510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

Hamilton-C2

K121225510(k) TraditionalCleared Nov 16, 2012207-day reviewClass II21 CFR 868.5895Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hamilton Medical AG · CBK Ventilator, Continuous, Facility Use

Cleared on Nov 16, 2012 as substantially equivalent to K102775 (Hamilton Medical AG, 2011) and 3 other predicates after a 207-day review. Its family tree reaches back 13 generations to 1991. 17 of the 123 devices in its family, counting itself, have a Class I recall. 8 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
Nov 16, 2012
Review
207 days
Class
Class II
Product code
CBK
Recalls
3 (1 Class I)
Predicates
4

Its longest line of descent runs back 13 generations to K902859 (1991). 114 ancestors in all, the oldest K791045 (1979). 8 later devices descend from it, 2 citing it directly. 17 of the 123 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2011: 1 other ancestor, 1 with a Class I recall12010: 3 other ancestors, 1 with a Class I recall12008: 4 other ancestors2004: 9 other ancestors, 1 with a Class I recall12001: 6 other ancestors1999: 9 other ancestors1998: 10 other ancestors, 1 with a Class I recall11994: 4 other ancestors1991: 6 other ancestors2007: 1 other ancestor2006: 6 other ancestors, 1 with a Class I recall12003: 4 other ancestors, 1 with a Class I recall12002: 4 other ancestors, 1 with a Class I recall11997: 4 other ancestors1995: 6 other ancestors1990: 6 other ancestors1979: 1 other ancestor2005: 3 other ancestors2000: 4 other ancestors, 1 with a Class I recall11996: 3 other ancestors1993: 1 other ancestor1992: 2 other ancestors1989: 1 other ancestor1988: 1 other ancestor1984: 2 other ancestors2012: 1 descendant cleared2015: 1 descendant cleared2017: 1 descendant cleared2019: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12026: 1 descendant clearedK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK022679 · Raphael · Hamilton Medical AG · 2003 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK061090 · Galileo Gold Asv · Hamilton Medical AG · 2006K070513 · Hamilton-G5 · Hamilton Medical AG · 2007 · recalledK081521 · Modification to Hamilton-G5 · Hamilton Medical AG · 2008 · Class I recallK102775 · Hamilton-C2 · Hamilton Medical AG · 2011×2: 2 generations cleared close together, K112006 (2012), K120670 (2012) 19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 2 with a Class I recall22 generations back: 14 other ancestors, 1 with a Class I recall13 generations back: 16 other ancestors, 1 with a Class I recall14 generations back: 23 other ancestors, 1 with a Class I recall15 generations back: 20 other ancestors6 generations back: 8 other ancestors, 1 with a Class I recall17 generations back: 8 other ancestors, 2 with a Class I recall28 generations back: 5 other ancestors9 generations back: 5 other ancestors 1 generation forward: 2 descendants2 generations forward: 3 descendants, 2 with a Class I recall23 generations forward: 2 descendants, 2 with a Class I recall24 generations forward: 1 descendant K120670 · Hamilton-T1 · Hamilton Medical AG · 2012 · Class I recallK112006 · Hamilton-TC1 · Hamilton Medical AG · 2012K102775 · Hamilton-C2 · Hamilton Medical AG · 2011K081521 · Modification to Hamilton-G5 · Hamilton Medical AG · 2008 · Class I recallK070513 · Hamilton-G5 · Hamilton Medical AG · 2007 · recalledK061090 · Galileo Gold Asv · Hamilton Medical AG · 2006K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K022679 · Raphael · Hamilton Medical AG · 2003 · recalledK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK961687 · EVITA 4 · Drager, Inc. · 1996K902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalled−10−50
  1. Root · 13 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 9 generations between
  3. 3 generations backK102775 Hamilton-C2Hamilton Medical AG · 2011
  4. 2 generations backK112006 Hamilton-TC1Hamilton Medical AG · 2012
  5. 1 generation backK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
  6. This deviceK121225 Hamilton-C2Hamilton Medical AG · 2012
  7. Descendant · cites it directly · 4 of its ownK123637 Hamilton-C3Hamilton Medical AG · 2012
  8. Descendant · 2 generations on · 3 of its ownK161450 Hamilton-C3Hamilton Medical AG · 2017 · Recalled
  9. Descendant · cites it directly · 2 of its ownK140939 Hamilton-T1,Hamilton-C1Hamilton Medical AG · 2015 · Recalled
  10. Descendant · 3 generations on · 1 of its ownK201658 Hamilton-C6Hamilton Medical AG · 2020 · Recalled (Class I)
  11. Descendant · 4 generations on · 0 of its ownK252663 Inbentus VersatileIconic Solutions by Murcia, SL · 2026
Show all 13 generations
  1. Root · 13 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 12 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 11 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 10 generations backK984535 Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Recalled (Class I)
  5. 9 generations backK022679 RaphaelHamilton Medical AG · 2003 · Recalled
  6. 8 generations backK040574 Galileo GoldHamilton Medical, Inc. · 2004
  7. 7 generations backK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
  8. 6 generations backK061090 Galileo Gold AsvHamilton Medical AG · 2006
  9. 5 generations backK070513 Hamilton-G5Hamilton Medical AG · 2007 · Recalled
  10. 4 generations backK081521 Modification to Hamilton-G5Hamilton Medical AG · 2008 · Recalled (Class I)
  11. 3 generations backK102775 Hamilton-C2Hamilton Medical AG · 2011
  12. 2 generations backK112006 Hamilton-TC1Hamilton Medical AG · 2012
  13. 1 generation backK120670 Hamilton-T1Hamilton Medical AG · 2012 · Recalled (Class I)
  14. This deviceK121225 Hamilton-C2Hamilton Medical AG · 2012

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2012: 12015: 12016: 12017: 12019: 12020: 12021: 12026: 1
2012high 1 (2012) · low 0 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K123637Hamilton-C3Hamilton Medical AG · 2012 · cites it directly4descendants
  • K161450Hamilton-C3Hamilton Medical AG · 2017 · 2 generations on3descendants
  • K140939Hamilton-T1,Hamilton-C1Hamilton Medical AG · 2015 · cites it directly2descendants
  • K201658Hamilton-C6Hamilton Medical AG · 2020 · 3 generations on1descendant
  • K252663Inbentus VersatileIconic Solutions by Murcia, SL · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K902859Servo Ventilator 300Siemens Life Support Systems19913
−12K961687EVITA 4Drager, Inc.1996—
−11K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−10K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp.1998Class I
−9K022679RaphaelHamilton Medical AG20033
−8K040574Galileo GoldHamilton Medical, Inc.2004—
−7K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa20051
−6K061090Galileo Gold AsvHamilton Medical AG2006—
−5K070513Hamilton-G5Hamilton Medical AG20071
−4K081521Modification to Hamilton-G5Hamilton Medical AG2008Class I
−3K102775Hamilton-C2Hamilton Medical AG2011—
−2K112006Hamilton-TC1Hamilton Medical AG2012—
−1K120670Hamilton-T1Hamilton Medical AG2012Class I
0 (this device)K121225Hamilton-C2Hamilton Medical AG2012—
+1K123637Hamilton-C3Hamilton Medical AG2012—
+2K161450Hamilton-C3Hamilton Medical AG20172
+1K140939Hamilton-T1,Hamilton-C1Hamilton Medical AG20151
+3K201658Hamilton-C6Hamilton Medical AG2020Class I
+4K252663Inbentus VersatileIconic Solutions by Murcia, SL2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198420
198810
198910
199060
199160
199220
199310
199440
199560
199630
199740
1998101
199990
200041
200160
200241
200341
200491
200530
200661
200710
200840
201031
201111

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201510
201611
201710
201911
202011
202111
202610
FIG. 1 — Lineage of K121225, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

17 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102775Hamilton-C2
Summary page 1
…Ventilator, Continuou s REGULATION NUMBER: 21 CFR 868.5895 PRODUCT CODE: CBI( PREDICATE DEVICE: ~ HAMILTON-C2 (K1 02775) (PRIMARY) PREDICATE DEVICES: HAMILTON-G5 (K1<103803), (SECONDARY) HAMILTON-Ti…
Hamilton Medical AG2011—
*K103803Hamilton-G5
Summary page 1
…PRODUCT CODE: CBI( PREDICATE DEVICE: ~ HAMILTON-C2 (K1 02775) (PRIMARY) PREDICATE DEVICES: HAMILTON-G5 (K1<103803), (SECONDARY) HAMILTON-Ti (K120670), Draeger's infinity Acute Care Syst…
Hamilton Medical AG2011Class I
*K120670Hamilton-T1
Summary page 1
…EVICE: ~ HAMILTON-C2 (K1 02775) (PRIMARY) PREDICATE DEVICES: HAMILTON-G5 (K1<103803), (SECONDARY) HAMILTON-Ti (K120670), Draeger's infinity Acute Care System Workstation - Neonat…
Hamilton Medical AG2012Class I
*K093632Infinity Acute Care System Workstation Neonatal Care
Summary page 1
…HAMILTON-Ti (K120670), Draeger's infinity Acute Care System Workstation - Neonatal Care (K093632). IbTPLIISYNC NEDI…
Drager Medical AG & Co. Kgaa2010Class I

Also cleared against K102775: 1 other device →

What descends from it? (8)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123637Hamilton-C3Hamilton Medical AG20124
K140939Hamilton-T1,Hamilton-C1Hamilton Medical AG20152

Has anything in its family been recalled? (17 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 15, 2023Class IZ-2141-2023Software designDue to a software anomaly after approximately 91 days of cumulative use, the ventilator may stop and enter the ambient mode without prior notice.
Mar 12, 2018Class IIZ-1371-2018OtherAfter two and a half years, the battery fuel gauge may indicate a higher battery charge than what is available.
Sep 26, 2013Class IIZ-0170-2014Device DesignManufacturer defined limits for improperly functioning loudspeakers of the HAMILTON C2 alarm system were exceeded.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K120670Hamilton-T1Hamilton Medical AG · 2012 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K103803Hamilton-G5Hamilton Medical AG · 2011 · Direct predicate · ◆ in the 2011 ancestors bar of FIG. 11 Class I recall
  • K093632Infinity Acute Care System Workstation Neonatal CareDrager Medical AG & Co. Kgaa · 2010 · Direct predicate · ◆ in the 2010 ancestors bar of FIG. 14 Class I recalls
  • K081521Modification to Hamilton-G5Hamilton Medical AG · 2008 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041223Modification to Servo-I Ventilator System, Model 64 87 800 E407EMaquet Critical Care AB · 2004 · 2 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K062267Oxylog 3000Drager Medical B.V. · 2006 · 3 generations back · ◆ in the 2006 ancestors bar of FIG. 13 Class I recalls
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 6 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013642AVEABird Products Corp. · 2002 · 7 generations back · ◆ in the 2002 ancestors bar of FIG. 113 Class I recalls
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 7 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

Nearest 12 of 16 shown.

What does this lineage say?

  • The line of descent spans 21 years and passes through 6 companies; a new generation was cleared every 1.0 years on median.
  • 35 of the 114 ancestors have a recall of any class; 12 have a Class I recall.
  • Of its descendants, K123637 (Hamilton Medical AG, 2012) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K102775.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The HAMILTON-C2 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates.

Intended areas of use:

- In the intensive care ward or in the recovery room

- During transfer of ventilated patients within the hospital

The HAMILTON-C2 ventilator is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121225: Hamilton-C2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121225

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026