510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Primus

K993425510(k) SpecialCleared Nov 26, 199945-day reviewClass II21 CFR 892.5050Recalls 50

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYE Accelerator, Linear, Medical

Cleared on Nov 26, 1999 as substantially equivalent to K882729 (Siemens Medical Laboratories, Inc., 1988) after a 45-day review. Its family tree reaches back 1 generation to 1988. 17 later devices descend from it. It has been recalled 50 times.

Decided
Nov 26, 1999
Review
45 days
Class
Class II
Product code
IYE
Recalls
50
Predicates
1

Its longest line of descent runs back 1 generation to K882729 (1988). 17 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2006: 2 descendants cleared2003: 1 descendant cleared2009: 4 descendants cleared2005: 1 descendant cleared2012: 1 descendant cleared2010: 2 descendants cleared2007: 1 descendant cleared2013: 1 descendant cleared2011: 1 descendant cleared2014: 2 descendants cleared2018: 1 descendant clearedK882729 · Mevatron, M 2 Class · Siemens Medical Laboratories, Inc. · 1988 · recalled1990200020102020
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 9 descendants3 generations forward: 3 descendants4 generations forward: 1 descendant K882729 · Mevatron, M 2 Class · Siemens Medical Laboratories, Inc. · 1988 · recalled0
  1. Root · 1 generation backK882729 Mevatron, M 2 ClassSiemens Medical Laboratories, Inc. · 1988 · Recalled
  2. This deviceK993425 PrimusSiemens Medical Solutions USA, Inc. · 1999
  3. Descendant · cites it directly · 16 of its ownK031764 Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  4. Descendant · cites it directly · 13 of its ownK060226 Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  5. Descendant · 2 generations on · 11 of its ownK072485 Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · Recalled
  6. Descendant · cites it directly · 9 of its ownK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  7. Descendant · 2 generations on · 7 of its ownK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · Recalled

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2003: 12005: 12006: 22007: 12009: 42010: 22011: 12012: 12013: 12014: 22018: 1
2003high 4 (2009) · low 0 (2004)2026

Most influential descendants

Showing 5 of 17 · See all
  • K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc. · 2003 · cites it directly16descendants
  • K060226Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006 · cites it directly13descendants
  • K072485Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · 2 generations on11descendants
  • K082775Prescision OptionSiemens Medical Solutions USA, Inc. · 2009 · cites it directly9descendants
  • K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · 2 generations on7descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K882729Mevatron, M 2 ClassSiemens Medical Laboratories, Inc.198838
0 (this device)K993425PrimusSiemens Medical Solutions USA, Inc.1999—
+1K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc.200317
+1K060226Oncor ExpressionSiemens Medical Solutions USA, Inc.200672
+2K072485Artiste MV SASiemens Medical Solutions USA, Inc.200724
+1K082775Prescision OptionSiemens Medical Solutions USA, Inc.2009—
+2K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20094

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200510
200620
200710
200940
201020
201110
201210
201310
201420
201810
FIG. 1 — Lineage of K993425, 1988–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K882729Mevatron, M 2 Class
Summary page 1
…fication Name: Medical Charged Particle Radiation Therapy System 21 CFR § 892.5710 Class II, Product Code: RA 90 IYE 3. Predicate Device: MEVATRON™ M Class, 510(k) No. K882729 4. Description of Device: The PRIMUS™ is a dual energy, high dose medical linear accelerator. The PRIMUS™ includes dual dose rat…
Siemens Medical Laboratories, Inc.198838

Also cleared against K882729: 2 other devices →

What descends from it? (17)

See all 17

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc.200316
K060226Oncor ExpressionSiemens Medical Solutions USA, Inc.200613
K082775Prescision OptionSiemens Medical Solutions USA, Inc.20099
K052315Simtec IM-Maxx 2 OptionSiemens Medical Solutions USA, Inc.20054

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 18, 2020Class IIZ-1680-2020Use errorPotential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
Mar 20, 2019Class IIZ-1204-2019Use errorAdvisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field…
Feb 14, 2018Class IIZ-1397-2018Software designControl Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Feb 14, 2018Class IIZ-1402-2018Software designControl Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
Feb 14, 2017Class IIZ-1489-2017Software designSoftware update
Jan 25, 2017Class IIZ-1215-2017Device DesignSiemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
Jan 25, 2017Class IIZ-1432-2017Under Investigation by firmSiemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any…
Jul 26, 2016Class IIZ-2877-2016Software designSoftware patch installation to address several safety issues.
Aug 5, 2015Class IIZ-2812-2015Software design (manufacturing process)A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.
May 5, 2015Class IIZ-0168-2016Software designpotential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.
Dec 19, 2014Class IIZ-1059-2015Component design/selectionA hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke…
Dec 2, 2014Class IIZ-0980-2015Component design/selectionUse of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in the overall dose…
Jul 18, 2013Class IIZ-2095-2013Device DesignSiemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rails in position on the tabletop might break off.
May 29, 2013Class IIZ-1663-2013Process change controlThe information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.
Apr 4, 2013Class IIZ-1182-2013Software designSoftware update to fix multiple safety related issues.
Apr 4, 2012Not classifiedZ-1494-2012Process controlThe firm became aware of a potential issue with Siemens-branded ONCOR or PRIMUS linear accelerator with an OPTIVUE or MVision imaging system that has the potential collision problem of the imaging panel if the panel does not retract…
Jan 22, 2012Not classifiedZ-1092-2012Software designSoftware upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.
Dec 1, 2011Not classifiedZ-0782-2012Nonconforming Material/ComponentBolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover.
Jun 3, 2011Not classifiedZ-3003-2011Software designA potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image.
May 25, 2011Not classifiedZ-2789-2011Software designPotential safety issue of possible data loss for the transfer of the same patients treatment data between the LANTIS OIS and the RTT v2.x software and the RT Therapist version 4.1 system
Mar 31, 2011Not classifiedZ-2243-2011Software designSiemens found a potential safety issue related to patient treatment with table offsets on syngo RT Therapist (RTT) 4.1 and 4.2 which could result in dose being delivered to wrong location.
Mar 31, 2011Not classifiedZ-2216-2011Software designSiemens has become aware of a potential safety issue that may result in mistreatment when customers are configured with the ARTISTE Linac RT Therapist v4.2 in combination with RT Oncologist v.4.2 and use the remote "Online 2D Image Review.
Mar 24, 2011Not classifiedZ-2190-2011PendingWhen using the ARTISTE Linac and RT Therapist v4.1 along with an imported cropped reference image, the image maybe processed incorrectly, thus resulting in a mistreatment.
Nov 4, 2010Not classifiedZ-2643-2011OtherThe firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
Nov 4, 2010Not classifiedZ-2644-2011OtherThe firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
Nov 4, 2010Not classifiedZ-2640-2011OtherThe firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
Nov 4, 2010Not classifiedZ-2641-2011OtherThe firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
Aug 21, 2009Not classifiedZ-1970-2011Software designWhen editing the note type of a previously entered note, then reverting to the original type, data may be lost.
Jan 11, 2008Not classifiedZ-1396-2008Software designLinear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
Jan 11, 2008Not classifiedZ-1397-2008Software designLinear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
Jan 11, 2008Not classifiedZ-1398-2008Software designLinear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
Dec 10, 2007Not classifiedZ-1759-2008Device DesignRadiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
Aug 7, 2007Not classifiedZ-0687-2008Device DesignIncorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial…
Aug 7, 2007Not classifiedZ-0686-2008Device DesignIncorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial…
Jul 27, 2007Not classifiedZ-0850-2008Component design/selectionDoor Hinge failures; loose or falling off (Stationary Structure Doors)
Mar 23, 2007Not classifiedZ-0380-2008Device DesignProduct's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
Mar 22, 2007Not classifiedZ-0166-2008Software designMistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of…
Mar 22, 2007Not classifiedZ-0169-2008Software designMistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of…
Mar 22, 2007Not classifiedZ-0167-2008Software designMistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of…
Mar 22, 2007Not classifiedZ-0168-2008Software designMistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of…
Jan 8, 2003Not classifiedZ-0355-06OtherIncomplete installation of the optional Beam Shield may allow unexpected doses of radiation outside the radiation protected areas.
Mar 5, 13Class IIZ-1040-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1041-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1043-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1036-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1039-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1044-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1042-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1038-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…
Mar 5, 13Class IIZ-1037-2013Software design (manufacturing process)Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K031764 (Siemens Medical Solutions USA, Inc., 2003) has gone furthest, with 16 descendants of its own.
  • 2 other devices also cite its predicate K882729.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The intended use of the PRIMUS™ is to deliver x-ray radiation for therapeutic treatment of cancer. The intended use of the PRIMUS™ is the same as the MEVATRON™ M and has not changed as a result of the modification.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993425: Primus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K993425

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026