510(k) Report

510(k) · Cleared

Recalled · 2002

I-Flow Elastomeric Pump

K020862510(k) SpecialCleared Mar 26, 20028-day reviewClass II21 CFR 880.5725Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

I-Flow Corp. · MEB Pump, Infusion, Elastomeric

Cleared on Mar 26, 2002 as substantially equivalent to K932740 (Block Medical, Inc., 1993) and 4 other predicates after an 8-day review. Its family tree reaches back 4 generations to 1990. 2 of the 33 devices in its family, counting itself, have a Class I recall. 19 later devices descend from it. It has been recalled once.

Decided
Mar 26, 2002
Review
8 days
Class
Class II
Product code
MEB
Recalls
1
Predicates
5

Its longest line of descent runs back 4 generations to K896422 (1990). 13 ancestors in all, the oldest K884505 (1989). 19 later devices descend from it, 2 citing it directly. 2 of the 33 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1999: 3 other ancestors1995: 2 other ancestors1993: 1 other ancestor1998: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor2002: 2 descendants cleared, 1 with a Class I recall12003: 2 descendants cleared2006: 3 descendants cleared2004: 2 descendants cleared2005: 4 descendants cleared2009: 1 descendant cleared2008: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant clearedK896422 · Burron Ambulatory Drug Delivery System · Burron Medical, Inc. · 1990K980558 · Painbuster Infusion System · I-Flow Corp. · 1998×2: 2 generations cleared close together, K982946 (1998), K984502 (1999) 1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors2 generations back: 5 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 3 descendants3 generations forward: 4 descendants4 generations forward: 4 descendants5 generations forward: 2 descendants, 1 with a Class I recall16 generations forward: 1 descendant7 generations forward: 3 descendants K984502 · Nerve Block Infusion Kit · I-Flow Corp. · 1999K982946 · Painbuster Infusion Kit · I-Flow Corp. · 1998K980558 · Painbuster Infusion System · I-Flow Corp. · 1998K896422 · Burron Ambulatory Drug Delivery System · Burron Medical, Inc. · 19900+5
  1. Root · 4 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 3 generations backK980558 Painbuster Infusion SystemI-Flow Corp. · 1998
  3. 2 generations backK982946 Painbuster Infusion KitI-Flow Corp. · 1998
  4. 1 generation backK984502 Nerve Block Infusion KitI-Flow Corp. · 1999
  5. This deviceK020862 I-Flow Elastomeric PumpI-Flow Corp. · 2002
  6. Descendant · cites it directly · 17 of its ownK023098 Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser…Mckinley Infuser · 2002
  7. Descendant · 2 generations on · 15 of its ownK033039 Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc. · 2003
  8. Descendant · 2 generations on · 8 of its ownK032642 Beeline SystemMckinley, Inc. · 2003
  9. Descendant · 3 generations on · 7 of its ownK042228 Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004
  10. Descendant · 4 generations on · 6 of its ownK050770 Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2002: 22003: 22004: 22005: 42006: 32008: 12009: 12016: 12024: 22026: 1
2002high 4 (2005) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 19 · See all
  • K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser · 2002 · cites it directly17descendants
  • K033039Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc. · 2003 · 2 generations on15descendants
  • K032642Beeline SystemMckinley, Inc. · 2003 · 2 generations on8descendants
  • K042228Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004 · 3 generations on7descendants
  • K050770Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005 · 4 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K896422Burron Ambulatory Drug Delivery SystemBurron Medical, Inc.1990—
−3K980558Painbuster Infusion SystemI-Flow Corp.1998—
−2K982946Painbuster Infusion KitI-Flow Corp.1998—
−1K984502Nerve Block Infusion KitI-Flow Corp.1999—
0 (this device)K020862I-Flow Elastomeric PumpI-Flow Corp.2002—
+1K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser2002—
+2K033039Modification to: Accufuser, Accufuser Plus, Standard Procedure KitsMckinley, Inc.2003—
+2K032642Beeline SystemMckinley, Inc.2003—
+3K042228Beeline Motiv, Propolis, PcaMckinley, Inc.2004—
+4K050770Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC2005—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199310
199520
199810
199930

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200221
200320
200420
200540
200630
200811
200910
201610
202420
202610
FIG. 1 — Lineage of K020862, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MEB device

  1. #1 K250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)score 14.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K932740Homepump NSF
Summary page 1
…ion Trade Name: I-Flow Elastomeric Pump Common Name: Infusion Pump and Administration Set Classification Name: Pump, Infusion Existing Device: I-Flow Elastomeric Pump (K932740, K944692, K984502, K991513, K992072, and K020660 (pending)). Device Description: The |-Flow Elastomeric Pump consists of an ela…
Block Medical, Inc.19935
*K944692C-Series Homepump
Summary page 1
…e Name: I-Flow Elastomeric Pump Common Name: Infusion Pump and Administration Set Classification Name: Pump, Infusion Existing Device: I-Flow Elastomeric Pump (K932740, K944692, K984502, K991513, K992072, and K020660 (pending)). Device Description: The |-Flow Elastomeric Pump consists of an elastomeric…
Block Medical, Inc.1995—
*K984502Nerve Block Infusion Kit
Summary page 1
…-Flow Elastomeric Pump Common Name: Infusion Pump and Administration Set Classification Name: Pump, Infusion Existing Device: I-Flow Elastomeric Pump (K932740, K944692, K984502, K991513, K992072, and K020660 (pending)). Device Description: The |-Flow Elastomeric Pump consists of an elastomeric pressure…
I-Flow Corp.1999—
*K991513Homepump C-Series and Homepump C-Series One-Step Kvo
Summary page 1
…stomeric Pump Common Name: Infusion Pump and Administration Set Classification Name: Pump, Infusion Existing Device: I-Flow Elastomeric Pump (K932740, K944692, K984502, K991513, K992072, and K020660 (pending)). Device Description: The |-Flow Elastomeric Pump consists of an elastomeric pressure source wi…
I-Flow Corp.19991
*K992072Bolus Accessory Set
Summary page 1
…Pump Common Name: Infusion Pump and Administration Set Classification Name: Pump, Infusion Existing Device: I-Flow Elastomeric Pump (K932740, K944692, K984502, K991513, K992072, and K020660 (pending)). Device Description: The |-Flow Elastomeric Pump consists of an elastomeric pressure source with an int…
I-Flow Corp.1999—

Also cleared against K932740: 2 other devices →

What descends from it? (19)

See all 19

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser200217
K023318I-Flow Elastomeric Pump with Bolus Class II-Flow Corp.20020

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 19, 2002Not classifiedZ-0312-04OtherSlower than expected infusion of medications, such as antibiotics.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K023318I-Flow Elastomeric Pump with BolusI-Flow Corp. · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 11 Class I recall
  • K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · 5 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 3.1 years on median.
  • 3 of the 13 ancestors have a recall of any class.
  • Of its descendants, K023098 (Mckinley Infuser, 2002) has gone furthest, with 17 descendants of its own.
  • 2 other devices also cite its predicate K932740.

How does its category compare?

All MEB devices

MEB Pump, Infusion, Elastomeric: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MEB

2in 2024

2007: 52008: 22009: 12013: 12014: 12016: 12019: 12024: 22026 (partial year): 1
2007high 5 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEB

42in 2007

2007: 42 days median
2007high 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MEB

17in 2018

2007: 12008: 32009: 482010: 22011: 8, 1 Class I2013: 2, 1 Class I2018: 17
2007high 48 (2009) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MEB

1,408in 2025

2007: 1102008: 4442009: 7982010: 1,8272011: 2,2652012: 1,9762013: 1,9562014: 2,6952015: 2,9802016: 2,1472017: 1,6532018: 1,4732019: 9742020: 2,7322021: 1,8022022: 1,2882023: 1,2482024: 1,4242025: 1,408
2007high 2,980 (2015) · low 110 (2007)2025

MAUDE reports naming product code MEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075421110
20082—3444
20091—48798
20100—21,827
20110—8 (1)2,265
20120—01,976
20131—2 (1)1,956
20141—02,695
20150—02,980
20161—02,147
20170—01,653
20180—171,473
20191—0974
20200—02,732
20210—01,802
20220—01,288
20230—01,248
20242—01,424
20250—01,408
20261—0—

What is it indicated for?

The I-Flow Elastomeric Pump is intended for continuous and/or intermittent infusion of medications for general infusion use, including antibiotic delivery, chemotherapy and pain management. Routes of administration include the following: intravenous, intra-arterial, subcutaneous, intramuscular and epidural.

The I-Flow Elastomeric Pump is also intended for continuous and/or intermittent delivery of medication (such as local anesthetics or narcotics) to surgical wound sites and/or close proximity to nerves for preoperative, perioperative and postoperative regional anesthesia and pain management. Routes of administration may be intraoperative, perineural or percutaneous.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020862: I-Flow Elastomeric Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020862

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026