510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ

K052412510(k) TraditionalCleared Sep 16, 200514-day reviewClass II21 CFR 892.1650Recalls 50

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare · OWB Interventional Fluoroscopic X-Ray System

Cleared on Sep 16, 2005 as substantially equivalent to K974715 (GE Medical Systems, 1998) and 3 other predicates after a 14-day review. Its family tree reaches back 5 generations to 1989. 74 later devices descend from it. It has been recalled 50 times.

Decided
Sep 16, 2005
Review
14 days
Class
Class II
Product code
OWB
Recalls
50
Predicates
4

Its longest line of descent runs back 5 generations to K890348 (1989). 14 ancestors in all. 74 later devices descend from it, 8 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1998: 1 other ancestor2002: 2 other ancestors1995: 3 other ancestors2001: 1 other ancestor1993: 1 other ancestor2006: 2 descendants cleared2010: 3 descendants cleared2009: 3 descendants cleared2016: 4 descendants cleared2008: 4 descendants cleared2011: 4 descendants cleared2018: 10 descendants cleared2012: 3 descendants cleared2022: 4 descendants cleared2017: 4 descendants cleared2013: 2 descendants cleared2015: 6 descendants cleared2014: 4 descendants cleared2021: 3 descendants cleared2024: 5 descendants cleared2020: 3 descendants cleared2025: 4 descendants cleared2023: 3 descendants cleared2026: 3 descendants clearedReference device: K954355 · Advantage Windows 3D with Navigator Option · General Electric Co. · 1995Reference device: K993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledReference device: K022322 · Innova and Innova 2000 S · Ge Medical Systems, Inc. · 2002 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK022322 · Innova and Innova 2000 S · Ge Medical Systems, Inc. · 2002 · recalledK031637 · Innova 3100 · GE Medical Systems · 2003 · recalledK050489 · Innova 2100 · Ge Medical Systems, LLC · 2005 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 4 other ancestors 1 generation forward: 8 descendants2 generations forward: 10 descendants3 generations forward: 15 descendants4 generations forward: 11 descendants5 generations forward: 9 descendants6 generations forward: 4 descendants7 generations forward: 5 descendants8 generations forward: 4 descendants9 generations forward: 3 descendants10 generations forward: 2 descendants11 generations forward: 2 descendants12 generations forward: 1 descendant K050489 · Innova 2100 · Ge Medical Systems, LLC · 2005 · recalledK031637 · Innova 3100 · GE Medical Systems · 2003 · recalledK022322 · Innova and Innova 2000 S · Ge Medical Systems, Inc. · 2002 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalled−50+5+10
  1. Root · 5 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 4 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 3 generations backK022322 Innova and Innova 2000 SGe Medical Systems, Inc. · 2002 · Recalled
  4. 2 generations backK031637 Innova 3100GE Medical Systems · 2003 · Recalled
  5. 1 generation backK050489 Innova 2100Ge Medical Systems, LLC · 2005 · Recalled
  6. This deviceK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005
  7. Descendant · cites it directly · 47 of its ownK060259 Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · Recalled
  8. Descendant · 2 generations on · 23 of its ownK081582 Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · Recalled
  9. Descendant · cites it directly · 22 of its ownK061163 Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova…Ge Medical Systems, LLC · 2006 · Recalled
  10. Descendant · cites it directly · 21 of its ownK101868 Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010
  11. Descendant · 2 generations on · 20 of its ownK120073 Infx-8000V withToshibamedical Systems Corporation · 2012 · Recalled

Showing 5 of 74 descendants. See all descendants

Descendants cleared per year

2006: 22008: 42009: 32010: 32011: 42012: 32013: 22014: 42015: 62016: 42017: 42018: 102020: 32021: 32022: 42023: 32024: 52025: 42026: 3
2006high 10 (2018) · low 0 (2007)2026

Most influential descendants

Showing 5 of 74 · See all
  • K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · cites it directly47descendants
  • K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C · 2008 · 2 generations on23descendants
  • K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…Ge Medical Systems, LLC · 2006 · cites it directly22descendants
  • K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C · 2010 · cites it directly21descendants
  • K120073Infx-8000V withToshibamedical Systems Corporation · 2012 · 2 generations on20descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−4K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−3K022322Innova and Innova 2000 SGe Medical Systems, Inc.20027
−2K031637Innova 3100GE Medical Systems200325
−1K050489Innova 2100Ge Medical Systems, LLC200527
0 (this device)K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare2005—
+1K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC20064
+2K081582Infx-8000V, Infinix-Cfi and Infinix VfiToshiba America Medical Systems, In.C20081
+1K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova…Ge Medical Systems, LLC200638
+1K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C2010—
+2K120073Infx-8000V withToshibamedical Systems Corporation20123

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199530
199810
200110
200220
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200620
200840
200930
201030
201140
201230
201320
201440
201560
201640
201740
2018100
202030
202130
202240
202330
202450
202540
202630
FIG. 1 — Lineage of K052412, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K974715Advantage 3D XR
Summary page 1
…for Innova 4100 Innova 4100'°, Innova 3100, Innova 3100'@ , Innova 2100'® devices Is substantlally equivalent to the currently marketed software option Advantage 3D XR (K974715). The GE Healthcare InnovaSpin optlon for Innova 4100, Innova 4100'%, Innova .3100, Innova 3100'® devices is substantially equnv…
GE Medical Systems1998—
*K050489Innova 2100
Summary page 1
…%, Innova .3100, Innova 3100'® devices is substantially equnvalent to the currently marketed InnovaSpin optuon of the Vascular Angiographic system Innova 2100 (2100-1Q) (K050489). The GE Healthcare Innova 4100'° and Innova 3100 devices are substantially equivalent to the currently marketed devices Innova…
Ge Medical Systems, LLC200527
*K033244Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option
Summary page 1
…-1Q) (K050489). The GE Healthcare Innova 4100'° and Innova 3100 devices are substantially equivalent to the currently marketed devices Innova 4100 device cleared under K033244 and Innova 3100 device cleared under K031637. This opinion is based on the information contained in the comparison table and the…
Ge Medical Systems, Inc.20038
*K031637Innova 3100
Summary page 1
…0'° and Innova 3100 devices are substantially equivalent to the currently marketed devices Innova 4100 device cleared under K033244 and Innova 3100 device cleared under K031637. This opinion is based on the information contained in the comparison table and the product data sheets. The Innova 3D is offer…
GE Medical Systems200325
†K954355Advantage Windows 3D with Navigator OptionGeneral Electric Co.1995—
†K993037LCV+ Version 2Ge Medical Systems, Inc.20006
†K022322Innova and Innova 2000 SGe Medical Systems, Inc.20027

Also cleared against K974715: 1 other device →

What descends from it? (74)

See all 74

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC200647
K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ…Ge Medical Systems, LLC200622
K101868Infinix-I Interventional Angiography System Model Infx-8000VToshiba America Medical Systems, In.C201021
K081624Infx-8000F (Infinix Csi and Infinix Dpi)Toshiba America Medical Systems, In.C200813
K091658GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models Innova 4100, 3100…GE Healthcare200911
K081621Infx-8000C (Infinix CC-I and VC-I)Toshiba America Medical Systems, In.C200810
K152352Stereo 3D option for Vision ApplicationsGE Medical Systems SCS20161
K092004GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQGE Healthcare20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 4, 2014Class IIZ-0453-2015Device DesignLoss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
Dec 20, 2011Class IIZ-1318-2013Nonconforming Material/ComponentWhile performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries…
Oct 21, 2011Class IIZ-1142-2013Labeling designAs part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff…
Aug 24, 2011Not classifiedZ-0133-2012Nonconforming Material/ComponentA potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails.
May 16, 2011Class IIZ-1616-2013Software designGE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
Feb 4, 2011Class IIZ-0934-2013Nonconforming Material/ComponentExposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
Mar 24, 2010Not classifiedZ-1660-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Mar 24, 2010Not classifiedZ-1659-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Mar 24, 2010Not classifiedZ-1661-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Dec 22, 2009Not classifiedZ-0618-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0617-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0619-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Aug 5, 2009Not classifiedZ-0408-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0410-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0409-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Apr 16, 2009Not classifiedZ-0136-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0135-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0137-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Nov 4, 2008Not classifiedZ-1036-2009Radiation Control for Health and Safety ActOne of three copper spectral filters could remain within the x-ray while the system displays no spectral filter. The displayed Air Kerma rate and thus cumulative Air Kerma are overestimated. The Air Kerma rate may be over estimated by up…
Oct 9, 2008Not classifiedZ-1691-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1688-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1687-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1690-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1689-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Jul 25, 2008Not classifiedZ-2438-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jul 25, 2008Not classifiedZ-2436-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jul 25, 2008Not classifiedZ-2437-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jun 24, 2008Not classifiedZ-2357-2008Employee errorGE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two…
Jun 24, 2008Not classifiedZ-2356-2008Employee errorGE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two…
Jun 24, 2008Not classifiedZ-2358-2008Employee errorGE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two…
Jun 17, 2008Not classifiedZ-1475-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1476-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1474-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jan 30, 2008Not classifiedZ-0473-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0475-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0474-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 14, 2008Not classifiedZ-2039-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2040-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2041-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Nov 15, 2007Not classifiedZ-0692-2008Device DesignFrozen Image Problems: During an acquisition (fluoro and/or record), there have been cases reported where an image became "frozen" on the DL (digital leader acquisition system) live monitor screen. In such cases, the system continued to…
Nov 15, 2007Not classifiedZ-0691-2008Device DesignFrozen Image Problems: During an acquisition (fluoro and/or record), there have been cases reported where an image became "frozen" on the DL (digital leader acquisition system) live monitor screen. In such cases, the system continued to…
Mar 26, 2007Not classifiedZ-0871-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-0870-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-0869-2007OtherMeasurement errors during the computation of vessel diameters while using the Stenosis Analysis software (also known as Quantitative Coronary Analysis function-QCA) using the auto-calibration function on Innova Digital (located in control…
Mar 26, 2007Not classifiedZ-1044-2007OtherCaution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense…
Mar 26, 2007Not classifiedZ-1043-2007OtherCaution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense…
Feb 5, 2007Not classifiedZ-0913-2007OtherUnintended Gantry Movements : GE Healthcare has recently become aware of accidental, operated-initiated activation of gantry movements associated with the new, metallic Smart Box controller, that may impact patient or staff safely.
Aug 17, 2006Not classifiedZ-1518-2008Radiation Control for Health and Safety ActUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to…
Aug 17, 2006Not classifiedZ-1519-2008Radiation Control for Health and Safety ActUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to…
Aug 17, 2006Not classifiedZ-1520-2008Radiation Control for Health and Safety ActUser was unable to hear the Innova 2100-IQ 5-minute fluoroscopy warning signal from the Table Control TSSC interface. Inability of the user to hear the audible signal indicating the completion of any preset exposure time may lead to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 1.6 years on median.
  • 10 of the 14 ancestors have a recall of any class.
  • Of its descendants, K060259 (Ge Medical Systems, LLC, 2006) has gone furthest, with 47 descendants of its own.
  • 1 other device also cites its predicate K974715.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscoplc images obtained through image intensifier technology. Those devices are…

Read the full statement (115 words)

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.

They are intended to replace fluoroscoplc images obtained through image intensifier technology. Those devices are not intended for mammography applications.

Innova 3D is a software option which reconstructs 3D volumes from Rotational Fluoroscopy acquisition to assist the physician in diagnosis, surgical planning, interventional procedures and treatment follow-up. It is not intended for mammography applications.

InnovaSpin is a software option that permits fast spin rotational angiography. 1t is not intended for mammography applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K052412: Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052412

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026