510(k) Report

510(k) · Cleared

Cleared

Sklarlite Rigid Sterilization Container System

K091720510(k) TraditionalCleared Apr 9, 2010302-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sklar Corp. · KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories

Cleared on Apr 9, 2010 as substantially equivalent to K050570 (Miltex, Inc., 2005) after a 302-day review. Its family tree reaches back 4 generations to 2000. 6 later devices descend from it.

Decided
Apr 9, 2010
Review
302 days
Class
Class II
Product code
KCT
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K991023 (2000). 6 later devices descend from it, 3 citing it directly.

  1. Root · 4 generations backK991023 Steritite Perforated Bottom Rigid Reusable Sterilization Container…Case Medical, Inc. · 2000
  2. 3 generations backK012931 Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp. · 2002
  3. 2 generations backK023614 Steritite Rigid Reusable Sterilization Container SystemCase Medical, Inc. · 2003 · Recalled
  4. 1 generation backK050570 Miltex Rigid Sterilization Container SystemMiltex, Inc. · 2005
  5. This deviceK091720 Sklarlite Rigid Sterilization Container SystemSklar Corp. · 2010
  6. Descendant · cites it directly · 3 of its ownK102146 Bahadir Sterilization TraysBahadir USA Corp. · 2010 · Recalled
  7. Descendant · 2 generations on · 2 of its ownK112090 Bahadir Sterilization TraysBahadir USA Corp. · 2012 · Recalled
  8. Descendant · cites it directly · 1 of its ownK131407 Bahadir Sterilization TraysBahadir USA Corp. · 2014
  9. Descendant · cites it directly · 1 of its ownK123234 Ermis Sterilization Container SystemErmis Medizintechnik · 2013
  10. Descendant · 2 generations on · 0 of its ownK233578 Bahadir Sterilization ContainersBahadir USA, LLC · 2024

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2010: 12012: 12013: 12014: 12018: 12024: 1
2010high 1 (2010) · low 0 (2011)2026

Most influential descendants

Showing 5 of 6 · See all
  • K102146Bahadir Sterilization TraysBahadir USA Corp. · 2010 · cites it directly3descendants
  • K112090Bahadir Sterilization TraysBahadir USA Corp. · 2012 · 2 generations on2descendants
  • K131407Bahadir Sterilization TraysBahadir USA Corp. · 2014 · cites it directly1descendant
  • K123234Ermis Sterilization Container SystemErmis Medizintechnik · 2013 · cites it directly1descendant
  • K233578Bahadir Sterilization ContainersBahadir USA, LLC · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K991023Steritite Perforated Bottom Rigid Reusable Sterilization Container System with…Case Medical, Inc.2000—
−3K012931Genesis Sterrad Compatible Reusable Sterilization Container SystemAllegiance Healthcare Corp.2002—
−2K023614Steritite Rigid Reusable Sterilization Container SystemCase Medical, Inc.20031
−1K050570Miltex Rigid Sterilization Container SystemMiltex, Inc.2005—
0 (this device)K091720Sklarlite Rigid Sterilization Container SystemSklar Corp.2010—
+1K102146Bahadir Sterilization TraysBahadir USA Corp.20101
+2K112090Bahadir Sterilization TraysBahadir USA Corp.20121
+1K131407Bahadir Sterilization TraysBahadir USA Corp.2014—
+1K123234Ermis Sterilization Container SystemErmis Medizintechnik2013—
+2K233578Bahadir Sterilization ContainersBahadir USA, LLC2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201210
201310
201410
201810
202410
FIG. 1 — Lineage of K091720, 2000–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KCT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KCT device

  1. #1 K214041 Aesculap AICON Containerscore 18.1 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050570Miltex Rigid Sterilization Container System
Summary page 1
…ual Name: Closed Sterilization Container 4. Predicate Device Information Miltex Rigid Sterilization Container System, Miltex, Incorporated, K050570 5. Device Description The SklarLite Rigid Sterilization Container System consists of a family of ri…
Miltex, Inc.2005—

Also cleared against K050570: 1 other device →

What descends from it? (6)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102146Bahadir Sterilization TraysBahadir USA Corp.20103
K131407Bahadir Sterilization TraysBahadir USA Corp.20141
K123234Ermis Sterilization Container SystemErmis Medizintechnik20131

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K102146 (Bahadir USA Corp., 2010) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K050570.

How does its category compare?

All KCT devices

KCT Sterilization Wrap Containers, Trays, Cassettes & Other Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · KCT

11in 2025

2007: 32008: 62009: 62010: 122011: 62012: 102013: 82014: 112015: 152016: 122017: 102018: 152019: 182020: 142021: 182022: 112023: 102024: 82025: 112026 (partial year): 9
2007high 18 (2019) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KCT

205in 2025

2007: 199 days median2008: 182 days median2009: 89 days median2010: 202 days median2011: 117 days median2012: 236 days median2013: 131 days median2014: 169 days median2015: 151 days median2016: 182 days median2017: 211 days median2018: 192 days median2019: 187 days median2020: 134 days median2021: 185 days median2022: 137 days median2023: 94 days median2024: 106 days median2025: 205 days median2026 (partial year): 122 days median
2007high 236 (2012) · low 89 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KCT

17in 2022

2010: 12011: 22013: 12014: 12016: 32017: 12021: 12022: 17
2007high 17 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KCT

50in 2025

2007: 22008: 82009: 52010: 32011: 32012: 22013: 62014: 52015: 62016: 162017: 32018: 222019: 152020: 662021: 262022: 82023: 302024: 182025: 50
2007high 66 (2020) · low 2 (2007)2025

MAUDE reports naming product code KCT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007319902
2008618208
200968905
20101220213
2011611723
20121023602
2013813116
20141116915
20151515106
201612182316
20171021113
201815192022
201918187015
202014134066
202118185126
202211137178
20231094030
20248106018
202511205050
202691220—

What is it indicated for?

The SklarLiteTm Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized. Containers allow sterilization of the enclosed medical devices, including surfaces and lumens, using high vacuum steam sterilizers at 270'F for 4 minutes with 30 minutes (minimum) dry time. The containers have been validated for sterilization of up to two instruments with lumens no…

Read the full statement (200 words)

The SklarLiteTm Rigid Sterilization Container System is intended for use in hospitals and health care facilities to contain other medical devices that are to be sterilized. Containers allow sterilization of the enclosed medical devices, including surfaces and lumens, using high vacuum steam sterilizers at 270'F for 4 minutes with 30 minutes (minimum) dry time.

The containers have been validated for sterilization of up to two instruments with lumens no smaller than 2.4 mm I.D. in diameter and no longer than 380 mm in length for the Full-Size (580mm x 280mm) and Mid-Size (465mm x 280mm) containers, and no smaller than 1.0 mm I.D. in diameter and no longer than 250 mm in length for the Half-Size (285mm x 280mm) container. Containers have been validated for a combined maximum load of 25lbs. with up to two internal stacked baskets and up to two silicon mats (one mat per basket) for the Full-Size, Mid-Size and Half-Size containers. Both solid and perforated bottom containers, maximum depth 150mm/each, can be stacked a maximum of 3 containers high during sterilization.

Sterilized devices may be stored and transported within the container. The container is intended to maintain sterility of the contents for up to six (6)months.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K091720: Sklarlite Rigid Sterilization Container System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091720

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026