Recalled · Class II · 2015Model TSX-101A/H/I Aquilion 32/64 SP CT System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Toshiba America Medical Systems, In.C · JAK System, X-Ray, Tomography, Computed
Cleared on Mar 14, 2008 as substantially equivalent to K051833 (Toshiba America Medical Systems, In.C, 2005) after a 45-day review. Its family tree reaches back 3 generations to 2003. 17 later devices descend from it. It has been recalled 8 times.
- Decided
- Mar 14, 2008
- Review
- 45 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 8
- Predicates
- 1
Its longest line of descent runs back 3 generations to K031469 (2003). 17 later devices descend from it, 2 citing it directly.
- Root · 3 generations backK031469 Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Toshiba America Medical Systems, In.C · 2003 · Recalled
- 2 generations backK033418 TSX-101A/a and TSX-101A/C Aquilion Multislice CT ScannersToshiba America Medical Systems, In.C · 2003 · Recalled
- 1 generation backK051833 TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT…Toshiba America Medical Systems, In.C · 2005
- This deviceK080211 Model TSX-101A/H/I Aquilion 32/64 SP CT SystemToshiba America Medical Systems, In.C · 2008
- Descendant · cites it directly · 15 of its ownK101888 Scenaria Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2011 · Recalled
- Descendant · 2 generations on · 14 of its ownK123509 Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2013
- Descendant · 3 generations on · 11 of its ownK161748 SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc. · 2016
- Descendant · 4 generations on · 10 of its ownK163528 HITACHI Supria Whole-body X-ray CT System Phase 3Hitachi Medical Systems America, Inc. · 2017
- Descendant · 5 generations on · 9 of its ownK171738 Supria True64Hitachi Healthcare Americas Corporation · 2017
Showing 5 of 17 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 17 · See all- K101888Scenaria Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2011 · cites it directly15descendants
- K123509Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2013 · 2 generations on14descendants
- K161748SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc. · 2016 · 3 generations on11descendants
- K163528HITACHI Supria Whole-body X-ray CT System Phase 3Hitachi Medical Systems America, Inc. · 2017 · 4 generations on10descendants
- K171738Supria True64Hitachi Healthcare Americas Corporation · 2017 · 5 generations on9descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K031469 | Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7 | Toshiba America Medical Systems, In.C | 2003 | 1 |
| −2 | K033418 | TSX-101A/a and TSX-101A/C Aquilion Multislice CT Scanners | Toshiba America Medical Systems, In.C | 2003 | 2 |
| −1 | K051833 | TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT Scanner | Toshiba America Medical Systems, In.C | 2005 | — |
| 0 (this device) | K080211 | Model TSX-101A/H/I Aquilion 32/64 SP CT System | Toshiba America Medical Systems, In.C | 2008 | — |
| +1 | K101888 | Scenaria Whole-Body X-Ray CT System | Hitachi Medical Systems America, Inc. | 2011 | 3 |
| +2 | K123509 | Scenaria Phase 2 Whole-Body X-Ray CT System | Hitachi Medical Systems America, Inc. | 2013 | — |
| +3 | K161748 | SUPRIA w/guideShot Option | Hitachi Medical Systems America, Inc. | 2016 | — |
| +4 | K163528 | HITACHI Supria Whole-body X-ray CT System Phase 3 | Hitachi Medical Systems America, Inc. | 2017 | — |
| +5 | K171738 | Supria True64 | Hitachi Healthcare Americas Corporation | 2017 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2015 | 2 | 0 |
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2019 | 2 | 0 |
| 2020 | 1 | 0 |
| 2022 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K051833 | TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT ScannerSummary page 2…ndards. Substantial Equivalence: This device is similar in materials and processes to that of the predicate device, Toshiba TSX-101AfH, /I Aquilion 32/64 SP CT scanner [k051833]. 510(k) Summary Statement. 2 of 2 === Page 3… | 2005 |
What descends from it? (17)
See all 172 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 8, 2015 | Class II | Z-0013-2016 | Radiation Control for Health and Safety Act | If two specific operations are performed in multi-phase helical scanning, the acquired raw data may not be saved. |
| Oct 1, 2013 | Class II | Z-1113-2014 | Software design | Toshiba intitiated this recall because their investigation revealed that due to a problem of the X-ray output control software, the X-ray output conditions may not be set correctly. As a result, X-ray exposure under the scan conditions… |
| Dec 8, 2010 | Not classified | Z-3221-2011 | Software design | There is the potential that the current may not be controlled correctly in examinations using the ECG modulation function, which adjusts the x-ray output in synchronization with the ECG waveform. |
| Apr 9, 2010 | Not classified | Z-2204-2011 | Software design | The recall was initiated by Toshiba America Medical Systems (TAMS) because the software associated with TSX-101A: Aquilion 64 can prevent scanning from being performed at the specified helical pitch (couch movement speed). |
| Jan 8, 2009 | Not classified | Z-1168-2010 | Component design/selection | Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large… |
| Jan 8, 2009 | Not classified | Z-1166-2010 | Component design/selection | Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large… |
| Jan 8, 2009 | Not classified | Z-1167-2010 | Component design/selection | Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large… |
| Jan 8, 2009 | Not classified | Z-1169-2010 | Component design/selection | Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large… |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 1.7 years on median.
- 2 of the 3 ancestors have a recall of any class.
- Of its descendants, K101888 (Hitachi Medical Systems America, Inc., 2011) has gone furthest, with 15 descendants of its own.
- 2 other devices also cite its predicate K051833.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
Acquisition and display of axial x-ray images of the whole body to include the head.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080211 · printed Nov 6, 2026
Cite this page
510(k) Report. “K080211: Model TSX-101A/H/I Aquilion 32/64 SP CT System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K080211
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026