510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Modification to Hamilton-G5

K081521510(k) TraditionalCleared Sep 25, 2008118-day reviewClass II21 CFR 868.5895Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hamilton Medical AG · CBK Ventilator, Continuous, Facility Use

Cleared on Sep 25, 2008 as substantially equivalent to K070513 (Hamilton Medical AG, 2007) and 5 other predicates after a 118-day review. Its family tree reaches back 9 generations to 1991. 13 of the 71 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 2 times, including a Class I recall.

Decided
Sep 25, 2008
Review
118 days
Class
Class II
Product code
CBK
Recalls
2 (1 Class I)
Predicates
6

Its longest line of descent runs back 9 generations to K902859 (1991). 54 ancestors in all, the oldest K833982 (1984). 16 later devices descend from it, 2 citing it directly. 13 of the 71 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 3 other ancestors2004: 5 other ancestors, 1 with a Class I recall11998: 5 other ancestors2005: 1 other ancestor2002: 2 other ancestors2001: 5 other ancestors1999: 4 other ancestors1997: 3 other ancestors2003: 2 other ancestors, 1 with a Class I recall12000: 2 other ancestors, 1 with a Class I recall11996: 2 other ancestors1994: 4 other ancestors1992: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestor1984: 2 other ancestors1995: 2 other ancestors2009: 1 descendant cleared2011: 1 descendant cleared2012: 5 descendants cleared, 2 with a Class I recall22015: 1 descendant cleared2017: 2 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared2026: 1 descendant clearedK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK022679 · Raphael · Hamilton Medical AG · 2003 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK061090 · Galileo Gold Asv · Hamilton Medical AG · 2006K070513 · Hamilton-G5 · Hamilton Medical AG · 2007 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors, 1 with a Class I recall12 generations back: 11 other ancestors3 generations back: 17 other ancestors, 1 with a Class I recall14 generations back: 8 other ancestors, 1 with a Class I recall15 generations back: 5 other ancestors 1 generation forward: 2 descendants2 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 4 descendants, 1 with a Class I recall14 generations forward: 4 descendants, 3 with a Class I recall35 generations forward: 3 descendants, 2 with a Class I recall26 generations forward: 1 descendant K070513 · Hamilton-G5 · Hamilton Medical AG · 2007 · recalledK061090 · Galileo Gold Asv · Hamilton Medical AG · 2006K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K022679 · Raphael · Hamilton Medical AG · 2003 · recalledK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK961687 · EVITA 4 · Drager, Inc. · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled−50+5
  1. Root · 9 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 5 generations between
  3. 3 generations backK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
  4. 2 generations backK061090 Galileo Gold AsvHamilton Medical AG · 2006
  5. 1 generation backK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled
  6. This deviceK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008
  7. Descendant · cites it directly · 14 of its ownK102775 Hamilton-C2Hamilton Medical AG · 2011
  8. Descendant · 2 generations on · 13 of its ownK112006 Hamilton-TC1Hamilton Medical AG · 2012
  9. Descendant · 3 generations on · 11 of its ownK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
  10. Descendant · 2 generations on · 8 of its ownK121225 Hamilton-C2Hamilton Medical AG · 2012 · Recalled (Class I)
  11. Descendant · 3 generations on · 4 of its ownK123637 Hamilton-C3Hamilton Medical AG · 2012
Show all 9 generations
  1. Root · 9 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 8 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 7 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 6 generations backK984535 Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Recalled (Class I)
  5. 5 generations backK022679 RaphaelHamilton Medical AG · 2003 · Recalled
  6. 4 generations backK040574 Galileo GoldHamilton Medical, Inc. · 2004
  7. 3 generations backK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
  8. 2 generations backK061090 Galileo Gold AsvHamilton Medical AG · 2006
  9. 1 generation backK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled
  10. This deviceK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2009: 12011: 12012: 52015: 12016: 12017: 22019: 12020: 12021: 12024: 12026: 1
2009high 5 (2012) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 16 · See all
  • K102775Hamilton-C2Hamilton Medical AG · 2011 · cites it directly14descendants
  • K112006Hamilton-TC1Hamilton Medical AG · 2012 · 2 generations on13descendants
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · 3 generations on11descendants
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 2 generations on8descendants
  • K123637Hamilton-C3Hamilton Medical AG · 2012 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K902859Servo Ventilator 300Siemens Life Support Systems19913
−8K961687EVITA 4Drager, Inc.1996—
−7K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−6K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp.1998Class I
−5K022679RaphaelHamilton Medical AG20033
−4K040574Galileo GoldHamilton Medical, Inc.2004—
−3K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa20051
−2K061090Galileo Gold AsvHamilton Medical AG2006—
−1K070513Hamilton-G5Hamilton Medical AG20071
0 (this device)K081521Modification to Hamilton-G5Hamilton Medical AG2008—
+1K102775Hamilton-C2Hamilton Medical AG2011—
+2K112006Hamilton-TC1Hamilton Medical AG2012—
+3K120670Hamilton-T1Hamilton Medical AG2012Class I
+2K121225Hamilton-C2Hamilton Medical AG2012Class I
+3K123637Hamilton-C3Hamilton Medical AG2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198420
198910
199110
199210
199440
199520
199620
199730
199850
199940
200021
200150
200220
200321
200451
200510
200630

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201110
201252
201510
201611
201721
201911
202011
202111
202410
202610
FIG. 1 — Lineage of K081521, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

13 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070513Hamilton-G5
Summary page 1
…entilator, Continuous, Facility Use (21 CFR 868.5895, Product Code: CB3K) HAMILTON-G5 510(k) Number: K070513 MAQUET Servo-i LEGALLY MARKETED 510(k) Number: K041223 DEVICES TO WHICH Viasys AVEA ventilator…
Hamilton Medical AG20071
*K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407E
Summary page 1
…Code: CB3K) HAMILTON-G5 510(k) Number: K070513 MAQUET Servo-i LEGALLY MARKETED 510(k) Number: K041223 DEVICES TO WHICH Viasys AVEA ventilator EQUIVALENCE IS 510(k) Number: K062093 BEING CLAIM ED: Datex-Ohmeda E…
Maquet Critical Care AB2004Class I
*K062093Avea Ventilator
Summary page 1
…3 MAQUET Servo-i LEGALLY MARKETED 510(k) Number: K041223 DEVICES TO WHICH Viasys AVEA ventilator EQUIVALENCE IS 510(k) Number: K062093 BEING CLAIM ED: Datex-Ohmeda Engstromn Carestation 510(k) Number: K062710 Dr…
Viasys Respiratory Care, Inc.20065
*K062710GE Datex-Ohmeda Engstrom Carestation
Summary page 1
…Viasys AVEA ventilator EQUIVALENCE IS 510(k) Number: K062093 BEING CLAIM ED: Datex-Ohmeda Engstromn Carestation 510(k) Number: K062710 Drager EvitaXL with NeoFlow Option 510(k) Numbers: K051263, K983219…
Datex-Ohmeda, Inc.20062
*K051263Evitaxl with Option Smartcare
Summary page 1
…meda Engstromn Carestation 510(k) Number: K062710 Drager EvitaXL with NeoFlow Option 510(k) Numbers: K051263, K983219 Page 1 of 3…
Drager Medical AG & Co. Kgaa20051
*K983219Neoflow Option Model Numbers 8413563, 8411900
Summary page 1
…tromn Carestation 510(k) Number: K062710 Drager EvitaXL with NeoFlow Option 510(k) Numbers: K051263, K983219 Page 1 of 3 HAMILTON…
Drager, Inc.1998—

Also cleared against K070513: 2 other devices →

What descends from it? (16)

See all 16

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102775Hamilton-C2Hamilton Medical AG201114
K090468Hamilton-C2Hamilton Medical AG20090

Has anything in its family been recalled? (13 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 25, 2019Class IZ-1874-2019Software designNew software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When…
Feb 16, 2011Not classifiedZ-1928-2011Software designThe adaptive volume/pressure regulation can be automatically and permanently reduced to PEEP + 3 mbar when a closed suctioning maneuver is performed on the patient and may be undetected by the operator.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · Direct predicate · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 3 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K121225Hamilton-C2Hamilton Medical AG · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K120670Hamilton-T1Hamilton Medical AG · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 4 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K181216Hamilton-T1, Hamilton-C1Hamilton Medical AG · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 15 Class I recalls
  • K153046Hamilton-MR1Hamilton Medical AG · 2016 · 4 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K201306Hamilton-C3Hamilton Medical AG · 2021 · 5 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K201658Hamilton-C6Hamilton Medical AG · 2020 · 5 generations on · ◆ in the 2020 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 17 years and passes through 6 companies; a new generation was cleared every 1.0 years on median.
  • 23 of the 54 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K102775 (Hamilton Medical AG, 2011) has gone furthest, with 14 descendants of its own.
  • 2 other devices also cite its predicate K070513.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The HAMILTON-G5 ventilator is intended to provide positive pressure ventilatory support to adult and pediatric patients, and optionally to infant and neonatal patients.

The device is intended for use in the hospital and institutional environment where healthcare professionals provide patient care, including use as a patient bedside for intra-facility transport, provided compressed gas is supplied.

The device is not intended for transportation outside the hospital or for use in the home environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K081521: Modification to Hamilton-G5.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K081521

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026