510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite System

K980130510(k) TraditionalCleared Feb 9, 199826-day reviewClass II21 CFR 862.1120Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer America, Inc. · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Feb 9, 1998 as substantially equivalent to K962334 (Radiometer America, Inc., 1996) and 1 other predicate after a 26-day review. Its family tree reaches back 2 generations to 1995. 44 later devices descend from it. It has been recalled 5 times.

Decided
Feb 9, 1998
Review
26 days
Class
Class II
Product code
CHL
Recalls
5
Predicates
2

Its longest line of descent runs back 2 generations to K952621 (1995). 4 ancestors in all, the oldest K914166 (1992). 44 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1992: 1 other ancestor2005: 4 descendants cleared1999: 1 descendant cleared2008: 2 descendants cleared2000: 1 descendant cleared2015: 3 descendants cleared2004: 2 descendants cleared2010: 2 descendants cleared2006: 1 descendant cleared2020: 1 descendant cleared2011: 2 descendants cleared2012: 4 descendants cleared2016: 5 descendants cleared2014: 2 descendants cleared2017: 2 descendants cleared2013: 3 descendants cleared2025: 1 descendant cleared2021: 2 descendants cleared2019: 2 descendants cleared2023: 2 descendants cleared2024: 1 descendant cleared2026: 1 descendant clearedK952621 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 1995K962334 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 1996 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 6 descendants3 generations forward: 9 descendants4 generations forward: 12 descendants5 generations forward: 9 descendants6 generations forward: 4 descendants K962334 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 1996 · recalledK952621 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 19950+5
  1. Root · 2 generations backK952621 EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc. · 1995
  2. 1 generation backK962334 EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc. · 1996 · Recalled
  3. This deviceK980130 ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite SystemRadiometer America, Inc. · 1998
  4. Descendant · cites it directly · 31 of its ownK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. Descendant · 2 generations on · 29 of its ownK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  6. Descendant · 3 generations on · 18 of its ownK050869 Modification to ABL800 FlexRadiometer Medical Aps · 2005 · Recalled
  7. Descendant · cites it directly · 16 of its ownK991417 Abl 735 AnalyzerRadiometer America, Inc. · 1999 · Recalled
  8. Descendant · 3 generations on · 12 of its ownK092686 ABL90 Flex SeriesRadiometer Medical Aps · 2010 · Recalled

Showing 5 of 44 descendants. See all descendants

Descendants cleared per year

1999: 12000: 12004: 22005: 42006: 12008: 22010: 22011: 22012: 42013: 32014: 22015: 32016: 52017: 22019: 22020: 12021: 22023: 22024: 12025: 12026: 1
1999high 5 (2016) · low 0 (2001)2026

Most influential descendants

Showing 5 of 44 · See all
  • K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · cites it directly31descendants
  • K041874ABL800 FlexRadiometer Medical Aps · 2004 · 2 generations on29descendants
  • K050869Modification to ABL800 FlexRadiometer Medical Aps · 2005 · 3 generations on18descendants
  • K991417Abl 735 AnalyzerRadiometer America, Inc. · 1999 · cites it directly16descendants
  • K092686ABL90 Flex SeriesRadiometer Medical Aps · 2010 · 3 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K952621EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc.1995—
−1K962334EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc.19961
0 (this device)K980130ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite SystemRadiometer America, Inc.1998—
+1K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
+2K041874ABL800 FlexRadiometer Medical Aps200422
+3K050869Modification to ABL800 FlexRadiometer Medical Aps20051
+1K991417Abl 735 AnalyzerRadiometer America, Inc.19992
+3K092686ABL90 Flex SeriesRadiometer Medical Aps201010

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200010
200420
200540
200610
200820
201020
201120
201240
201330
201420
201530
201650
201720
201920
202010
202120
202320
202410
202510
202610
FIG. 1 — Lineage of K980130, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.1 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K962334EML105 Electrolyte Metabolite Analyzer
Summary page 1
…nologically similar to the ABL625 and ABL555. SUBSTANTIAL EQUIVALENCE: ABL700 Series is substantially equivalent in features and characteristics to the current ABL625 (K962334) and ABL555 (K973367) marketed by Radiometer America Inc. WIALTY of ., W SEVICES T %, (7 DEPARTMENT OF HEALTH…
Radiometer America, Inc.19961
*K973367ABL555 Blood Gas, Electrolyte & Metabolite System
Summary page 1
…o the ABL625 and ABL555. SUBSTANTIAL EQUIVALENCE: ABL700 Series is substantially equivalent in features and characteristics to the current ABL625 (K962334) and ABL555 (K973367) marketed by Radiometer America Inc. WIALTY of ., W SEVICES T %, (7 DEPARTMENT OF HEALTH & HUMAN SERVICES Pub…
Radiometer America, Inc.1997—

What descends from it? (44)

See all 44

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.200031
K991417Abl 735 AnalyzerRadiometer America, Inc.199916
K051804ABL80 Flex AnalyzerSendx Medical, Inc.20052
K080370Abl 80 CO-OX SystemSendx Medical, Inc.20081

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 5, 2013Class IIZ-1075-2014Under Investigation by firmRADIOMETER has become aware that some D788 pCO2 membranes can cause biased measurement results on patient results as well as QC results.
Apr 7, 2011Not classifiedZ-2571-2011PendingBased on internal testing and user/customer complaint reports received, the firm determined that wrinkles may arise on the outer membrane of the calcium membrane units upon installation. Retention of fluids between the wrinkles can cause…
Sep 29, 2005Not classifiedZ-1386-06OtherUnder certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and…
Sep 19, 2005Not classifiedZ-1323-06OtherReference electrodes for the firm's ABL 700/800 Blood Gas Analyzers are oversized and will not fit into the instrument.
Feb 1, 2005Not classifiedZ-1314-06OtherThe CI membrane units generate too low values for status and sensitivity during calibration.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 3 years and passes through 1 company; a new generation was cleared every 1.3 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K992859 (Radiometer America, Inc., 2000) has gone furthest, with 31 descendants of its own.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

ABL700 Series is a stand-alone blood gas analyzer that measures pH, pCO,, pO, , cNa*, cK*, cCa**, cCL’, Glucose, Lactate, and Co-oximetry parameters on human arterial/venous and capillary whole blood, and pO, and pCO, on expired air.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K980130: ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K980130

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026