510(k) Report

510(k) · Cleared

Recalled · 2008

Non-Invasive Ventilation Niv Option, Model 84 14 474

K010093510(k) TraditionalCleared Jun 14, 2001154-day reviewClass II21 CFR 868.5895Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Drager Medizintechnik GmbH · CBK Ventilator, Continuous, Facility Use

Cleared on Jun 14, 2001 after a 154-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 30 of the 96 devices in its family, counting itself, have a Class I recall. 95 later devices descend from it. It has been recalled 3 times.

Decided
Jun 14, 2001
Review
154 days
Class
Class II
Product code
CBK
Recalls
3
Predicates
0

95 later devices descend from it, 6 citing it directly. 30 of the 96 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2005: 1 descendant cleared2004: 3 descendants cleared, 1 with a Class I recall12010: 4 descendants cleared, 1 with a Class I recall12008: 8 descendants cleared, 1 with a Class I recall12003: 1 descendant cleared2009: 4 descendants cleared, 1 with a Class I recall12006: 4 descendants cleared, 1 with a Class I recall12014: 3 descendants cleared, 1 with a Class I recall12007: 4 descendants cleared2011: 8 descendants cleared, 6 with a Class I recall62023: 5 descendants cleared, 1 with a Class I recall12025: 3 descendants cleared2012: 6 descendants cleared, 2 with a Class I recall22019: 5 descendants cleared, 1 with a Class I recall12015: 6 descendants cleared, 1 with a Class I recall12017: 6 descendants cleared, 3 with a Class I recall32013: 6 descendants cleared, 3 with a Class I recall32016: 2 descendants cleared, 2 with a Class I recall22021: 5 descendants cleared, 2 with a Class I recall22026: 3 descendants cleared2024: 3 descendants cleared2020: 4 descendants cleared, 3 with a Class I recall32018: 1 descendant cleared200020052010201520202025
  1. This deviceK010093 Non-Invasive Ventilation Niv Option, Model 84 14 474Drager Medizintechnik GmbH · 2001
  2. Descendant · cites it directly · 73 of its ownK041223 Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Recalled (Class I)
  3. Descendant · 2 generations on · 61 of its ownK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007
  4. Descendant · cites it directly · 53 of its ownK040574 Galileo GoldHamilton Medical, Inc. · 2004
  5. Descendant · cites it directly · 51 of its ownK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
  6. Descendant · 2 generations on · 37 of its ownK081842 Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008

Showing 5 of 95 descendants. See all descendants

Descendants cleared per year

2003: 12004: 32005: 12006: 42007: 42008: 82009: 42010: 42011: 82012: 62013: 62014: 32015: 62016: 22017: 62018: 12019: 52020: 42021: 52023: 52024: 32025: 32026: 3
2003high 8 (2008) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 95 · See all
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · cites it directly73descendants
  • K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007 · 2 generations on61descendants
  • K040574Galileo GoldHamilton Medical, Inc. · 2004 · cites it directly53descendants
  • K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · cites it directly51descendants
  • K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc. · 2008 · 2 generations on37descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010093Non-Invasive Ventilation Niv Option, Model 84 14 474Drager Medizintechnik GmbH2001—
+1K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB2004Class I
+2K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—
+1K040574Galileo GoldHamilton Medical, Inc.2004—
+1K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa20051
+2K081842Engstrom Carestation, Engstrom ProDatex-Ohmeda, Inc.2008—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200431
200510
200641
200740
200881
200941
201041
201186
201262
201363
201431
201561
201622
201763
201810
201951
202043
202152
202351
202430
202530
202630
FIG. 1 — Lineage of K010093, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

30 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (95)

See all 95

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407E Class IMaquet Critical Care AB200473
K040574Galileo GoldHamilton Medical, Inc.200453
K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa200551
K023289Option Niv, Model 8414115, Upgrade Kit Niv, Model 8414072Drager Medizintechnik GmbH200317
K093632Infinity Acute Care System Workstation Neonatal Care Class IDrager Medical AG & Co. Kgaa201010
K083050Evite XLDraeger Medical AG & Co. KG20082

Has anything in its family been recalled? (30 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 2, 2008Not classifiedZ-2325-2008Nonconforming Material/ComponentAudible power failure alarm was not enunciated at the required volume
Jul 2, 2008Not classifiedZ-2326-2008Nonconforming Material/Componentaudible power failure alarm was not enunciated at the required volume
May 7, 2003Not classifiedZ-0954-03Othersoftware problems - ventilation mode changes without user input, settings change without corresponding display change, false auto-set alarm.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K093632Infinity Acute Care System Workstation Neonatal CareDrager Medical AG & Co. Kgaa · 2010 · Cites it directly · ◆ in the 2010 descendants bar of FIG. 14 Class I recalls
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 2 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall
  • K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software OptionPuritan Bennett Corp. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K111116Engstrom Carestation, Engstrom Pro, Models 1505-9000-000, 1505-9003-000Datex-Ohmeda · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 13 Class I recalls
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K142679Carescape R860Datex-Ohmeda, Inc. · 2015 · 3 generations on · ◆ in the 2015 descendants bar of FIG. 13 Class I recalls
  • K131774Hamilton-G5Hamilton Medical AG · 2013 · 4 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 4 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls

Nearest 12 of 30 shown.

What does this lineage say?

  • Of its descendants, K041223 (Maquet Critical Care AB, 2004) has gone furthest, with 73 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—
Cite this page

510(k) Report. “K010093: Non-Invasive Ventilation Niv Option, Model 84 14 474.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K010093

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026