510(k) Report

510(k) · Cleared

Recalled · Class III · 2013

Steritite Rigid Reusable Sterilization Container System

K023614510(k) TraditionalCleared Apr 7, 2003161-day reviewClass II21 CFR 880.6850Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Case Medical, Inc. · FRG Wrap, Sterilization

Cleared on Apr 7, 2003 as substantially equivalent to K012931 (Allegiance Healthcare Corp., 2002) after a 161-day review. Its family tree reaches back 2 generations to 2000. 9 later devices descend from it. It has been recalled once.

Decided
Apr 7, 2003
Review
161 days
Class
Class II
Product code
FRG
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K991023 (2000). 9 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2005: 1 descendant cleared2006: 1 descendant cleared2010: 2 descendants cleared2014: 1 descendant cleared2013: 1 descendant cleared2012: 1 descendant cleared2018: 1 descendant cleared2024: 1 descendant clearedK991023 · Steritite Perforated Bottom Rigid Reusable Sterilization Container System with SCF02 - Polypropylene Non-Woven Disposabl · Case Medical, Inc. · 2000K012931 · Genesis Sterrad Compatible Reusable Sterilization Container System · Allegiance Healthcare Corp. · 2002200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 2 descendants K012931 · Genesis Sterrad Compatible Reusable Sterilization Container System · Allegiance Healthcare Corp. · 2002K991023 · Steritite Perforated Bottom Rigid Reusable Sterilization Container System with SCF02 - Polypropylene Non-Woven Disposabl · Case Medical, Inc. · 20000
  1. Root · 2 generations backK991023 Steritite Perforated Bottom Rigid Reusable Sterilization Container…Case Medical, Inc. · 2000
  2. 1 generation backK012931 Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp. · 2002
  3. This deviceK023614 Steritite Rigid Reusable Sterilization Container SystemCase Medical, Inc. · 2003
  4. Descendant · cites it directly · 7 of its ownK050570 Miltex Rigid Sterilization Container SystemMiltex, Inc. · 2005
  5. Descendant · 2 generations on · 6 of its ownK091720 Sklarlite Rigid Sterilization Container SystemSklar Corp. · 2010
  6. Descendant · 2 generations on · 3 of its ownK102146 Bahadir Sterilization TraysBahadir USA Corp. · 2010 · Recalled
  7. Descendant · 3 generations on · 2 of its ownK112090 Bahadir Sterilization TraysBahadir USA Corp. · 2012 · Recalled
  8. Descendant · 3 generations on · 1 of its ownK131407 Bahadir Sterilization TraysBahadir USA Corp. · 2014

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2005: 12006: 12010: 22012: 12013: 12014: 12018: 12024: 1
2005high 2 (2010) · low 0 (2007)2026

Most influential descendants

Showing 5 of 9 · See all
  • K050570Miltex Rigid Sterilization Container SystemMiltex, Inc. · 2005 · cites it directly7descendants
  • K091720Sklarlite Rigid Sterilization Container SystemSklar Corp. · 2010 · 2 generations on6descendants
  • K102146Bahadir Sterilization TraysBahadir USA Corp. · 2010 · 2 generations on3descendants
  • K112090Bahadir Sterilization TraysBahadir USA Corp. · 2012 · 3 generations on2descendants
  • K131407Bahadir Sterilization TraysBahadir USA Corp. · 2014 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K991023Steritite Perforated Bottom Rigid Reusable Sterilization Container System with…Case Medical, Inc.2000—
−1K012931Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp.2002—
0 (this device)K023614Steritite Rigid Reusable Sterilization Container SystemCase Medical, Inc.2003—
+1K050570Miltex Rigid Sterilization Container SystemMiltex, Inc.2005—
+2K091720Sklarlite Rigid Sterilization Container SystemSklar Corp.2010—
+2K102146Bahadir Sterilization TraysBahadir USA Corp.20101
+3K112090Bahadir Sterilization TraysBahadir USA Corp.20121
+3K131407Bahadir Sterilization TraysBahadir USA Corp.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200610
201020
201210
201310
201410
201810
202410
FIG. 1 — Lineage of K023614, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K012931Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp.2002—

Also cleared against K012931: 1 other device →

What descends from it? (9)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050570Miltex Rigid Sterilization Container SystemMiltex, Inc.20057
K061033Miltex Rigid Sterilization Container System, (47 Components)Miltex, Inc.20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 3, 2013Class IIIZ-0434-2014Device DesignAdvanced Sterilization Products (ASP) is recalling the SteriTite containers because they do not have consistent data to show that sterility assurance level of 10(-6) is achieved during the half-cycle test runs on new, unprocessed…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 3 years and passes through 2 companies; a new generation was cleared every 2.6 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K050570 (Miltex, Inc., 2005) has gone furthest, with 7 descendants of its own.
  • 1 other device also cites its predicate K012931.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The SteriTite container system and MediTray product is intended to be used to contain medical devices for sterilization, storage,

transportation and aseptic presentation of contents in health care facilities. This includes sterilization of surfaces and lumened dev

in EO, Metal lumens 2.2mm diameter or larger and length up to 45.7 cm and porous (silicone) lumened devices 3 mrn diameter or

larger and length up to 40 cm were validated. The SteriTite container is designed to maintain sterility of the contents until it is used

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023614: Steritite Rigid Reusable Sterilization Container System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023614

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026