510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ, Innova 2100IQ with Innova Ivus Opt

K061163510(k) TraditionalCleared Jun 7, 200642-day reviewClass II21 CFR 892.1650Recalls 38

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · OWB Interventional Fluoroscopic X-Ray System

Cleared on Jun 7, 2006 as substantially equivalent to K060259 (Ge Medical Systems, LLC, 2006) and 4 other predicates after a 42-day review. Its family tree reaches back 7 generations to 1989. 22 later devices descend from it. It has been recalled 38 times.

Decided
Jun 7, 2006
Review
42 days
Class
Class II
Product code
OWB
Recalls
38
Predicates
5

Its longest line of descent runs back 7 generations to K890348 (1989). 17 ancestors in all. 22 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2002: 2 other ancestors1998: 2 other ancestors2001: 1 other ancestor1995: 3 other ancestors1993: 1 other ancestor2009: 3 descendants cleared2011: 2 descendants cleared2010: 2 descendants cleared2013: 1 descendant cleared2012: 2 descendants cleared2017: 1 descendant cleared2014: 1 descendant cleared2018: 1 descendant cleared2015: 1 descendant cleared2021: 1 descendant cleared2024: 1 descendant cleared2020: 1 descendant cleared2025: 2 descendants cleared2023: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant clearedK890348 · LC, Cardiac Positioner #B5078A · General Electric Co. · 1989 · recalledK993037 · LCV+ Version 2 · Ge Medical Systems, Inc. · 2000 · recalledK022322 · Innova and Innova 2000 S · Ge Medical Systems, Inc. · 2002 · recalledK031637 · Innova 3100 · GE Medical Systems · 2003 · recalledK050489 · Innova 2100 · Ge Medical Systems, LLC · 2005 · recalled×2: 2 generations cleared close together, K052412 (2005), K060259 (2006) 1990200020102020
  1. Root · 7 generations backK890348 LC, Cardiac Positioner #B5078AGeneral Electric Co. · 1989 · Recalled
  2. 6 generations backK993037 LCV+ Version 2Ge Medical Systems, Inc. · 2000 · Recalled
  3. 5 generations backK022322 Innova and Innova 2000 SGe Medical Systems, Inc. · 2002 · Recalled
  4. 4 generations backK031637 Innova 3100GE Medical Systems · 2003 · Recalled
  5. 3 generations backK050489 Innova 2100Ge Medical Systems, LLC · 2005 · Recalled
  6. 2 generations backK052412 Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova…GE Healthcare · 2005 · Recalled
  7. 1 generation backK060259 Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC · 2006 · Recalled
  8. This deviceK061163 Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova…Ge Medical Systems, LLC · 2006
  9. Descendant · cites it directly · 11 of its ownK091658 GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models…GE Healthcare · 2009 · Recalled
  10. Descendant · 2 generations on · 10 of its ownK092439 Canon Digital Fluoroscopy and Radiography, Model Cxdi-50RFVirtual Imaging, Inc. · 2009
  11. Descendant · cites it directly · 8 of its ownK111209 Integrated Innova - S5I System OptionGE Healthcare · 2011 · Recalled
  12. Descendant · 2 generations on · 6 of its ownK113403 GE Discovery Igs Angiographic,Fluoroscopic X-Ray SystemGE Medical Systems SCS · 2012 · Recalled
  13. Descendant · 2 generations on · 6 of its ownK113034 GE Innova Solid State Xray Imager Flurosopic X Ray SystemGE Healthcare · 2011

Showing 5 of 22 descendants. See all descendants

Descendants cleared per year

2009: 32010: 22011: 22012: 22013: 12014: 12015: 12017: 12018: 12020: 12021: 12022: 12023: 12024: 12025: 22026: 1
2009high 3 (2009) · low 0 (2016)2026

Most influential descendants

Showing 5 of 22 · See all
  • K091658GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models Innova 4100, 3100…GE Healthcare · 2009 · cites it directly11descendants
  • K092439Canon Digital Fluoroscopy and Radiography, Model Cxdi-50RFVirtual Imaging, Inc. · 2009 · 2 generations on10descendants
  • K111209Integrated Innova - S5I System OptionGE Healthcare · 2011 · cites it directly8descendants
  • K113403GE Discovery Igs Angiographic,Fluoroscopic X-Ray SystemGE Medical Systems SCS · 2012 · 2 generations on6descendants
  • K113034GE Innova Solid State Xray Imager Flurosopic X Ray SystemGE Healthcare · 2011 · 2 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K890348LC, Cardiac Positioner #B5078AGeneral Electric Co.19895
−6K993037LCV+ Version 2Ge Medical Systems, Inc.20006
−5K022322Innova and Innova 2000 SGe Medical Systems, Inc.20027
−4K031637Innova 3100GE Medical Systems200325
−3K050489Innova 2100Ge Medical Systems, LLC200527
−2K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQGE Healthcare200550
−1K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121Ge Medical Systems, LLC20064
0 (this device)K061163Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova…Ge Medical Systems, LLC2006—
+1K091658GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models Innova…GE Healthcare20091
+2K092439Canon Digital Fluoroscopy and Radiography, Model Cxdi-50RFVirtual Imaging, Inc.2009—
+1K111209Integrated Innova - S5I System OptionGE Healthcare20112
+2K113403GE Discovery Igs Angiographic,Fluoroscopic X-Ray SystemGE Medical Systems SCS20124
+2K113034GE Innova Solid State Xray Imager Flurosopic X Ray SystemGE Healthcare2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199530
199820
200110
200220
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200930
201020
201120
201220
201310
201410
201510
201710
201810
202010
202110
202210
202310
202410
202520
202610
FIG. 1 — Lineage of K061163, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.2 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060259Digital Fluoroscopic Imaging Systems - Innova 3131 & Innova 2121
Summary page 1
…re substantlally eguuvalent to the currently marketed GE Healthcare Innova 2121'“, Innova 3131 , Innova 4100blnnova 4100 Innova 3100, Innova 31009, Innova 2100" devices (K060259 K033244, K052412, K031 637 K050489) This opinion is based on the information contained in the comparison table and the product d…
Ge Medical Systems, LLC20064
*K033244Digital Fluoroscopic Imaging System, Model Innova 4100 with Tilt Table Option
Summary page 1
…antlally eguuvalent to the currently marketed GE Healthcare Innova 2121'“, Innova 3131 , Innova 4100blnnova 4100 Innova 3100, Innova 31009, Innova 2100" devices (K060259 K033244, K052412, K031 637 K050489) This opinion is based on the information contained in the comparison table and the product data shee…
Ge Medical Systems, Inc.20038
*K052412Innova 2100-IQ, Innova 3100, Innova 3100-IQ, Innova 4100, and Innova 4100-IQ
Summary page 1
…eguuvalent to the currently marketed GE Healthcare Innova 2121'“, Innova 3131 , Innova 4100blnnova 4100 Innova 3100, Innova 31009, Innova 2100" devices (K060259 K033244, K052412, K031 637 K050489) This opinion is based on the information contained in the comparison table and the product data sheets. The…
GE Healthcare200550
*K031637Innova 3100
Summary page 1
…t to the currently marketed GE Healthcare Innova 2121'“, Innova 3131 , Innova 4100blnnova 4100 Innova 3100, Innova 31009, Innova 2100" devices (K060259 K033244, K052412, K031 637 K050489) This opinion is based on the information contained in the comparison table and the product data sheets. The Innova IVU…
GE Medical Systems200325
*K050489Innova 2100
Summary page 1
…currently marketed GE Healthcare Innova 2121'“, Innova 3131 , Innova 4100blnnova 4100 Innova 3100, Innova 31009, Innova 2100" devices (K060259 K033244, K052412, K031 637 K050489) This opinion is based on the information contained in the comparison table and the product data sheets. The Innova IVUS is off…
Ge Medical Systems, LLC200527

Also cleared against K060259: 3 other devices →

What descends from it? (22)

See all 22

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091658GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Models Innova 4100, 3100…GE Healthcare200911
K111209Integrated Innova - S5I System OptionGE Healthcare20118
K092004GE Innova Solid State Xray-Imager Fluoroscopic X-Ray System, Model: 4100IQ, 3100IQ, 2100IQGE Healthcare20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 16, 2012Class IIZ-2376-2012Software designGE Healthcare has recently become aware of a potential safety issue with respect to the Innova 2121IQ & Innova 3131IQ Systems. While performing a fluoroscopic acquisition with the Innova 2121IQ or Innova 3131IQ systems, there is a…
Dec 20, 2011Class IIZ-1318-2013Nonconforming Material/ComponentWhile performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries…
Aug 24, 2011Not classifiedZ-0133-2012Nonconforming Material/ComponentA potential weakness on the monitor suspension end-stop device may impact user safety. The end-stops may break if the monitor-suspension is repeatedly pushed with excessive force on the end rails.
May 16, 2011Class IIZ-1616-2013Software designGE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.
Feb 4, 2011Class IIZ-0934-2013Nonconforming Material/ComponentExposed high voltage wires inside the cabinet . On a cable located in the system cabinet, two of the shrink tubings covering the 400V live cable leave an area potentially unprotected.
Mar 24, 2010Not classifiedZ-1661-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Mar 24, 2010Not classifiedZ-1659-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Mar 24, 2010Not classifiedZ-1660-2010Software designA potential software failure related to the Worklist functionality has been identified on a subset of Innova 2100IQ, 3100IQ and 4100IQ systems. When using the Worklist function to import patient data to Innova, there are two scenarios…
Dec 22, 2009Not classifiedZ-0618-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0619-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0621-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Dec 22, 2009Not classifiedZ-0617-2010Device DesignPotential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational…
Aug 5, 2009Not classifiedZ-0410-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0409-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0407-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Aug 5, 2009Not classifiedZ-0408-2011Nonconforming Material/ComponentGE Healthcare has become aware of a potential issue regarding Innova systems with GE Fluoro UPS 20KVA that may impact patient safety. GE Fluoro UPS 20KVA (Uninterrupted Power Supply) is used to maintain fluoroscopy availability for…
Apr 16, 2009Not classifiedZ-0135-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0137-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0138-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Apr 16, 2009Not classifiedZ-0136-2010Device DesignPotential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room…
Nov 21, 2008Not classifiedZ-1195-2009Device DesignGE Healthcare has become aware of a potential image storage failure on Innova 2121 IQ/ 3131 IQ cardiovascular imaging systems that may impact patient safety. This intermittent communication failure on the image storage chain may result in…
Oct 9, 2008Not classifiedZ-1691-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1690-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1688-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1689-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Oct 9, 2008Not classifiedZ-1687-2009Radiation Control for Health and Safety ActGE Healthcare has become aware of an Air Kerma Rate issue associated with a large Equivalent Patient Thickness (EPT) on the Innova system that my impact patient safety. During a Maximum Air Kerma Rate (AKR) calibration procedure check, GE…
Jul 25, 2008Not classifiedZ-2438-2008Device DesignThe charger and / or battery of the PDB may fail earlier than expected with no advance warning and thus cause the system to shut down.
Jun 24, 2008Not classifiedZ-2360-2008Employee errorGE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two…
Jun 17, 2008Not classifiedZ-1475-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1476-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jun 17, 2008Not classifiedZ-1474-2009Process controlGE Healthcare has become aware that a few customers have experienced an unusual occurrence of system lockups on Innova single plane cardiovascular X-ray imaging systems while using the image auto-archiving function. when occurring during…
Jan 30, 2008Not classifiedZ-0473-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0475-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 30, 2008Not classifiedZ-0474-2009Software designSoftware error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and…
Jan 14, 2008Not classifiedZ-2041-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2043-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2039-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system
Jan 14, 2008Not classifiedZ-2040-2008Process controlRisk of sudden table drop on Omega IV and Omega V tables on GE Healthcare Advantx-E as well as on Innova 2000, 2100IQ, 3100/3100 IQ and 4100/4100IQ cardiovascular X-ray system

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 10 months on median.
  • 13 of the 17 ancestors have a recall of any class.
  • Of its descendants, K091658 (GE Healthcare, 2009) has gone furthest, with 11 descendants of its own.
  • 3 other devices also cite its predicate K060259.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are…

Read the full statement (138 words)

The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.

They are intended to replace fluoroscopic images obtained through image intensifier technology. Those devices are not intended for mammography applications.

Innova IVUS option:

The Innova IVUS software option simplifies the clinical workflow associated with the use of Volcano IVUS products by:

(1) automatically synchronizing the patient demographic information (patient name, date of birth, DICOM attributes etc.) from Innova system with an IVUS imaging system,

(2) providing a remote access to commonly used IVUS functions from the Innova table side user interface.

(3) displaying the IVUS images on the multi-monitor display of the Innova cathlab system.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K061163: Innova 2121IQ, Innova 3131IQ, Innova 4100, Innova 4100IQ, Innova 3100, Innova 3100IQ, Innova 2100IQ with Innova Ivus Opt.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K061163

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026