510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Intera I/T

K013344510(k) TraditionalCleared Dec 11, 200163-day reviewClass II21 CFR 892.1000Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems North America Co. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Dec 11, 2001 as substantially equivalent to K001796 (Philips Medical Systems North America, Inc., 2000) after a 63-day review. Its family tree reaches back 1 generation to 2000. 1 of the 56 devices in its family, counting itself, has a Class I recall. 54 later devices descend from it. It has been recalled 2 times.

Decided
Dec 11, 2001
Review
63 days
Class
Class II
Product code
LNH
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K001796 (2000). 54 later devices descend from it, 1 citing it directly. 1 of the 56 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2007: 1 descendant cleared, 1 with a Class I recall12019: 2 descendants cleared2017: 3 descendants cleared2010: 2 descendants cleared2011: 3 descendants cleared2018: 9 descendants cleared2015: 4 descendants cleared2016: 3 descendants cleared2013: 1 descendant cleared2012: 1 descendant cleared2020: 3 descendants cleared2026: 2 descendants cleared2021: 4 descendants cleared2022: 8 descendants cleared2025: 3 descendants cleared2023: 5 descendants clearedK001796 · Gyroscan Intera (R7.5) · Philips Medical Systems North America, Inc. · 2000 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 7 descendants3 generations forward: 11 descendants4 generations forward: 9 descendants5 generations forward: 11 descendants6 generations forward: 5 descendants7 generations forward: 2 descendants8 generations forward: 3 descendants9 generations forward: 3 descendants10 generations forward: 2 descendants K001796 · Gyroscan Intera (R7.5) · Philips Medical Systems North America, Inc. · 2000 · recalled0+5+10
  1. Root · 1 generation backK001796 Gyroscan Intera (R7.5)Philips Medical Systems North America, Inc. · 2000 · Recalled
  2. This deviceK013344 Intera I/TPhilips Medical Systems North America Co. · 2001
  3. Descendant · cites it directly · 53 of its ownK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
  4. Descendant · 2 generations on · 34 of its ownK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
  5. Descendant · 3 generations on · 19 of its ownK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
  6. Descendant · 4 generations on · 15 of its ownK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled
  7. Descendant · 5 generations on · 13 of its ownK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled

Showing 5 of 54 descendants. See all descendants

Descendants cleared per year

2007: 12010: 22011: 32012: 12013: 12015: 42016: 32017: 32018: 92019: 22020: 32021: 42022: 82023: 52025: 32026: 2
2007high 9 (2018) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 54 · See all
  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · cites it directly53descendants
  • K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · 2 generations on34descendants
  • K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · 3 generations on19descendants
  • K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · 4 generations on15descendants
  • K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · 5 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K001796Gyroscan Intera (R7.5)Philips Medical Systems North America, Inc.20004
0 (this device)K013344Intera I/TPhilips Medical Systems North America Co.2001—
+1K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems2007Class I
+2K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care201120
+3K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V.20177
+4K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V.201811
+5K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V.201925

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200711
201020
201130
201210
201310
201540
201630
201730
201890
201920
202030
202140
202280
202350
202530
202620
FIG. 1 — Lineage of K013344, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001796Gyroscan Intera (R7.5)
Summary page 1
…fety standards and/or draft standards are used. Common/Usual Name : INTERA I/T. Predicate Device(s). The Philips cleared MRDD Philips INTERA 1.5T system with FDA ref. K001796. Intended Use. INTERA I/T is a whole body 1.5T Magnetic Resonance Diagnostic Device extended with optional hardware extensions…
Philips Medical Systems North America, Inc.20004

Also cleared against K001796: 2 other devices →

What descends from it? (54)

See all 54

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063559Achieva, Intera and Panorama 1.0T, Release 2.5 Class IPhilips Medical Systems200753

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 21, 2018Class IIZ-0885-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Jul 6, 2015Class IIZ-1024-2016Device DesignWrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K063559 (Philips Medical Systems, 2007) has gone furthest, with 53 descendants of its own.
  • 2 other devices also cite its predicate K001796.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

INTERA I/T is a whole body 1.5T Magnetic Resonance Diagnostic Device being extended with optional hardware extensions to aid in the performance of interventional procedures in the head, body and extremities, which may be facilitated by MR techniques, such as real time imaging.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013344: Intera I/T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K013344

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026