Recalled · Class I · 2026Achieva, Intera and Panorama 1.0T, Release 2.5
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medical Systems · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Jan 4, 2007 as substantially equivalent to K043147 (Philips Medical Systems North America Co., 2005) and 3 other predicates after a 38-day review. Its family tree reaches back 3 generations to 2003. 4 of the 60 devices in its family, counting itself, have a Class I recall. 53 later devices descend from it. It has been recalled 10 times, including a Class I recall.
- Decided
- Jan 4, 2007
- Review
- 38 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 10 (1 Class I)
- Predicates
- 4
Its longest line of descent runs back 3 generations to K031815 (2003). 6 ancestors in all, the oldest K001796 (2000). 53 later devices descend from it, 7 citing it directly. 4 of the 60 devices in this family, counting this one, have a Class I recall.
- Root · 3 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
- 2 generations backK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
- 1 generation backK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
- This deviceK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007
- Descendant · cites it directly · 34 of its ownK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · Recalled
- Descendant · 2 generations on · 19 of its ownK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
- Descendant · 3 generations on · 15 of its ownK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled
- Descendant · 4 generations on · 13 of its ownK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
- Descendant · 2 generations on · 13 of its ownK143606 MultiBand SENSEPhilips Medical Systems Nethrlands BV · 2015
Showing 5 of 53 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 53 · See all- K110151Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011 · cites it directly34descendants
- K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · 2 generations on19descendants
- K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · 3 generations on15descendants
- K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · 4 generations on13descendants
- K143606MultiBand SENSEPhilips Medical Systems Nethrlands BV · 2015 · 2 generations on13descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K031815 | Intera-Sensation Family | Philips Medical Systems | 2003 | 3 |
| −2 | K041602 | Panorama 1.0T | Philips Medical Systems North America, Inc. | 2004 | Class I |
| −1 | K052078 | Achieva, Intera and Panorama 1.0 Release 2-Series | Philips Medical Systems Nederland B.V. | 2005 | Class I |
| 0 (this device) | K063559 | Achieva, Intera and Panorama 1.0T, Release 2.5 | Philips Medical Systems | 2007 | — |
| +1 | K110151 | Achieva R4 1.5T and Achieva R4 3.0T | Philips Health Care | 2011 | 20 |
| +2 | K163116 | Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 | Philips Medical Systems Nederland B.V. | 2017 | 7 |
| +3 | K173079 | Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4 | Philips Medical Systems Nederland B.V. | 2018 | 11 |
| +4 | K183063 | Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems | Philips Medical Systems Nederland B.V. | 2019 | 25 |
| +2 | K143606 | MultiBand SENSE | Philips Medical Systems Nethrlands BV | 2015 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2005 | 1 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 2 | 0 |
| 2011 | 3 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2015 | 4 | 0 |
| 2016 | 3 | 0 |
| 2017 | 3 | 0 |
| 2018 | 9 | 0 |
| 2019 | 2 | 0 |
| 2020 | 3 | 0 |
| 2021 | 4 | 0 |
| 2022 | 8 | 0 |
| 2023 | 5 | 0 |
| 2025 | 3 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K043147 | Achieva FamilySummary page 1…, INTERA and PANORAMA 1.0T Release 2.5-series are the successor of the predicate devices ACHIEVA, INTERA and PANORAMA LOT release 2-series. (FDA references K043147, K041602, K052078 and KO 13344). Indications for use: The ACHIEVA, INTERA and PANORAMA 1.0T Release 2.5-series are… | 2005 |
| * | K041602 | Panorama 1.0TSummary page 1…and PANORAMA 1.0T Release 2.5-series are the successor of the predicate devices ACHIEVA, INTERA and PANORAMA LOT release 2-series. (FDA references K043147, K041602, K052078 and KO 13344). Indications for use: The ACHIEVA, INTERA and PANORAMA 1.0T Release 2.5-series are… | 2004 |
| * | K052078 | Achieva, Intera and Panorama 1.0 Release 2-SeriesSummary page 1…AMA 1.0T Release 2.5-series are the successor of the predicate devices ACHIEVA, INTERA and PANORAMA LOT release 2-series. (FDA references K043147, K041602, K052078 and KO 13344). Indications for use: The ACHIEVA, INTERA and PANORAMA 1.0T Release 2.5-series are magnetic… | 2005 |
| * | K013344 | Intera I/T | 2001 |
What descends from it? (53)
See all 537 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K110151 | Achieva R4 1.5T and Achieva R4 3.0T | 2011 | 34 |
| K100836 | Clearpoint System | 2010 | 12 |
| K181892 | Ziostation2 | 2018 | 1 |
| K162931 | Ingenia 1.5T CX and Ingenia 3.0T CX R5.3 | 2017 | 1 |
| K102344 | Mdixon Software Option for Intera 1.5T, Achieva 1.5T & Achieva 3.0T MR Systems | 2010 | 1 |
| K190461 | Achieva 1.5T, 3.0T and Intera 1.5T MR Systems | 2019 | 0 |
| K103483 | Ingenuity TF Pet/MR | 2011 | 0 |
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 21, 2026 | Class II | Z-3232-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| May 3, 2024 | Class II | Z-1855-2024 | Under Investigation by firm | The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire |
| Nov 17, 2023 | Class I | Z-0505-2024 | Component design/selection | Firm was informed of an event in which the structural integrity of the MR system components failed due to unintended excessive pressure buildup of helium gas during a magnet quench. If the system is unable to contain the pressured helium… |
| Oct 23, 2023 | Class II | Z-0393-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0389-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0390-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0392-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Aug 21, 2014 | Class II | Z-0325-2016 | Software Design Change | When a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed. This occurs with software release R5.1.7/R5.1.8. |
| May 5, 2014 | Class II | Z-0135-2016 | Software design | In spine clinical workflows, cross reference lines may be used to determine the position of slices. In cases, where MobiView fused Images are used to show the cross reference lines, the cross reference lines may be positioned incorrectly. |
| Aug 13, 2008 | Not classified | Z-0971-2009 | Process design | Battery Failure: The Invivo Lithium Polymer Wireless Module Battery Part Number 9065 and the Lithium Polymer Fiber Optic Module Battery Part Number 9067 is being recalled because the battery, itself, when dropped from 3 feet/1 meter onto… |
The 5 latest of 10 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
- K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
- 6 of the 6 ancestors have a recall of any class; 3 have a Class I recall.
- Of its descendants, K110151 (Philips Health Care, 2011) has gone furthest, with 34 descendants of its own.
- 5 other devices also cite its predicate K043147.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The ACIREVA, INTERA and PANORAMA LOT Release 2.5-series are magnetic resonance diagnostic devices that produce cross-sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
In addition the Achieva, Intera and Panorama 1.OT devices provide capabilities to perform interventional procedures in the head, body and extremities, which may be facilitated by MR techniques, such as real time imaging. Such procedures must be performed with MR compatible instrumentation as selected and evaluated by the clinical user.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K063559 · printed Nov 4, 2026
Cite this page
510(k) Report. “K063559: Achieva, Intera and Panorama 1.0T, Release 2.5.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K063559