NPT7 Series PH/Blood Gas and CO-Oximetry System

K982928510(k) TraditionalCleared Jan 6, 1999139-day reviewClass II21 CFR 862.1120Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer America, Inc. · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Jan 6, 1999 after a 139-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 13 later devices descend from it. It has been recalled 2 times.

Decided
Jan 6, 1999
Review
139 days
Class
Class II
Product code
CHL
Recalls
2
Predicates
0

13 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared2012: 3 descendants cleared2013: 3 descendants cleared2025: 1 descendant cleared2021: 1 descendant cleared2011: 1 descendant cleared2016: 1 descendant cleared2024: 1 descendant cleared2026: 1 descendant cleared1995200020052010201520202025
  1. This deviceK982928 NPT7 Series PH/Blood Gas and CO-Oximetry SystemRadiometer America, Inc. · 1999
  2. Descendant · cites it directly · 12 of its ownK092686 ABL90 Flex SeriesRadiometer Medical Aps · 2010 · Recalled
  3. Descendant · 2 generations on · 6 of its ownK120197 ABL90 FlexRadiometer Medical Aps · 2012 · Recalled
  4. Descendant · 3 generations on · 5 of its ownK122729 ABL90 FlexRadiometer Medical Aps · 2012
  5. Descendant · 4 generations on · 3 of its ownK131988 ABL90 Flex AnalyzerRadiometer Medical Aps · 2013
  6. Descendant · 5 generations on · 2 of its ownK160153 ABL90 Flex PlusRadiometer Medical Aps · 2016 · Recalled

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2010: 12011: 12012: 32013: 32016: 12021: 12024: 12025: 12026: 1
2010high 3 (2012) · low 0 (2014)2026

Most influential descendants

Showing 5 of 13 · See all
  • K092686ABL90 Flex SeriesRadiometer Medical Aps · 2010 · cites it directly12descendants
  • K120197ABL90 FlexRadiometer Medical Aps · 2012 · 2 generations on6descendants
  • K122729ABL90 FlexRadiometer Medical Aps · 2012 · 3 generations on5descendants
  • K131988ABL90 Flex AnalyzerRadiometer Medical Aps · 2013 · 4 generations on3descendants
  • K160153ABL90 Flex PlusRadiometer Medical Aps · 2016 · 5 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K982928NPT7 Series PH/Blood Gas and CO-Oximetry SystemRadiometer America, Inc.1999—
+1K092686ABL90 Flex SeriesRadiometer Medical Aps201010
+2K120197ABL90 FlexRadiometer Medical Aps20123
+3K122729ABL90 FlexRadiometer Medical Aps2012—
+4K131988ABL90 Flex AnalyzerRadiometer Medical Aps2013—
+5K160153ABL90 Flex PlusRadiometer Medical Aps20165

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201110
201230
201330
201610
202110
202410
202510
202610
FIG. 1 — Lineage of K982928, 1999–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.1 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (13)

See all 13

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K092686ABL90 Flex SeriesRadiometer Medical Aps201012

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2006Not classifiedZ-1261-06OtherAir bubbles may be trapped on or close to the O2 sensor in the pO2 chamber of the cuvettes in cartridges used for blood gas analysis on NPT7 analyzer. This produces too high pO2 readings without a question mark.
Jan 12, 2005Not classifiedZ-1406-06OtherQC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K092686 (Radiometer Medical Aps, 2010) has gone furthest, with 12 descendants of its own.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

NPT7 Series is a stand-alone blood gas analyzer that measures pH, pCO, pO, ,and Co-oximetry parameters on human arterial/venous and capillary whole blood. The NPT7 Near Patient Testing System in its various configurations is intended for use in laboratory and point-of-care testing of pH, blood gases and co-oximetry parameters on whole blood samples by minimally trained personnel. The device is designed for stationary use as well as use in moving vehicles and air crafts.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982928: NPT7 Series PH/Blood Gas and CO-Oximetry System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K982928

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026