Sterrad 100S Sterilizer

K991999510(k) TraditionalCleared Jun 15, 2000367-day reviewClass II21 CFR 880.6860Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advanced Sterilization Products · MLR Sterilizer, Chemical

Cleared on Jun 15, 2000 after a 367-day review. The FDA summary names its predicate without a number. 19 later devices descend from it. It has been recalled 6 times.

Decided
Jun 15, 2000
Review
367 days
Class
Class II
Product code
MLR
Recalls
6
Predicates
0

19 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2003: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared2024: 2 descendants cleared2016: 5 descendants cleared2018: 2 descendants cleared2017: 1 descendant cleared2021: 1 descendant cleared2022: 1 descendant cleared2019: 1 descendant cleared2025: 1 descendant cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK991999 Sterrad 100S SterilizerAdvanced Sterilization Products · 2000
  2. Descendant · cites it directly · 18 of its ownK023290 Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products · 2003 · Recalled
  3. Descendant · cites it directly · 17 of its ownK111377 Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products · 2012 · Recalled
  4. Descendant · 2 generations on · 9 of its ownK142454 STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · Recalled
  5. Descendant · 2 generations on · 5 of its ownK141163 Sterizone SterilizerTso3, Inc. · 2014
  6. Descendant · 2 generations on · 4 of its ownK160903 STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products · 2016

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2003: 12012: 12014: 12015: 12016: 52017: 12018: 22019: 12021: 12022: 12024: 22025: 12026: 1
2003high 5 (2016) · low 0 (2004)2026

Most influential descendants

Showing 5 of 19 · See all
  • K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products · 2003 · cites it directly18descendants
  • K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products · 2012 · cites it directly17descendants
  • K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products · 2015 · 2 generations on9descendants
  • K141163Sterizone SterilizerTso3, Inc. · 2014 · 2 generations on5descendants
  • K160903STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products · 2016 · 2 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K991999Sterrad 100S SterilizerAdvanced Sterilization Products2000—
+1K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products20039
+1K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products20121
+2K142454STERRAD NX Sterilizer, STERRAD 100NX SterilizerAdvanced Sterilization Products20152
+2K141163Sterizone SterilizerTso3, Inc.2014—
+2K160903STERRAD 100NX Sterilizer with ALLClearTM TechnologyAdvanced Sterilization Products2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
201210
201410
201510
201650
201710
201820
201910
202110
202210
202420
202510
202610
FIG. 1 — Lineage of K991999, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MLR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MLR device

  1. #1 K250802 STERRAD® 100NX Sterilizer with ALLClear™…score 14.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—STERRAD® 100 Sterilization System
Named without a submission number
—
*—STERRAD® 50 Sterilization System
Named without a submission number
—

What descends from it? (19)

See all 19

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K023290Sterrad 50 & Sterrad 100S SterilizersAdvanced Sterilization Products200318
K111377Sterrad (R) 100NX Sterilizer Duo CycleAdvanced Sterilization Products201217

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 19, 2012Class IIIZ-2337-2012Error in labelingAdvanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and…
Oct 17, 2008Not classifiedZ-0340-2009Component design/selection1) Inability of the sterilizer to detect when an injection takes place without hydrogen peroxide being transferred to the vaporizer bowl. This situation can result in a cycle completion with the injection of insufficient hydrogen peroxide…
Feb 25, 2008Not classifiedZ-1628-2008Device DesignOil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."
Nov 12, 2007Not classifiedZ-0844-2008OtherOutdated Compatibility Lists: Recalling firm has decided to discontinue dissemination of Compatible Medical Device Reference Lists and instrument assessment activities
Nov 12, 2007Not classifiedZ-0845-2008Employee errorUser Guide Update: based on user reports about contact with residual hydrogen peroxide from instruments, pouches and trays after sterilization and subsequent light colored residue on these devices after sterilization updated user guidance…
Sep 16, 2005Not classifiedZ-0197-06OtherThe STERRAD 100 and the STERRAD 100S vaporizer plates are prone to displacement allowing micro-droplets of H2O2 to contact the load. If the vaporizer plate is dislodged, there is an increased chance for operators to have contact with H2O2…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K023290 (Advanced Sterilization Products, 2003) has gone furthest, with 18 descendants of its own.

How does its category compare?

All MLR devices

MLR Sterilizer, Chemical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MLR

1in 2025

2007: 22009: 12011: 62012: 22013: 12014: 22015: 12016: 42017: 22018: 22019: 42021: 22022: 22023: 52024: 32025: 12026 (partial year): 1
2007high 6 (2011) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MLR

216in 2024

2011: 214 days median2016: 183 days median2019: 90 days median2023: 222 days median2024: 216 days median
2007high 222 (2023) · low 90 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MLR

1in 2017

2007: 52008: 82009: 22010: 12011: 22012: 32013: 92014: 32017: 1
2007high 9 (2013) · low 0 (2015)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MLR

46in 2025

2007: 2722008: 5362009: 4332010: 2112011: 342012: 1012013: 5412014: 3992015: 4122016: 2862017: 2792018: 2892019: 2222020: 1742021: 2022022: 1882023: 1402024: 442025: 46
2007high 541 (2013) · low 34 (2011)2025

MAUDE reports naming product code MLR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—5272
20080—8536
20091—2433
20100—1211
20116214234
20122—3101
20131—9541
20142—3399
20151—0412
201641830286
20172—1279
20182—0289
20194900222
20200—0174
20212—0202
20222—0188
202352220140
20243216044
20251—046
20261—0—

What is it indicated for?

The STERRAD® 100S Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. (See list of recommended Materials in Section 3 of the Operator’s Manual.) The STERRAD® 100S Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the…

Read the full statement (152 words)

The STERRAD® 100S Sterilizer is designed for sterilization of both metal and nonmetal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is especially suitable for instruments sensitive to heat and moisture. (See list of recommended Materials in Section 3 of the Operator’s Manual.)

The STERRAD® 100S Sterilizer can sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.

Metal and nonmetal lumened instruments with inside diameters of 6 mm or larger and lengths of 310 mm or shorter can be processed in the STERRAD® 100S Sterilizer. (See list of recommended Materials in Section 3 of the Operator’s Manual)

Medical devices with only a single stainless steel lumen which has an inside diameter equal to or greater than 3 mm and a length less than or equal to 400 mm can be processed in the STERRAD® 100S Sterilizer.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K991999: Sterrad 100S Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K991999

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026