510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Abbott Plum a+ Infusion Pump

K982159510(k) TraditionalCleared Jan 12, 1999207-day reviewClass II21 CFR 880.5725Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Laboratories · FRN Pump, Infusion

Cleared on Jan 12, 1999 after a 207-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 5 of the 13 devices in its family, counting itself, have a Class I recall. 12 later devices descend from it. It has been recalled 11 times.

Decided
Jan 12, 1999
Review
207 days
Class
Class II
Product code
FRN
Recalls
11
Predicates
0

12 later devices descend from it, 3 citing it directly. 5 of the 13 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2011: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12004: 1 descendant cleared2017: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2025: 2 descendants cleared1995200020052010201520202025
  1. This deviceK982159 Abbott Plum a+ Infusion PumpAbbott Laboratories · 1999
  2. Descendant · cites it directly · 10 of its ownK103224 Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe…Hospira, Inc. · 2011 · Recalled (Class I)
  3. Descendant · cites it directly · 9 of its ownK141789 Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play…Hospira, Inc. · 2015 · Recalled (Class I)
  4. Descendant · 2 generations on · 5 of its ownK161469 Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play…Hospira, Inc. · 2017 · Recalled (Class I)
  5. Descendant · 3 generations on · 2 of its ownK223607 Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · Recalled (Class I)
  6. Descendant · 2 generations on · 2 of its ownK153074 Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding…Covidien · 2016 · Recalled

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2004: 12011: 12015: 12016: 12017: 12019: 12021: 12022: 12023: 12025: 22026: 1
2004high 2 (2025) · low 0 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · cites it directly10descendants
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for…Hospira, Inc. · 2015 · cites it directly9descendants
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations on5descendants
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 3 generations on2descendants
  • K153074Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets, Kangaroo…Covidien · 2016 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K982159Abbott Plum a+ Infusion PumpAbbott Laboratories1999—
+1K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc.2011Class I
+1K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0…Hospira, Inc.2015Class I
+2K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc.2017Class I
+3K223607Plum Duo™ Infusion SystemIcu Medical, Inc.2023Class I
+2K153074Kangaroo Connect Enteral Feeding Pump with Kangaroo Connect Feeding Sets…Covidien20165

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
201111
201511
201610
201711
201911
202110
202210
202311
202520
202610
FIG. 1 — Lineage of K982159, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (12)

See all 12

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe Tubing Class IHospira, Inc.201110
K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for… Class IHospira, Inc.20159
K040114Abbott Data Port PC, Model 12181Abbott Laboratories20040

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 29, 2019Class IIZ-0126-2020Under Investigation by firmPotential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.
Feb 25, 2014Class IIZ-1479-2014Mixed-up of materials/componentsHospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood…
Sep 9, 2013Class IIZ-2059-2016Device DesignIntravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.
Sep 9, 2013Class IIZ-2062-2016Device DesignIntravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.
Feb 1, 2013Class IIZ-0285-2014Component design/selectionPlum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and…
Dec 7, 2012Class IIZ-0624-2013Device DesignThe door roller assembly on the Plum A+/A+3 infuser family of infusers has the potential to break which can lead to possible unrestricted flow and/or over-delivery during the removal of the IV administration set's cassette from the pump.
Oct 26, 2012Class IIZ-0481-2013Device DesignThe volume control knob on some Plum A+ single channel infusers (located on the back of the infuser) may not function as described in the System Operating Manual - the direction for Loud and Quiet may be reversed. The manual indicates a…
Feb 5, 2012Class IIZ-0205-2014Nonconforming Material/ComponentHospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186…
Sep 16, 2011Not classifiedZ-0009-2012Software designHospira has received reports of continuous recycling and/or rebooting of Plum A+ infusion pumps when the backlight intensity and/or display contrast settings for the LED display has been adjusted from the original default setting. This…
Sep 16, 2011Not classifiedZ-3284-2011Nonconforming Material/ComponentHospira has received reports of incorrect seating of the regulator closer on the Plum A+ Infusion Pumps. If the regulator closer does not close the flow regulator actuator when the pump door is opened, and if the clinician has not engaged…
Feb 14, 2011Not classifiedZ-1644-2011Nonconforming Material/ComponentHospira has received customer reports of the Plum A+ infusion pumps with no audible alarm conditions, which have been associated with failure of the piezoelectric assembly due to component quality issues.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K103224Hospira Plum Infusion Set; Hospira Infusion Set with Yellow Stripe TubingHospira, Inc. · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K141789Plum 360 Infusion System with Hospira Mednet, Smart Card Plug N Play CE 3.0 Module for Plum a+ Infusion…Hospira, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K161469Plum 360(TM) Infusion System with MedNet/ Smart Card Plug and Play ModuleHospira, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K223607Plum Duo™ Infusion SystemIcu Medical, Inc. · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K183311Ivenix Infusion System (IIS)Ivenix, Inc. · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 113 Class I recalls

What does this lineage say?

  • Of its descendants, K103224 (Hospira, Inc., 2011) has gone furthest, with 10 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Plum A+ Infusion Pump is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982159: Abbott Plum a+ Infusion Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K982159

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026