510(k) · Cleared
Recalled · Class II · 2026Achieva R4 1.5T and Achieva R4 3.0T
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Health Care · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Mar 22, 2011 as substantially equivalent to K063559 (Philips Medical Systems, 2007) and 1 other predicate after a 63-day review. Its family tree reaches back 4 generations to 2003. 4 of the 42 devices in its family, counting itself, have a Class I recall. 34 later devices descend from it. It has been recalled 20 times.
- Decided
- Mar 22, 2011
- Review
- 63 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 20
- Predicates
- 2
Its longest line of descent runs back 4 generations to K031815 (2003). 7 ancestors in all, the oldest K001796 (2000). 34 later devices descend from it, 8 citing it directly. 4 of the 42 devices in this family, counting this one, have a Class I recall.
- Root · 4 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
- 3 generations backK041602 Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · Recalled (Class I)
- 2 generations backK052078 Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · Recalled (Class I)
- 1 generation backK063559 Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Recalled (Class I)
- This deviceK110151 Achieva R4 1.5T and Achieva R4 3.0TPhilips Health Care · 2011
- Descendant · cites it directly · 19 of its ownK163116 Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · Recalled
- Descendant · 2 generations on · 15 of its ownK173079 Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · Recalled
- Descendant · 3 generations on · 13 of its ownK183063 Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · Recalled
- Descendant · cites it directly · 13 of its ownK143606 MultiBand SENSEPhilips Medical Systems Nethrlands BV · 2015
- Descendant · cites it directly · 12 of its ownK162940 MultiBand SENSEPhilips Medical Systems Nederland B.V. · 2016
Showing 5 of 34 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 34 · See all- K163116Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3Philips Medical Systems Nederland B.V. · 2017 · cites it directly19descendants
- K173079Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4Philips Medical Systems Nederland B.V. · 2018 · 2 generations on15descendants
- K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019 · 3 generations on13descendants
- K143606MultiBand SENSEPhilips Medical Systems Nethrlands BV · 2015 · cites it directly13descendants
- K162940MultiBand SENSEPhilips Medical Systems Nederland B.V. · 2016 · cites it directly12descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K031815 | Intera-Sensation Family | Philips Medical Systems | 2003 | 3 |
| −3 | K041602 | Panorama 1.0T | Philips Medical Systems North America, Inc. | 2004 | Class I |
| −2 | K052078 | Achieva, Intera and Panorama 1.0 Release 2-Series | Philips Medical Systems Nederland B.V. | 2005 | Class I |
| −1 | K063559 | Achieva, Intera and Panorama 1.0T, Release 2.5 | Philips Medical Systems | 2007 | Class I |
| 0 (this device) | K110151 | Achieva R4 1.5T and Achieva R4 3.0T | Philips Health Care | 2011 | — |
| +1 | K163116 | Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 | Philips Medical Systems Nederland B.V. | 2017 | 7 |
| +2 | K173079 | Ingenia 1.5T, Ingenia 1 5T S, Ingenia 1.5T CX, Ingenia 3.0T CX R5.4 | Philips Medical Systems Nederland B.V. | 2018 | 11 |
| +3 | K183063 | Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems | Philips Medical Systems Nederland B.V. | 2019 | 25 |
| +1 | K143606 | MultiBand SENSE | Philips Medical Systems Nethrlands BV | 2015 | — |
| +1 | K162940 | MultiBand SENSE | Philips Medical Systems Nederland B.V. | 2016 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2005 | 1 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2015 | 2 | 0 |
| 2016 | 2 | 0 |
| 2017 | 1 | 0 |
| 2018 | 7 | 0 |
| 2019 | 2 | 0 |
| 2020 | 2 | 0 |
| 2021 | 3 | 0 |
| 2022 | 6 | 0 |
| 2023 | 3 | 0 |
| 2025 | 2 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K063559 | Achieva, Intera and Panorama 1.0T, Release 2.5Summary page 1…T and ACHIEVA R4 3.OT with enhancements are the successors of their predicate devices ACHIEVA 1.5T and ACHIEVA 3.OT MR systems Release 2.5-series (FDA references K063559, K043 147, K04 1602, K052078 and K013344.) Intended Use: The ACHIEVA R4 1.5T and ACHIEVA R4 3.OT are magnetic resonance… | 2007 |
| * | K043147 | Achieva FamilySummary page 1…IEVA R4 3.OT with enhancements are the successors of their predicate devices ACHIEVA 1.5T and ACHIEVA 3.OT MR systems Release 2.5-series (FDA references K063559, K043 147, K04 1602, K052078 and K013344.) Intended Use: The ACHIEVA R4 1.5T and ACHIEVA R4 3.OT are magnetic resonance diagnostic… | 2005 |
What descends from it? (34)
See all 348 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K163116 | Ingenia 1.5T, Ingenia 1.5T S and Ingenia 3.0T R5.3 | 2017 | 19 |
| K143606 | MultiBand SENSE | 2015 | 13 |
| K162940 | MultiBand SENSE | 2016 | 12 |
| K131241 | SWIP | 2013 | 5 |
| K151212 | Ziostation2 | 2015 | 2 |
| K123492 | DS Anterior, DS Base, DS Head-Neck, DS Head, Flex (S,M,L) | 2012 | 1 |
| K190461 | Achieva 1.5T, 3.0T and Intera 1.5T MR Systems | 2019 | 0 |
| K153324 | Ingenia 1.5T and Ingenia 1.5T S R5.2 | 2016 | 0 |
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 28, 2026 | Class II | Z-3314-2026 | Software design | When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party… |
| Aug 21, 2026 | Class II | Z-3243-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 11, 2021 | Class II | Z-2411-2021 | Device Design | When using manual mode, the tabletop may not completely move in or out and result in potential delay in patient evacuation from the bore, which could result in a delay of treatment |
| Jul 30, 2021 | Class II | Z-2346-2021 | Nonconforming Material/Component | Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury |
| Dec 21, 2018 | Class II | Z-0924-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0926-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0925-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 1, 2016 | Class II | Z-0825-2017 | Device Design | Possibility that a patient or user may be exposed to a very small level of touch current. |
| Aug 21, 2014 | Class II | Z-0325-2016 | Software Design Change | When a fused series of a sagittal, coronal or radial multi station scan is generated in MobiView, the resultant image order is reversed. This occurs with software release R5.1.7/R5.1.8. |
| Jul 27, 2014 | Class II | Z-2237-2014 | Software design | The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction… |
| Jun 13, 2014 | Class II | Z-1865-2014 | Nonconforming Material/Component | For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts. |
| Jun 13, 2014 | Class II | Z-1862-2014 | Nonconforming Material/Component | For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts. |
| Jun 13, 2014 | Class II | Z-1864-2014 | Nonconforming Material/Component | For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts. |
| Jun 13, 2014 | Class II | Z-1863-2014 | Nonconforming Material/Component | For some identified Ingenia systems shipped between December 2013 and March 2014, cover set rear magnetic M10 nuts were erroneously supplied instead of A4 stainless steel nuts. |
| May 15, 2014 | Class II | Z-1827-2014 | Device Design | Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF. |
| May 15, 2014 | Class II | Z-1826-2014 | Device Design | Ingenia customers have experienced clamping of the foot under the central column of the Height Adjustable Flex Trak trolley and the Trolley Variable Height IRF. |
| May 5, 2014 | Class II | Z-0135-2016 | Software design | In spine clinical workflows, cross reference lines may be used to determine the position of slices. In cases, where MobiView fused Images are used to show the cross reference lines, the cross reference lines may be positioned incorrectly. |
| Mar 11, 2014 | Class II | Z-1678-2014 | Software design | During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in… |
| Jan 27, 2012 | Not classified | Z-1125-2012 | Software design | There maybe incorrect values in the diffusion/perfusion maps. |
| Aug 3, 2011 | Not classified | Z-0293-2012 | Software design | The firm determined that during scanning procedures, for some larger patients, some reddening of the skin of can occur in certain instances due to RF heating when the torso of the patient is positioned within the body coil. |
The 5 latest of 20 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K063559Achieva, Intera and Panorama 1.0T, Release 2.5Philips Medical Systems · 2007 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
- K052078Achieva, Intera and Panorama 1.0 Release 2-SeriesPhilips Medical Systems Nederland B.V. · 2005 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K041602Panorama 1.0TPhilips Medical Systems North America, Inc. · 2004 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 8 years and passes through 4 companies; a new generation was cleared every 1.3 years on median.
- 7 of the 7 ancestors have a recall of any class; 4 have a Class I recall.
- Of its descendants, K163116 (Philips Medical Systems Nederland B.V., 2017) has gone furthest, with 19 descendants of its own.
- 6 other devices also cite its predicate K063559.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The ACHIEVA R4 1.5T and ACHIEVA R14 3.OT are magnetic resonance diagnostic devices that produce cross-sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
In addition, the ACHIEVA R4 1.5T and ACHIEVA R4 3.OT devices provide capabilities to perform interventional procedures in the head, body and extremities, which may be facilitated by MR techniques, such as real time imaging. Such procedures must be performed with MR compatible instrumentation as selected and evaluated by the clinical user.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K110151 · printed Nov 6, 2026
Cite this page
510(k) Report. “K110151: Achieva R4 1.5T and Achieva R4 3.0T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110151
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026