510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Optima MR450W

K091536510(k) TraditionalCleared Jul 17, 200952-day reviewClass II21 CFR 892.1000Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jul 17, 2009 as substantially equivalent to K083147 (Ge Medical Systems, LLC, 2008) after a 52-day review. Its family tree reaches back 8 generations to 1997. 5 of the 83 devices in its family, counting itself, have a Class I recall. 59 later devices descend from it. It has been recalled 8 times.

Decided
Jul 17, 2009
Review
52 days
Class
Class II
Product code
LNH
Recalls
8
Predicates
1

Its longest line of descent runs back 8 generations to K971684 (1997). 23 ancestors in all, the oldest K962061 (1996). 59 later devices descend from it, 3 citing it directly. 5 of the 83 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 3 other ancestors2004: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor, 1 with a Class I recall12002: 2 other ancestors, 1 with a Class I recall12001: 4 other ancestors, 1 with a Class I recall11999: 2 other ancestors1998: 1 other ancestor1996: 1 other ancestor2011: 2 descendants cleared2012: 2 descendants cleared2013: 5 descendants cleared2020: 6 descendants cleared2015: 5 descendants cleared2023: 6 descendants cleared2014: 2 descendants cleared2022: 5 descendants cleared2018: 2 descendants cleared2016: 4 descendants cleared2017: 5 descendants cleared2025: 4 descendants cleared2019: 4 descendants cleared2024: 2 descendants cleared2026: 4 descendants cleared2021: 1 descendant clearedReference device: K052293 · GE 1.5T and 3.0T Signa HDX MR System · Ge Medical Systems, LLC · 2005 · recalledK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK052164 · Magnetom Systems with Software Syngo MR 2006A · Siemens Medical Solutions USA, Inc. · 2005K062454 · Magnetom Systems - I-Class and T-Class Releases · Siemens Medi Cal Solutions, Inc. · 2006K072237 · Magnetom Verio · Siemens Medical Solutions USA, Inc. · 2007 · Class I recallK081028 · Signa 3.0T MR750 System · Ge Medical Systems, LLC · 2008 · recalledK083147 · Discovery MR450 · Ge Medical Systems, LLC · 2008 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor, 1 with a Class I recall14 generations back: 3 other ancestors, 2 with a Class I recall25 generations back: 5 other ancestors, 1 with a Class I recall16 generations back: 5 other ancestors 1 generation forward: 3 descendants2 generations forward: 5 descendants3 generations forward: 6 descendants4 generations forward: 7 descendants5 generations forward: 7 descendants6 generations forward: 7 descendants7 generations forward: 8 descendants8 generations forward: 5 descendants9 generations forward: 2 descendants10 generations forward: 5 descendants11 generations forward: 4 descendants K083147 · Discovery MR450 · Ge Medical Systems, LLC · 2008 · recalledK081028 · Signa 3.0T MR750 System · Ge Medical Systems, LLC · 2008 · recalledK072237 · Magnetom Verio · Siemens Medical Solutions USA, Inc. · 2007 · Class I recallK062454 · Magnetom Systems - I-Class and T-Class Releases · Siemens Medi Cal Solutions, Inc. · 2006K052164 · Magnetom Systems with Software Syngo MR 2006A · Siemens Medical Solutions USA, Inc. · 2005K041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalled−50+5+10
  1. Root · 8 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 7 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 6 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 5 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
  5. 4 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
  6. 3 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
  7. 2 generations backK081028 Signa 3.0T MR750 SystemGe Medical Systems, LLC · 2008 · Recalled
  8. 1 generation backK083147 Discovery MR450Ge Medical Systems, LLC · 2008 · Recalled
  9. This deviceK091536 Optima MR450WGe Medical Systems, LLC · 2009
  10. Descendant · cites it directly · 52 of its ownK103327 Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · Recalled
  11. Descendant · cites it directly · 44 of its ownK113490 Optima MR450WGe Medical Systems, LLC · 2012 · Recalled
  12. Descendant · 2 generations on · 43 of its ownK130115 Discovery MR750W 3.0TGe Medical Systems, LLC · 2013 · Recalled
  13. Descendant · 2 generations on · 43 of its ownK123522 Optima MR450WGe Healthcare (Ge Medical Systems, LLC) · 2013 · Recalled
  14. Descendant · 3 generations on · 42 of its ownK132376 Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T…Ge Healthcare (Ge Medical Systems, LLC) · 2013 · Recalled

Showing 5 of 59 descendants. See all descendants

Descendants cleared per year

2011: 22012: 22013: 52014: 22015: 52016: 42017: 52018: 22019: 42020: 62021: 12022: 52023: 62024: 22025: 42026: 4
2011high 6 (2020) · low 1 (2021)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 59 · See all
  • K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation · 2011 · cites it directly52descendants
  • K113490Optima MR450WGe Medical Systems, LLC · 2012 · cites it directly44descendants
  • K130115Discovery MR750W 3.0TGe Medical Systems, LLC · 2013 · 2 generations on43descendants
  • K123522Optima MR450WGe Healthcare (Ge Medical Systems, LLC) · 2013 · 2 generations on43descendants
  • K132376Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima MR450W 1.5TGe Healthcare (Ge Medical Systems, LLC) · 2013 · 3 generations on42descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−7K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−6K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−5K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc.2005—
−4K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc.2006—
−3K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I
−2K081028Signa 3.0T MR750 SystemGe Medical Systems, LLC200811
−1K083147Discovery MR450Ge Medical Systems, LLC200810
0 (this device)K091536Optima MR450WGe Medical Systems, LLC2009—
+1K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation20116
+1K113490Optima MR450WGe Medical Systems, LLC20123
+2K130115Discovery MR750W 3.0TGe Medical Systems, LLC20131
+2K123522Optima MR450WGe Healthcare (Ge Medical Systems, LLC)20131
+3K132376Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750W 3.0T, Optima…Ge Healthcare (Ge Medical Systems, LLC)201315

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199810
199920
200141
200221
200311
200411
200530

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201120
201220
201350
201420
201550
201640
201750
201820
201940
202060
202110
202250
202360
202420
202540
202640
FIG. 1 — Lineage of K091536, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K083147Discovery MR450
Summary page 1
…ce Name: Proprietor/Nome: Optima MR450w ClassificationNome: Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH Predicate Device: GE Discovery® MR450 System (K083147) Device Description: The 1.5T GE Optima MR450w features a superconducting magnet operating at 1.5 Tesla. The data acquisition sy…
Ge Medical Systems, LLC200810
†K052293GE 1.5T and 3.0T Signa HDX MR SystemGe Medical Systems, LLC200521

Also cleared against K083147: 3 other devices →

What descends from it? (59)

See all 59

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103327Discovery MR750W 3.0TGe Healthcare Japan Corporation201152
K113490Optima MR450WGe Medical Systems, LLC201244
K103335Gem Option for 1.5 Mri SystemsGE Healthcare20113

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 8, 2020Class IIZ-0128-2021Process change controlPotential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
May 11, 2017Class IIZ-3024-2017Mixed-up of materials/componentsPossible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Apr 23, 2014Class IIZ-1533-2014Process designGE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products. Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block mechanism malfunctions or…
Apr 2, 2014Class IIZ-1480-2014Process change controlPotential safety issue with gradient coil electromechanical connections associated with GE MR Products. Gradient coil electromechanical connections may loosen causing an increase in resistance. The increased resistance can lead to…
Jun 8, 2012Class IIZ-1207-2013Software designGE Healthcare has recently become aware of a Reformat Image Flip Issue associated with GE MR Products. The axial and coronal reformatted images from a sagittal 3D series (such as with applications of FSPGR-3D T1, BRAVO T1, or CUBE T1) of…
Jul 29, 2011Not classifiedZ-3121-2011Software designGE Healthcare has become aware of two potential safety issues associated with MR products (Optima MR450w, Discovery MR450, Discovery MR750, Brivo MR355, Optima MR360) that may impact patient safety. 1) Inline post-processed (i.e.…
May 10, 2011Class IIZ-0935-2013Software designAn error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports.
Nov 18, 2010Not classifiedZ-1377-2011Nonconforming Material/ComponentGE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may impact service or…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K041476GE Signa Excite 1.5T MR System, GE Signa Excite 3.0T MR SystemGe Medical Systems, LLC · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K030874GE 3.0T Signa Infinity Twinspeed with Excite MR SystemGeneral Electric Co. · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K013636Signa Infinity MR System with Excite TechnologyGE Medical Systems · 2002 · 4 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K0122001.5T Infinity Twinspeed MR SystemGE Medical Systems · 2001 · 5 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 11 months on median.
  • 16 of the 23 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K103327 (Ge Healthcare Japan Corporation, 2011) has gone furthest, with 52 descendants of its own.
  • 3 other devices also cite its predicate K083147.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The Optima MR450w is a whole body magnetic resonance scanner designed to support high resolution and high signal-to-noise ratio images in short exam times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique anatomical images, spectroscopic data, parametric maps, or dynamic images of the structures or functions of the entire body. The indication for use includes, but is not limited to, head, neck, TMJ…

Read the full statement (136 words)

The Optima MR450w is a whole body magnetic resonance scanner designed to support high resolution and high signal-to-noise ratio images in short exam times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique anatomical images, spectroscopic data, parametric maps, or dynamic images of the structures or functions of the entire body. The indication for use includes, but is not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the Optima MR450w reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and spectra, when interpreted by a trained physician yield information that may assist in diagnosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091536: Optima MR450W.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091536

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026