

Cleared on Apr 3, 2003 as substantially equivalent to K981625 (Advanced Sterilization Products, 1999) and 1 other predicate after a 183-day review. Its family tree reaches back 1 generation to 1999. 18 later devices descend from it. It has been recalled 9 times.

Decided
Apr 3, 2003
Review
183 days
Class
Class II
Product code
FLF
Recalls
9
Predicates
2

Its longest line of descent runs back 1 generation to K981625 (1999). 2 ancestors in all. 18 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2012: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared2024: 2 descendants cleared2016: 5 descendants cleared2018: 2 descendants cleared2017: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared2019: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant clearedK981625 · Sterrad 50 Sterilizer · Advanced Sterilization Products · 1999 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 4 descendants3 generations forward: 7 descendants4 generations forward: 3 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant K981625 · Sterrad 50 Sterilizer · Advanced Sterilization Products · 1999 · recalled0+5
  1. Root · 1 generation backK981625 Sterrad 50 SterilizerAdvanced Sterilization Products · 1999 · Recalled
  2. This deviceK023290 Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products · 2003
  3. Descendant · cites it directly · 17 of its ownK111377 Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products · 2012 · Recalled
  4. Descendant · 2 generations on · 9 of its ownK142454 STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · Recalled
  5. Descendant · 2 generations on · 5 of its ownK141163 Sterizone SterilizerTso3, Inc. · 2014
  6. Descendant · 2 generations on · 4 of its ownK160903 STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products · 2016
  7. Descendant · 3 generations on · 4 of its ownK153392 STERIZONE VP4 SterilizerTs03, Inc. · 2016

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2012: 12014: 12015: 12016: 52017: 12018: 22019: 12021: 12022: 12024: 22025: 12026: 1
2012high 5 (2016) · low 0 (2013)2026

Most influential descendants

Showing 5 of 18 · See all
  • K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products · 2012 · cites it directly17descendants
  • K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · 2 generations on9descendants
  • K141163Sterizone SterilizerTso3, Inc. · 2014 · 2 generations on5descendants
  • K160903STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products · 2016 · 2 generations on4descendants
  • K153392STERIZONE VP4 SterilizerTs03, Inc. · 2016 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K981625Sterrad 50 SterilizerAdvanced Sterilization Products19995
0 (this device)K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products2003—
+1K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products20121
+2K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products20152
+2K141163Sterizone SterilizerTso3, Inc.2014—
+2K160903STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products2016—
+3K153392STERIZONE VP4 SterilizerTs03, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201410
201510
201650
201710
201820
201910
202110
202210
202420
202510
202610
FIG. 1 — Lineage of K023290, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLF device

  1. #1 K192978 EOGas 4 Ethylene Oxide Gas Sterilizerscore 0.6 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K981625Sterrad 50 Sterilizer
Summary page 1
…e: STERRAD@50 Sterilization System and STERRAD@100s Sterilization System 2.0 PREDICATE DEVICES STERRAD@ 50 Sterilization System (K98 1625) and STERRAD@ 100s Sterilization System (K99 1999) 3.0 INDICATIONS FOR USE This Premarket Notification is to…
Advanced Sterilization Products19995
*K991999Sterrad 100S Sterilizer
Summary page 1
…Sterilization System 2.0 PREDICATE DEVICES STERRAD@ 50 Sterilization System (K98 1625) and STERRAD@ 100s Sterilization System (K99 1999) 3.0 INDICATIONS FOR USE This Premarket Notification is to expand the Indications for Use for the STERRAD 50…
Advanced Sterilization Products20006

What descends from it? (18)

See all 18

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products201217

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 4, 2014Class IIZ-0842-2015Nonconforming Material/ComponentAdvanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
Nov 4, 2014Class IIZ-0843-2015Nonconforming Material/ComponentAdvanced Sterilization Products is recalling the STERRAD System cassettes because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.
Dec 16, 2013Class IIZ-0744-2014Under Investigation by firmAdvanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
Dec 16, 2013Class IIZ-0745-2014Under Investigation by firmAdvanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
Dec 16, 2013Class IIZ-0746-2014Under Investigation by firmAdvanced Sterilization Products (ASP) is recalling the STERRAD NX, STERRAD 50, and STERRAD 100S Sterilization Systems because some healthcare workers detected odors or smells possibly emanating from STERRAD Systems.
Mar 9, 2009Not classifiedZ-1419-2009Component design/selectionASP has identified a secondary, lower-occurring cause of oil mist within certain STERRAD NX Systems and STERRAD 50 Systems due to potentially defective oil fill plugs.
Oct 17, 2008Not classifiedZ-0340-2009Component design/selection1) Inability of the sterilizer to detect when an injection takes place without hydrogen peroxide being transferred to the vaporizer bowl. This situation can result in a cycle completion with the injection of insufficient hydrogen peroxide…
Feb 25, 2008Not classifiedZ-1629-2008Device DesignOil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."
Feb 25, 2008Not classifiedZ-1628-2008Device DesignOil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K111377 (Advanced Sterilization Products, 2012) has gone furthest, with 17 descendants of its own.

How does its category compare?

All FLF devices

FLF Sterilizer, Ethylene-Oxide Gas: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FLF

1in 2020

2009: 12015: 22017: 12020: 1
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLF

Too few FLF 510(k)s in any year since 2007 to give a median.

Recalls per year · FLF

2in 2014

2009: 12012: 12013: 22014: 2
2007high 2 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLF

4in 2021

2007: 22008: 12009: 22010: 32011: 42012: 52013: 32014: 62015: 32016: 32017: 32019: 22021: 4
2007high 6 (2014) · low 0 (2018)2025

MAUDE reports naming product code FLF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—02
20080—01
20091—12
20100—03
20110—04
20120—15
20130—23
20140—26
20152—03
20160—03
20171—03
20180—00
20190—02
20201—00
20210—04
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

This Premarket Notification is to expand the Indications for Use for the STERRAD 50 Sterilizer (K981625) and the STERRAD 100s Sterilizer (K991999) to include titanium as a compatible material and to expand the lumen sizes to include medical devices with only a single stainless steel lumen in the following configurations: An inside diameter of 1 mm or larger and a length of 125 mm or shorter An inside diameter of 2 mm or larger and a length of 250 mm or…

Read the full statement (130 words)

This Premarket Notification is to expand the Indications for Use for the STERRAD 50 Sterilizer (K981625) and the STERRAD 100s Sterilizer (K991999) to include titanium as a compatible material and to expand the lumen sizes to include medical devices with only a single stainless steel lumen in the following configurations:

An inside diameter of 1 mm or larger and a length of 125 mm or shorter

An inside diameter of 2 mm or larger and a length of 250 mm or shorter

The validation testing for these two new lumen sizes was conducted using a maximum of 10 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing.

For information pertaining to previously cleared materials and lumen sizes, please reference K981625 and K991999.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023290: Sterrad 50 & Sterrad 100S Sterilizers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023290

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026