510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Iolmaster

K993357510(k) TraditionalCleared Mar 20, 2000166-day reviewClass II21 CFR 886.1850Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carl Zeiss, Inc. · HJO Biomicroscope, Slit-Lamp, Ac-Powered

Cleared on Mar 20, 2000 as substantially equivalent to K961171 (Carl Zeiss, Inc., 1996) and 3 other predicates after a 166-day review. Its family tree reaches back 1 generation to 1979. 17 later devices descend from it. It has been recalled 4 times.

Decided
Mar 20, 2000
Review
166 days
Class
Class II
Product code
HJO
Recalls
4
Predicates
4

Its longest line of descent runs back 1 generation to K781994 (1979). 4 ancestors in all. 17 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1993: 2 other ancestors2003: 1 descendant cleared2009: 1 descendant cleared2010: 1 descendant cleared2006: 1 descendant cleared2014: 1 descendant cleared2025: 2 descendants cleared2015: 2 descendants cleared2013: 1 descendant cleared2016: 1 descendant cleared2019: 2 descendants cleared2017: 1 descendant cleared2021: 1 descendant cleared2020: 1 descendant cleared2018: 1 descendant clearedK781994 · Autokeratometer, Instrument · Zeiss Humphrey System · 197919801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 6 descendants3 generations forward: 3 descendants4 generations forward: 4 descendants5 generations forward: 1 descendant K781994 · Autokeratometer, Instrument · Zeiss Humphrey System · 19790+5
  1. Root · 1 generation backK781994 Autokeratometer, InstrumentZeiss Humphrey System · 1979
  2. This deviceK993357 IolmasterCarl Zeiss, Inc. · 2000
  3. Descendant · cites it directly · 16 of its ownK030393 Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG · 2003
  4. Descendant · cites it directly · 14 of its ownK082891 Lenstar, Model LS900Haag-Streit AG · 2009 · Recalled
  5. Descendant · cites it directly · 10 of its ownK101182 Iolmaster 500Carl Zeiss Meditec, AG · 2010 · Recalled
  6. Descendant · 2 generations on · 9 of its ownK122418 Iolmaster 500Carl Zeiss Meditec, AG · 2013 · Recalled
  7. Descendant · 2 generations on · 4 of its ownK143275 IOLMaster700Carl Zeiss Meditec, Inc. · 2015

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2003: 12006: 12009: 12010: 12013: 12014: 12015: 22016: 12017: 12018: 12019: 22020: 12021: 12025: 2
2003high 2 (2015) · low 0 (2004)2026

Most influential descendants

Showing 5 of 17 · See all
  • K030393Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG · 2003 · cites it directly16descendants
  • K082891Lenstar, Model LS900Haag-Streit AG · 2009 · cites it directly14descendants
  • K101182Iolmaster 500Carl Zeiss Meditec, AG · 2010 · cites it directly10descendants
  • K122418Iolmaster 500Carl Zeiss Meditec, AG · 2013 · 2 generations on9descendants
  • K143275IOLMaster700Carl Zeiss Meditec, Inc. · 2015 · 2 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K781994Autokeratometer, InstrumentZeiss Humphrey System1979—
0 (this device)K993357IolmasterCarl Zeiss, Inc.2000—
+1K030393Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG2003—
+1K082891Lenstar, Model LS900Haag-Streit AG20093
+1K101182Iolmaster 500Carl Zeiss Meditec, AG20104
+2K122418Iolmaster 500Carl Zeiss Meditec, AG20134
+2K143275IOLMaster700Carl Zeiss Meditec, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199320
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200610
200910
201010
201310
201410
201520
201610
201710
201810
201920
202010
202110
202520
FIG. 1 — Lineage of K993357, 1979–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HJO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HJO device

  1. #1 K253834 Visuref 600score 13.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K961171Humphrey Oct Scanner
Summary page 1
…cation Name Keratoscope: § 886.1350 ' AC-Powered slitlamp biomicroscope: 886.1850 We are claiming substantial equivalence to the technology of the Humphrey Systems OCT (K961171) and the technology and indications of the Zeiss SL 120 slit lamp (K925641) and the indications of the Quantel Medical B-Scan ‘S’…
Carl Zeiss, Inc.1996—
*K925641Zeiss 20 SL Slit-Lamp & Mechanical Micromanipulatr
Summary page 1
…6.1850 We are claiming substantial equivalence to the technology of the Humphrey Systems OCT (K961171) and the technology and indications of the Zeiss SL 120 slit lamp (K925641) and the indications of the Quantel Medical B-Scan ‘S’ Model (K926251), for determining axial eye length measurement and anterior…
Carl Zeiss, Inc.1993—
*K926251B-Scan Opthalmic Ultrasound Imaging System
Summary page 1
…he Humphrey Systems OCT (K961171) and the technology and indications of the Zeiss SL 120 slit lamp (K925641) and the indications of the Quantel Medical B-Scan ‘S’ Model (K926251), for determining axial eye length measurement and anterior chamber depth measurement, and the Humphrey Auto Keratometer Model 42…
Biovision, Inc.1993—
*K781994Autokeratometer, Instrument
Summary page 1
…tel Medical B-Scan ‘S’ Model (K926251), for determining axial eye length measurement and anterior chamber depth measurement, and the Humphrey Auto Keratometer Model 420 (K781994) for corneal radius measurement, which already have pre-market clearance for the indications of providing measurements and genera…
Zeiss Humphrey System1979—

What descends from it? (17)

See all 17

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K030393Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG200316
K082891Lenstar, Model LS900Haag-Streit AG200914
K101182Iolmaster 500Carl Zeiss Meditec, AG201010

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 27, 2015Class IIZ-0357-2016Software designIOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Oct 27, 2015Class IIZ-0358-2016Software designIOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
Oct 23, 2013Class IIZ-0333-2014Software in the Use EnvironmentThe user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could…
Oct 23, 2013Class IIZ-0334-2014Software in the Use EnvironmentThe user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 4 ancestors has a recall.
  • Of its descendants, K030393 (Haag-Streit Holding AG, 2003) has gone furthest, with 16 descendants of its own.

How does its category compare?

All HJO devices

HJO Biomicroscope, Slit-Lamp, Ac-Powered: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HJO

1in 2024

2007: 12008: 42009: 32010: 52011: 22013: 22014: 102015: 22016: 52017: 92018: 22020: 12022: 12024: 12026 (partial year): 2
2007high 10 (2014) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HJO

232in 2017

2008: 117 days median2009: 91 days median2010: 121 days median2014: 196 days median2016: 243 days median2017: 232 days median
2007high 243 (2016) · low 91 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HJO

1in 2018

2013: 22015: 22018: 1
2007high 2 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HJO

20in 2025

2007: 12008: 32009: 32010: 22011: 52012: 72013: 162014: 42015: 82016: 152017: 152018: 112019: 132020: 192021: 142022: 62023: 182024: 182025: 20
2007high 20 (2025) · low 1 (2007)2025

MAUDE reports naming product code HJO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
2008411703
200939103
2010512102
20112—05
20120—07
20132—216
20141019604
20152—28
20165243015
20179232015
20182—111
20190—013
20201—019
20210—014
20221—06
20230—018
20241—018
20250—020
20262—0—

What is it indicated for?

This device will be used in the same manner as all ophthalmic diagnostic devices used to obtain ocular measurements (for axial length, anterior chamber depth and corneal radius), and perform calculations to allow physicians to determine appropriate IOL power and type for implantation.

This device is intended for use primarily by physicians and health care workers and may only be used under the supervision of a physician. This device will not be sold to the general public.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993357: Iolmaster.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K993357

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026