510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4

K173079510(k) AbbreviatedCleared Apr 4, 2018187-day reviewClass II21 CFR 892.1000Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Apr 4, 2018 as substantially equivalent to K163116 (Philips Medical Systems Nederland B.V., 2017) after a 187-day review. Its family tree reaches back 6 generations to 2003. 4 of the 25 devices in its family, counting itself, have a Class I recall. 15 later devices descend from it. It has been recalled 11 times.

Decided
Apr 4, 2018
Review
187 days
Class
Class II
Product code
LNH
Recalls
11
Predicates
1

Its longest line of descent runs back 6 generations to K031815 (2003). 9 ancestors in all, the oldest K001796 (2000). 15 later devices descend from it, 3 citing it directly. 4 of the 25 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor, 1 with a Class I recall12001: 1 other ancestor2000: 1 other ancestor2018: 1 descendant cleared2020: 1 descendant cleared2019: 2 descendants cleared2022: 4 descendants cleared2021: 1 descendant cleared2025: 2 descendants cleared2023: 3 descendants cleared2026: 1 descendant clearedK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalledK041602 · Panorama 1.0T · Philips Medical Systems North America, Inc. · 2004 · Class I recallK052078 · Achieva, Intera and Panorama 1.0 Release 2-Series · Philips Medical Systems Nederland B.V. · 2005 · Class I recallK063559 · Achieva, Intera and Panorama 1.0T, Release 2.5 · Philips Medical Systems · 2007 · Class I recallK110151 · Achieva R4 1.5T and Achieva R4 3.0T · Philips Health Care · 2011 · recalledK163116 · Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 · Philips Medical Systems Nederland B.V. · 2017 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor, 1 with a Class I recall14 generations back: 1 other ancestor5 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 7 descendants3 generations forward: 3 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant K163116 · Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 · Philips Medical Systems Nederland B.V. · 2017 · recalledK110151 · Achieva R4 1.5T and Achieva R4 3.0T · Philips Health Care · 2011 · recalledK063559 · Achieva, Intera and Panorama 1.0T, Release 2.5 · Philips Medical Systems · 2007 · Class I recallK052078 · Achieva, Intera and Panorama 1.0 Release 2-Series · Philips Medical Systems Nederland B.V. · 2005 · Class I recallK041602 · Panorama 1.0T · Philips Medical Systems North America, Inc. · 2004 · Class I recallK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalled−50+5
  1. Root · 6 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
  2. 5 generations backK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
  3. 4 generations backK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
  4. 3 generations backK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
  5. 2 generations backK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
  6. 1 generation backK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
  7. This deviceK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018
  8. Descendant · cites it directly · 13 of its ownK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
  9. Descendant · cites it directly · 11 of its ownK193215 Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia…Philips Medical Systems Nederland B.V. · 2020 · Recalled
  10. Descendant · 2 generations on · 6 of its ownK212673 MR 5300Philips Medical Systems Nederland B.V. · 2021 · Recalled
  11. Descendant · 2 generations on · 5 of its ownK213583 Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition…Philips Medical Systems Nederland B.V. · 2022 · Recalled
  12. Descendant · 2 generations on · 5 of its ownK213516 Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with…Philips Medical Systems Nederland B.V. · 2022 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2018: 12019: 22020: 12021: 12022: 42023: 32025: 22026: 1
2018high 4 (2022) · low 0 (2024)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 15 · See all
  • K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · cites it directly13descendants
  • K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2020 · cites it directly11descendants
  • K212673MR 5300Philips Medical Systems Nederland B.V. · 2021 · 2 generations on6descendants
  • K213583Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2022 · 2 generations on5descendants
  • K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi NudcleiPhilips Medical Systems Nederland B.V. · 2022 · 2 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K031815Intera-Sensation FamilyPhilips Medical Systems20033
−5K041602Panorama 1.0TPhilips Medical Systems North America, Inc.2004Class I
−4K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V.2005Class I
−3K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems2007Class I
−2K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care201120
−1K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V.20177
0 (this device)K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V.2018—
+1K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V.201925
+1K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR…Philips Medical Systems Nederland B.V.2020152
+2K212673MR 5300Philips Medical Systems Nederland B.V.202117
+2K213583Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V.202239
+2K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed…Philips Medical Systems Nederland B.V.202245

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200511

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
201920
202010
202110
202240
202330
202520
202610
FIG. 1 — Lineage of K173079, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3
Summary page 4
…Manufacturer: Philips Medical Systems Nederland B.V. 510(k) Clearance: K163116 Classification Regulation: 21CFR 892.1000 Cla…
Philips Medical Systems Nederland B.V.20177

Also cleared against K163116: 3 other devices →

What descends from it? (15)

See all 15

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V.201913
K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V.202011
K181479Ingenia Ambition S and Ingenia Ambition XPhilips Medical Systems Nederland B.V.20180

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 14, 2026Class IIZ-1959-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1952-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
Apr 14, 2026Class IIZ-1956-2026Under Investigation by firmThe potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol…
May 13, 2025Class IIZ-1788-2025Software designPotential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
May 13, 2025Class IIZ-1791-2025Software designPotential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
May 13, 2025Class IIZ-1790-2025Software designPotential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
May 13, 2025Class IIZ-1789-2025Software designPotential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
Oct 23, 2023Class IIZ-0393-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Oct 23, 2023Class IIZ-0389-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Oct 23, 2023Class IIZ-0390-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Oct 23, 2023Class IIZ-0392-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · 3 generations back · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
  • K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 1.3 years on median.
  • 9 of the 9 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K183063 (Philips Medical Systems Nederland B.V., 2019) has gone furthest, with 13 descendants of its own.
  • 3 other devices also cite its predicate K163116.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities. Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the…

Read the full statement (266 words)

This system is a Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities.

Magnetic Resonance images represent the spatial distribution of protons or other nuclei with spin. Image appearance is determined by many different physical properties of the tissue and the anatomy, and the MR scan technique applied. The image acquisition process can be synchronized with the patient’s breathing or cardiac cycle. The systems can use combinations of images to produce physical parameters, and related derived images.

Images, spectra, and measurements of physical parameters, when interpreted by a trained physician, provide information that may assist the diagnosis and therapy planning. The accuracy of determined physical parameters depends on system and scan parameters, and must be controlled and validated by the clinical user. The use of contrast agents for diagnostic imaging applications should be performed consistent with the approved labeling for the contrast agent.

In addition the Philips MR systems provide imaging capabilities, such as MR fluoroscopy, to guide and evaluate interventional and minimally invasive procedures in the head, body and extremities.

MR Interventional procedures, performed inside or adjacent to the Philips MR system, must be performed with MR Conditional or MR Safe instrumentation as selected and evaluated by the clinical user for use with the specific MR system configuration in the hospital. The appropriateness and use of information from a Philips MR system for a specific interventional procedure and specific MR system configuration must be validated by the clinical user.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173079: Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173079

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026