510(k) Report

510(k) · Cleared

Cleared

SMARTez elastomeric infusion pump

K151650510(k) TraditionalCleared Apr 4, 2016291-day reviewClass II21 CFR 880.5725

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Epic Medical Pte., Ltd. · MEB Pump, Infusion, Elastomeric

Cleared on Apr 4, 2016 as substantially equivalent to K052117 (I-Flow Corp., 2005) and 3 other predicates after a 291-day review. Its family tree reaches back 11 generations to 1990. 1 of the 51 devices in its family, counting itself, has a Class I recall. 3 later devices descend from it.

Decided
Apr 4, 2016
Review
291 days
Class
Class II
Product code
MEB
Recalls
None
Predicates
4

Its longest line of descent runs back 11 generations to K896422 (1990). 47 ancestors in all, the oldest K802820 (1980). 3 later devices descend from it, 3 citing it directly. 1 of the 51 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2005: 1 other ancestor1999: 6 other ancestors2006: 1 other ancestor2004: 2 other ancestors2000: 2 other ancestors1995: 2 other ancestors1990: 1 other ancestor2002: 1 other ancestor2001: 5 other ancestors1998: 2 other ancestors1991: 1 other ancestor1987: 1 other ancestor1985: 2 other ancestors1984: 1 other ancestor1980: 1 other ancestor2003: 1 other ancestor1996: 1 other ancestor1993: 2 other ancestors1989: 2 other ancestors2024: 2 descendants cleared2026: 1 descendant clearedK896422 · Burron Ambulatory Drug Delivery System · Burron Medical, Inc. · 1990×3: 3 generations cleared close together, K980558 (1998), K982946 (1998), K984502 (1999) K020862 · I-Flow Elastomeric Pump · I-Flow Corp. · 2002 · recalledK023098 · Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion Kit · Mckinley Infuser · 2002K032642 · Beeline System · Mckinley, Inc. · 2003×2: 2 generations cleared close together, K042228 (2004), K050770 (2005) K052451 · MPS Acacia Medflo Elastomeric Pump · Mps Acacia · 2005K081905 · Accuflow and Accuflux Elastomeric Infusion Pump · Westmed, Inc. · 2008 · Class I recall19801990200020102020
  1. Root · 11 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 7 generations between
  3. 3 generations backK050770 Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005
  4. 2 generations backK052451 MPS Acacia Medflo Elastomeric PumpMps Acacia · 2005
  5. 1 generation backK081905 Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · Recalled (Class I)
  6. This deviceK151650 SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd. · 2016
  7. Descendant · cites it directly · 0 of its ownK250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd. · 2026
  8. Descendant · cites it directly · 0 of its ownK240624 InfuLifeFirst Medical Source, LLC · 2024
  9. Descendant · cites it directly · 0 of its ownK242152 SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd. · 2024
Show all 11 generations
  1. Root · 11 generations backK896422 Burron Ambulatory Drug Delivery SystemBurron Medical, Inc. · 1990
  2. 10 generations backK980558 Painbuster Infusion SystemI-Flow Corp. · 1998
  3. 9 generations backK982946 Painbuster Infusion KitI-Flow Corp. · 1998
  4. 8 generations backK984502 Nerve Block Infusion KitI-Flow Corp. · 1999
  5. 7 generations backK020862 I-Flow Elastomeric PumpI-Flow Corp. · 2002 · Recalled
  6. 6 generations backK023098 Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser…Mckinley Infuser · 2002
  7. 5 generations backK032642 Beeline SystemMckinley, Inc. · 2003
  8. 4 generations backK042228 Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004
  9. 3 generations backK050770 Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005
  10. 2 generations backK052451 MPS Acacia Medflo Elastomeric PumpMps Acacia · 2005
  11. 1 generation backK081905 Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · Recalled (Class I)
  12. This deviceK151650 SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd. · 2016

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K250234SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd. · 2026 · cites it directly0descendants
  • K240624InfuLifeFirst Medical Source, LLC · 2024 · cites it directly0descendants
  • K242152SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd. · 2024 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K896422Burron Ambulatory Drug Delivery SystemBurron Medical, Inc.1990—
−10K980558Painbuster Infusion SystemI-Flow Corp.1998—
−9K982946Painbuster Infusion KitI-Flow Corp.1998—
−8K984502Nerve Block Infusion KitI-Flow Corp.1999—
−7K020862I-Flow Elastomeric PumpI-Flow Corp.20021
−6K023098Accufuser, Accufuser Plus, Standard Procedure Kit, Accufuser Infusion KitMckinley Infuser2002—
−5K032642Beeline SystemMckinley, Inc.2003—
−4K042228Beeline Motiv, Propolis, PcaMckinley, Inc.2004—
−3K050770Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC2005—
−2K052451MPS Acacia Medflo Elastomeric PumpMps Acacia2005—
−1K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc.2008Class I
0 (this device)K151650SMARTez elastomeric infusion pumpEpic Medical Pte., Ltd.2016—
+1K250234SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd.2026—
+1K240624InfuLifeFirst Medical Source, LLC2024—
+1K242152SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198410
198520
198710
198920
199010
199110
199320
199520
199610
199820
199960
200020
200150
200210
200310
200420
200510
200610
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202420
202610
FIG. 1 — Lineage of K151650, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MEB device

  1. #1 K250234 SMARTeZ™ Elastomeric Infusion Pump (RS series)score 14.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052117I-Flow Elastomeric Pump (on-Q, Painbuster, C-Block, Homepump, Easypump)
Summary page 5
…Table 5-2. Predicate devices SMARTeZ Pump is substantial equivalent (SE*) with Predicate Devices K052117 510(k) No. K071222 K081905 K991513 (Primar…
I-Flow Corp.20053
*K071222Infusor SV and LV Elastomeric Infusion Devices
Summary page 5
…antial equivalent (SE*) with Predicate Devices K052117 510(k) No. K071222 K081905 K991513 (Primary)…
Baxter Healthcare Corp20074
*K081905Accuflow and Accuflux Elastomeric Infusion Pump
Summary page 5
…with Predicate Devices K052117 510(k) No. K071222 K081905 K991513 (Primary) Westmed, Inc.…
Westmed, Inc.2008Class I
*K991513Homepump C-Series and Homepump C-Series One-Step Kvo
Summary page 5
…Predicate Devices K052117 510(k) No. K071222 K081905 K991513 (Primary) Westmed, Inc. 510(k)…
I-Flow Corp.19991

Also cleared against K052117: 2 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250234SMARTeZ™ Elastomeric Infusion Pump (RS series)Epic Medical Pte., Ltd.20260
K240624InfuLifeFirst Medical Source, LLC20240
K242152SMARTeZ™ Elastomeric Infusion Pump (498111, 498121, 498131, 498141)Epic Medical Pte., Ltd.20240

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 8 companies; a new generation was cleared every 9 months on median.
  • 12 of the 47 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K250234 (Epic Medical Pte., Ltd., 2026) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K052117.

How does its category compare?

All MEB devices

MEB Pump, Infusion, Elastomeric: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · MEB

2in 2024

2007: 52008: 22009: 12013: 12014: 12016: 12019: 12024: 22026 (partial year): 1
2007high 5 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEB

42in 2007

2007: 42 days median
2007high 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MEB

17in 2018

2007: 12008: 32009: 482010: 22011: 8, 1 Class I2013: 2, 1 Class I2018: 17
2007high 48 (2009) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MEB

1,408in 2025

2007: 1102008: 4442009: 7982010: 1,8272011: 2,2652012: 1,9762013: 1,9562014: 2,6952015: 2,9802016: 2,1472017: 1,6532018: 1,4732019: 9742020: 2,7322021: 1,8022022: 1,2882023: 1,2482024: 1,4242025: 1,408
2007high 2,980 (2015) · low 110 (2007)2025

MAUDE reports naming product code MEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075421110
20082—3444
20091—48798
20100—21,827
20110—8 (1)2,265
20120—01,976
20131—2 (1)1,956
20141—02,695
20150—02,980
20161—02,147
20170—01,653
20180—171,473
20191—0974
20200—02,732
20210—01,802
20220—01,288
20230—01,248
20242—01,424
20250—01,408
20261—0—

What is it indicated for?

The SMARTeZ Pump (Long infusion time article) is intended for continuous infusion of medications for general infusion use, including pain management.

* Routes of administration: intravenous and subcutaneous.

The SMARTeZ Pump (Short infusion time article) is intended for continuous infusion of medications for general infusion use, including antibiotic delivery.

* Route of administration: intravenous.

The SMARTeZ Pump (Chemotherapy article) is intended for continuous infusion of chemotherapy medications.

* Routes of administration: intravenous and intra-arterial.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K151650: SMARTez elastomeric infusion pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151650

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026