510(k) Report

510(k) · Cleared

Recalled · 2005

Abl 735 Analyzer

K991417510(k) TraditionalCleared Nov 24, 1999215-day reviewClass II21 CFR 862.1120Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer America, Inc. · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Nov 24, 1999 as substantially equivalent to K980130 (Radiometer America, Inc., 1998) and 3 other predicates after a 215-day review. Its family tree reaches back 3 generations to 1995. 16 later devices descend from it. It has been recalled 2 times.

Decided
Nov 24, 1999
Review
215 days
Class
Class II
Product code
CHL
Recalls
2
Predicates
5

Its longest line of descent runs back 3 generations to K952621 (1995). 8 ancestors in all, the oldest K790608 (1979). 16 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1992: 2 other ancestors1984: 1 other ancestor1979: 1 other ancestor1997: 1 other ancestor2008: 1 descendant cleared2004: 1 descendant cleared2010: 1 descendant cleared2011: 1 descendant cleared2012: 1 descendant cleared2016: 4 descendants cleared2014: 1 descendant cleared2021: 1 descendant cleared2017: 1 descendant cleared2019: 2 descendants cleared2023: 2 descendants clearedK952621 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 1995K962334 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 1996 · recalledK980130 · ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite System · Radiometer America, Inc. · 1998 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 7 descendants5 generations forward: 1 descendant K980130 · ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite System · Radiometer America, Inc. · 1998 · recalledK962334 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 1996 · recalledK952621 · EML105 Electrolyte Metabolite Analyzer · Radiometer America, Inc. · 19950+5
  1. Root · 3 generations backK952621 EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc. · 1995
  2. 2 generations backK962334 EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc. · 1996 · Recalled
  3. 1 generation backK980130 ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite SystemRadiometer America, Inc. · 1998 · Recalled
  4. This deviceK991417 Abl 735 AnalyzerRadiometer America, Inc. · 1999
  5. Descendant · cites it directly · 12 of its ownK093623 Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter)…Instrumentation Laboratory CO · 2010 · Recalled
  6. Descendant · 2 generations on · 11 of its ownK112995 Gem Premier 4000Instrumentation Laboratory CO · 2012
  7. Descendant · 3 generations on · 10 of its ownK133407 Gem Premier 4000Instrumentation Laboratory CO · 2014
  8. Descendant · 4 generations on · 2 of its ownK173403 GEM Premier 5000Instrumentation Laboratory CO · 2017
  9. Descendant · 2 generations on · 2 of its ownK160225 GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride…Instrumentation Laboratory CO · 2016

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2004: 12008: 12010: 12011: 12012: 12014: 12016: 42017: 12019: 22021: 12023: 2
2004high 4 (2016) · low 0 (2005)2026

Most influential descendants

Showing 5 of 16 · See all
  • K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2 (Addition…Instrumentation Laboratory CO · 2010 · cites it directly12descendants
  • K112995Gem Premier 4000Instrumentation Laboratory CO · 2012 · 2 generations on11descendants
  • K133407Gem Premier 4000Instrumentation Laboratory CO · 2014 · 3 generations on10descendants
  • K173403GEM Premier 5000Instrumentation Laboratory CO · 2017 · 4 generations on2descendants
  • K160225GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride, Ionized Calcium), GEM…Instrumentation Laboratory CO · 2016 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K952621EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc.1995—
−2K962334EML105 Electrolyte Metabolite AnalyzerRadiometer America, Inc.19961
−1K980130ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite SystemRadiometer America, Inc.19985
0 (this device)K991417Abl 735 AnalyzerRadiometer America, Inc.1999—
+1K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2…Instrumentation Laboratory CO20105
+2K112995Gem Premier 4000Instrumentation Laboratory CO2012—
+3K133407Gem Premier 4000Instrumentation Laboratory CO2014—
+4K173403GEM Premier 5000Instrumentation Laboratory CO2017—
+2K160225GEM Premier 5000 (Measured Parameters:Sodium, Potassium, Chloride, Ionized…Instrumentation Laboratory CO2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198410
199220
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200810
201010
201110
201210
201410
201640
201710
201920
202110
202320
FIG. 1 — Lineage of K991417, 1995–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.1 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K980130ABL700 Series Blood Gas , CO-Oximetry, Electrolyte & Metabolite System
Summary page 1
…BL735 Common Name: pH/Blood Gas/Co-oximetry/Electrolyte/Metabolyte Analyzer Classification Name: Blood Gas and Blood pH Test System PREDICATE DEVICE: Radiometer ABL700 (K980130), for neonate bilirubin testing the Unistat (Leica Inc., K922770) and the Advanced Bilirubin Stat Analyzer Model BR2 (Advanced In…
Radiometer America, Inc.19985
*K922770Unistat Bilirubinometer Models 10310 and 10311
Summary page 1
…e Analyzer Classification Name: Blood Gas and Blood pH Test System PREDICATE DEVICE: Radiometer ABL700 (K980130), for neonate bilirubin testing the Unistat (Leica Inc., K922770) and the Advanced Bilirubin Stat Analyzer Model BR2 (Advanced Instruments Inc., K790608) and for adult bilirubin testing the Vitr…
Leica, Inc.1992—
*K790608BL Bilirubinometer
Summary page 1
…diometer ABL700 (K980130), for neonate bilirubin testing the Unistat (Leica Inc., K922770) and the Advanced Bilirubin Stat Analyzer Model BR2 (Advanced Instruments Inc., K790608) and for adult bilirubin testing the Vitros (Johnson and Johnson Clinical Diagnostics, formerly Kodak Clinical Diagnostics, K3406…
Advanced Instruments, Inc.1979—
Kept off the family treeK340606Vitros
Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree
Summary page 1
…Model BR2 (Advanced Instruments Inc., K790608) and for adult bilirubin testing the Vitros (Johnson and Johnson Clinical Diagnostics, formerly Kodak Clinical Diagnostics, K340606; K840880). PRODUCT DESCRIPTION: ABL73S is an automated pH/Blood Gas/Co-oximetry/Electrolyte/Metabo- lyte Analyzer that is 44cm…
*K840880Ektachem Clinical Chem. Slides
Summary page 1
…(Advanced Instruments Inc., K790608) and for adult bilirubin testing the Vitros (Johnson and Johnson Clinical Diagnostics, formerly Kodak Clinical Diagnostics, K340606; K840880). PRODUCT DESCRIPTION: ABL73S is an automated pH/Blood Gas/Co-oximetry/Electrolyte/Metabo- lyte Analyzer that is 44cm X 70cm X…
Eastman Kodak Company19842

Also cleared against K980130: 3 other devices →

What descends from it? (16)

See all 16

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093623Gem Premier 4000 with Iqm (Addition of Total Bilirubin Parameter), Gem CVP 2 (Addition…Instrumentation Laboratory CO201012
K073537Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics20081
K040174Bilirubin Assay on the Omni S AnalyzerRoche Diagnostics Corp.20040

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 29, 2005Not classifiedZ-1386-06OtherUnder certain circumstances, the ABL 700 & 800 Series Blood Gas Analyzers may experience leakage current into the measuring system. Consequently, the analyzers intermittently provide incorrect (too low) result values for Calcium (Ca) and…
Jun 17, 2005Not classifiedZ-1317-06OtherSoftware defect. pO2 and pCO2 sample test results run on the firm's ABL700/800 Series Blood Gas Analyzers are not being properly flagged during the sample calibration phase.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 1 company; a new generation was cleared every 1.3 years on median.
  • 4 of the 8 ancestors have a recall of any class.
  • Of its descendants, K093623 (Instrumentation Laboratory CO, 2010) has gone furthest, with 12 descendants of its own.
  • 3 other devices also cite its predicate K980130.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The ABL735 is intended for in vitro testing of samples of whole blood for the parameters pH,p02,.pCO02, potassium, sodium, calcium, chloride, glucose, 1actate, total bilirubin, and coOoximetry parameters (total hemoglobin, oxygen saturation, and the hemoglobin fractions f02Hb, fCOHb fMetHb fHHD, and fHbLF)

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K991417: Abl 735 Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K991417

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026