510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Genesis Imrt Linear Accelerator

K982502510(k) SpecialCleared Jul 31, 199811-day reviewClass II21 CFR 892.5050Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYE Accelerator, Linear, Medical

Cleared on Jul 31, 1998 as substantially equivalent to K882729 (Siemens Medical Laboratories, Inc., 1988) after an 11-day review. Its family tree reaches back 1 generation to 1988. 28 later devices descend from it. It has been recalled 4 times.

Decided
Jul 31, 1998
Review
11 days
Class
Class II
Product code
IYE
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K882729 (1988). 28 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2002: 1 descendant cleared2003: 1 descendant cleared2004: 1 descendant cleared2006: 1 descendant cleared2009: 3 descendants cleared2008: 1 descendant cleared2011: 3 descendants cleared2010: 1 descendant cleared2013: 1 descendant cleared2014: 2 descendants cleared2012: 2 descendants cleared2016: 2 descendants cleared2015: 1 descendant cleared2020: 2 descendants cleared2017: 2 descendants cleared2018: 2 descendants cleared2021: 1 descendant cleared2023: 1 descendant clearedK882729 · Mevatron, M 2 Class · Siemens Medical Laboratories, Inc. · 1988 · recalled1990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 2 descendants6 generations forward: 5 descendants7 generations forward: 5 descendants8 generations forward: 6 descendants9 generations forward: 4 descendants10 generations forward: 2 descendants K882729 · Mevatron, M 2 Class · Siemens Medical Laboratories, Inc. · 1988 · recalled0+5+10
  1. Root · 1 generation backK882729 Mevatron, M 2 ClassSiemens Medical Laboratories, Inc. · 1988 · Recalled
  2. This deviceK982502 Genesis Imrt Linear AcceleratorSiemens Medical Solutions USA, Inc. · 1998
  3. Descendant · cites it directly · 27 of its ownK013673 HI-Art SystemTomo Therapy, Inc. · 2002
  4. Descendant · 2 generations on · 26 of its ownK033347 Modification to: HI-Art SystemTomotherapy Incorporated · 2003
  5. Descendant · 3 generations on · 25 of its ownK042739 HI-Art SystemTomotherapy Incorporated · 2004
  6. Descendant · 4 generations on · 24 of its ownK060912 Modification to: HI-Art SystemTomotherapy Incorporated · 2006 · Recalled
  7. Descendant · 5 generations on · 13 of its ownK082005 Modification to HI-Art Treatment SystemTomotherapy Incorporated · 2008 · Recalled

Showing 5 of 28 descendants. See all descendants

Descendants cleared per year

2002: 12003: 12004: 12006: 12008: 12009: 32010: 12011: 32012: 22013: 12014: 22015: 12016: 22017: 22018: 22020: 22021: 12023: 1
2002high 3 (2009) · low 0 (2005)2026

Most influential descendants

Showing 5 of 28 · See all
  • K013673HI-Art SystemTomo Therapy, Inc. · 2002 · cites it directly27descendants
  • K033347Modification to: HI-Art SystemTomotherapy Incorporated · 2003 · 2 generations on26descendants
  • K042739HI-Art SystemTomotherapy Incorporated · 2004 · 3 generations on25descendants
  • K060912Modification to: HI-Art SystemTomotherapy Incorporated · 2006 · 4 generations on24descendants
  • K082005Modification to HI-Art Treatment SystemTomotherapy Incorporated · 2008 · 5 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K882729Mevatron, M 2 ClassSiemens Medical Laboratories, Inc.198838
0 (this device)K982502Genesis Imrt Linear AcceleratorSiemens Medical Solutions USA, Inc.1998—
+1K013673HI-Art SystemTomo Therapy, Inc.2002—
+2K033347Modification to: HI-Art SystemTomotherapy Incorporated2003—
+3K042739HI-Art SystemTomotherapy Incorporated2004—
+4K060912Modification to: HI-Art SystemTomotherapy Incorporated20064
+5K082005Modification to HI-Art Treatment SystemTomotherapy Incorporated20087

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200310
200410
200610
200810
200930
201010
201130
201220
201310
201420
201510
201620
201720
201820
202020
202110
202310
FIG. 1 — Lineage of K982502, 1988–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K882729Mevatron, M 2 Class
Summary page 1
…sification Name: Medical Charged Particle Radiation Therapy System 21 CFR § 892.5710 Class I, Product Code: RA 90 IYE 3. Predicate Device: MEVATRON M Class, 510(k) No. K882729 4. Description of Device: The GENESIS™ Linear Accelerator is a single energy, high dose medical linear accelerator. The GENESIS™…
Siemens Medical Laboratories, Inc.198838

Also cleared against K882729: 2 other devices →

What descends from it? (28)

See all 28

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K013673HI-Art SystemTomo Therapy, Inc.200227

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 19, 2014Class IIZ-1059-2015Component design/selectionA hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke…
May 28, 2014Class IIZ-2174-2014Use errorA safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to lead to a collision…
May 29, 2013Class IIZ-1663-2013Process change controlThe information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.
Nov 4, 2010Not classifiedZ-2639-2011OtherThe firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K013673 (Tomo Therapy, Inc., 2002) has gone furthest, with 27 descendants of its own.
  • 2 other devices also cite its predicate K882729.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The GENESIS IMRT Linear Accelerator is intended to deliver megavoltage

x-ray treatment for therapeutic application in the treatment of cancer. The

GENESIS delivers these treatments conformally, shaping the field of

radiation to the tumor volume. The depth of penetration emulating from the

GENESIS ideally suits itself for treating deep-seated tumors while providing

a skin sparing effect.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982502: Genesis Imrt Linear Accelerator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K982502

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026