510(k) · Cleared
Recalled · Class II · 2013Innova 3100
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
GE Medical Systems · OWB Interventional Fluoroscopic X-Ray System
Cleared on Aug 22, 2003 as substantially equivalent to K023178 (Ge Medical Systems, Inc., 2002) and 1 other predicate after an 87-day review. Its family tree reaches back 3 generations to 1989. 76 later devices descend from it. It has been recalled 25 times.
- Decided
- Aug 22, 2003
- Review
- 87 days
- Class
- Class II
- Product code
- OWB
- Recalls
- 25
- Predicates
- 2
Its longest line of descent runs back 3 generations to K890348 (1989). 7 ancestors in all. 76 later devices descend from it, 3 citing it directly.
- Root · 3 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
- 2 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
- 1 generation backK022322 Innova and Innova 2000 SGe Medical Systems, Inc. · 2002 · Recalled
- This deviceK031637 Innova 3100GE Medical Systems · 2003
- Descendant · cites it directly · 75 of its ownK050489 Innova 2100Ge Medical Systems, LLC · 2005 · Recalled
- Descendant · cites it directly · 74 of its ownK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
- Descendant · 2 generations on · 47 of its ownK060259 Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · Recalled
- Descendant · 3 generations on · 23 of its ownK081582 Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · Recalled
- Descendant · cites it directly · 22 of its ownK061163 Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova…Ge Medical Systems, LLC · 2006 · Recalled
Showing 5 of 76 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 76 · See all- K050489Innova 2100Ge Medical Systems, LLC · 2005 · cites it directly75descendants
- K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare · 2005 · cites it directly74descendants
- K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · 2 generations on47descendants
- K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · 3 generations on23descendants
- K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…Ge Medical Systems, LLC · 2006 · cites it directly22descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K890348 | LC, Cardiac Positioner #B5078A | General Electric Co. | 1989 | 5 |
| −2 | K993037 | LCV+ Version 2 | Ge Medical Systems, Inc. | 2000 | 6 |
| −1 | K022322 | Innova and Innova 2000 S | Ge Medical Systems, Inc. | 2002 | 7 |
| 0 (this device) | K031637 | Innova 3100 | GE Medical Systems | 2003 | — |
| +1 | K050489 | Innova 2100 | Ge Medical Systems, LLC | 2005 | 27 |
| +1 | K052412 | Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ | GE Healthcare | 2005 | 50 |
| +2 | K060259 | Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121 | Ge Medical Systems, LLC | 2006 | 4 |
| +3 | K081582 | Infx-8000V, Infinix-Cfi and Infinix Vfi | Toshiba America Medical Systems, In.C | 2008 | 1 |
| +1 | K061163 | Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova… | Ge Medical Systems, LLC | 2006 | 38 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1995 | 2 | 0 |
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 2 | 0 |
| 2006 | 2 | 0 |
| 2008 | 4 | 0 |
| 2009 | 3 | 0 |
| 2010 | 3 | 0 |
| 2011 | 4 | 0 |
| 2012 | 3 | 0 |
| 2013 | 2 | 0 |
| 2014 | 4 | 0 |
| 2015 | 6 | 0 |
| 2016 | 4 | 0 |
| 2017 | 4 | 0 |
| 2018 | 10 | 0 |
| 2020 | 3 | 0 |
| 2021 | 3 | 0 |
| 2022 | 4 | 0 |
| 2023 | 3 | 0 |
| 2024 | 5 | 0 |
| 2025 | 4 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OWB device
- #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K023178 | Innova 4100Summary page 1…: GE Medical Systems, Milwaukee, Wi Marketed Devices: * The Innova 3100 is substantially equivalent to the currently marketed Vascular Angiographic system Innova 4100 (K023178) and complies with the same or equivalent standards. The SuperFast Gantry (InnovaTrace) with the capacitive sensor feature in the… | 2002 |
| * | K022322 | Innova and Innova 2000 SSummary page 1…rds. The SuperFast Gantry (InnovaTrace) with the capacitive sensor feature in the Innova 3100 system is substantially equivalent to the Innova 2000/Innova 2000S systems (K022322) for this feature. Device Description: The Innova 3100 is designed to perform fluoroscopic x-ray examinations. The detector is c… | 2002 |
What descends from it? (76)
See all 763 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 11, 2013 | Class II | Z-1163-2014 | Software design | GE Healthcare has become aware of a potential safety issue involving the innova systems. The Innova system can unexpectedly stop delivering X-rays after a power-on or a reset cycle which may result in the loss of real-time interventional… |
| Jan 21, 2013 | Class II | Z-0947-2013 | Nonconforming Material/Component | A locking nut might loosen out of the overhead Video Monitor Suspension yoke affecting Advantx, Innova 2000, Innova 2100IQ, Innova 3100, Innova 3100IQ, Innova 4100, and Innova 4100IQ Cardiovascular X-ray imaging systems. |
| Jun 3, 2010 | Not classified | Z-1922-2010 | Device Design | GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. |
| Jun 3, 2010 | Not classified | Z-1923-2010 | Device Design | GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. |
| Jun 3, 2010 | Not classified | Z-1921-2010 | Device Design | GE Healthcare has become aware of a potential loss of gantry and table motions affecting Innova 2100IQ, 3100, 3100IQ, 4100 and 4100IQ systems with InnovaIQ Table that may impact patient safety. |
| Dec 22, 2009 | Not classified | Z-0618-2010 | Device Design | Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational… |
| Jul 31, 2009 | Not classified | Z-2241-2009 | Component design/selection | GE Healthcare has recently became aware of a potential power supply failure with the Innova 2000, 3100 and 4100 systems that may impact patient safety. A potential power supply failure inside the GE equipment cabinet could occur with no… |
| Apr 16, 2009 | Not classified | Z-0135-2010 | Device Design | Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room… |
| Oct 9, 2008 | Not classified | Z-1688-2009 | Radiation Control for Health and Safety Act | GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE… |
| Oct 9, 2008 | Not classified | Z-1689-2009 | Radiation Control for Health and Safety Act | GE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE… |
| Jul 25, 2008 | Not classified | Z-2437-2008 | Device Design | The charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down. |
| Jun 24, 2008 | Not classified | Z-2356-2008 | Employee error | GE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two… |
| Jun 17, 2008 | Not classified | Z-1475-2009 | Process control | GE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during… |
| May 19, 2008 | Not classified | Z-1522-2008 | Software design | GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room). |
| May 19, 2008 | Not classified | Z-1521-2008 | Software design | GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room). |
| Mar 6, 2008 | Not classified | Z-1092-2009 | Software design | GE Healthcare reported an issue with the Exposure Switch Redundancy system with Innova Systems 3100 & 4100, in which this error stays on, even after the pedal is released, inhibiting further x-rays, and a system reset is required to… |
| Jan 30, 2008 | Not classified | Z-0474-2009 | Software design | Software error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and… |
| Jan 14, 2008 | Not classified | Z-2040-2008 | Process control | Risk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system |
| Mar 26, 2007 | Not classified | Z-0871-2007 | Other | Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control… |
| Mar 26, 2007 | Not classified | Z-0870-2007 | Other | Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control… |
| Mar 26, 2007 | Not classified | Z-0869-2007 | Other | Measurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control… |
| Mar 26, 2007 | Not classified | Z-1044-2007 | Other | Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense… |
| Mar 26, 2007 | Not classified | Z-1043-2007 | Other | Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense… |
| Feb 5, 2007 | Not classified | Z-0914-2007 | Other | Unintended Gantry Movements : GE Healthcare has recently become aware of accidental, operated-initiated activation of gantry movements associated with the new, metallic Smart Box controller that may impact patient or staff safely. |
| Mar 6, 2006 | Not classified | Z-0897-06 | Other | Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient. |
The 5 latest of 25 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 2.6 years on median.
- 5 of the 7 ancestors have a recall of any class.
- Of its descendants, K050489 (Ge Medical Systems, LLC, 2005) has gone furthest, with 75 descendants of its own.
- 4 other devices also cite its predicate K023178.
How does its category compare?
All OWB devicesOWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · OWB
22in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OWB
148in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OWB
42so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OWB
14,008in 2025
MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 52 | 0 | 1 |
| 2008 | 4 | 135 | 1 | 0 |
| 2009 | 9 | 99 | 0 | 0 |
| 2010 | 7 | 66 | 0 | 0 |
| 2011 | 6 | 79 | 0 | 1 |
| 2012 | 6 | 37 | 0 | 1 |
| 2013 | 11 | 119 | 10 | 27 |
| 2014 | 15 | 158 | 9 | 103 |
| 2015 | 11 | 66 | 35 | 2,134 |
| 2016 | 16 | 103 | 13 | 4,074 |
| 2017 | 20 | 76 | 13 | 1,604 |
| 2018 | 23 | 54 | 6 | 1,449 |
| 2019 | 9 | 174 | 7 | 1,558 |
| 2020 | 16 | 53 | 8 | 1,581 |
| 2021 | 24 | 60 | 31 | 1,981 |
| 2022 | 21 | 90 | 24 | 2,932 |
| 2023 | 12 | 152 | 22 | 7,604 |
| 2024 | 13 | 126 | 11 | 7,942 |
| 2025 | 22 | 148 | 30 | 14,008 |
| 2026 | 15 | 221 | 42 | — |
What is it indicated for?
The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and interventional angiography procedures of human anatomy. It is intended fo replace image intensifier fluoroscopic systems in all diagnostic or interventional angiography procedures. This device is not intended for mammography applications.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K031637 · printed Nov 6, 2026
Cite this page
510(k) Report. “K031637: Innova 3100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031637
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026