Scenaria Whole-Body X-Ray CT System

K101888510(k) TraditionalCleared Apr 8, 2011275-day reviewClass II21 CFR 892.1750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hitachi Medical Systems America, Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Apr 8, 2011 as substantially equivalent to K080211 (Toshiba America Medical Systems, In.C, 2008) after a 275-day review. Its family tree reaches back 4 generations to 2003. 15 later devices descend from it. It has been recalled 3 times.

Decided
Apr 8, 2011
Review
275 days
Class
Class II
Product code
JAK
Recalls
3
Predicates
1

Its longest line of descent runs back 4 generations to K031469 (2003). 15 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2013: 1 descendant cleared2016: 1 descendant cleared2015: 2 descendants cleared2019: 2 descendants cleared2017: 2 descendants cleared2020: 1 descendant cleared2022: 1 descendant cleared2023: 1 descendant cleared2025: 1 descendant cleared2024: 1 descendant cleared2026: 2 descendants clearedK031469 · Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7 · Toshiba America Medical Systems, In.C · 2003 · recalledK033418 · TSX-101A/a and TSX-101A/C Aquilion Multislice CT Scanners · Toshiba America Medical Systems, In.C · 2003 · recalledK051833 · TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT Scanner · Toshiba America Medical Systems, In.C · 2005K080211 · Model TSX-101A/H/I Aquilion 32/64 SP CT System · Toshiba America Medical Systems, In.C · 2008 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant7 generations forward: 2 descendants8 generations forward: 2 descendants K080211 · Model TSX-101A/H/I Aquilion 32/64 SP CT System · Toshiba America Medical Systems, In.C · 2008 · recalledK051833 · TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT Scanner · Toshiba America Medical Systems, In.C · 2005K033418 · TSX-101A/a and TSX-101A/C Aquilion Multislice CT Scanners · Toshiba America Medical Systems, In.C · 2003 · recalledK031469 · Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7 · Toshiba America Medical Systems, In.C · 2003 · recalled0+5
  1. Root · 4 generations backK031469 Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Toshiba America Medical Systems, In.C · 2003 · Recalled
  2. 3 generations backK033418 TSX-101A/a and TSX-101A/C Aquilion Multislice CT ScannersToshiba America Medical Systems, In.C · 2003 · Recalled
  3. 2 generations backK051833 TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT…Toshiba America Medical Systems, In.C · 2005
  4. 1 generation backK080211 Model TSX-101A/H/I Aquilion 32/64 SP CT SystemToshiba America Medical Systems, In.C · 2008 · Recalled
  5. This deviceK101888 Scenaria Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2011
  6. Descendant · cites it directly · 14 of its ownK123509 Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2013
  7. Descendant · 2 generations on · 11 of its ownK161748 SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc. · 2016
  8. Descendant · 3 generations on · 10 of its ownK163528 HITACHI Supria Whole-body X-ray CT System Phase 3Hitachi Medical Systems America, Inc. · 2017
  9. Descendant · 4 generations on · 9 of its ownK171738 Supria True64Hitachi Healthcare Americas Corporation · 2017
  10. Descendant · 2 generations on · 8 of its ownK150595 SCENARIA Phase 3 Whole-body X-ray CT SystemHitachi Medical Systems America, Inc. · 2015 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2013: 12015: 22016: 12017: 22019: 22020: 12022: 12023: 12024: 12025: 12026: 2
2013high 2 (2015) · low 0 (2014)2026

Most influential descendants

Showing 5 of 15 · See all
  • K123509Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2013 · cites it directly14descendants
  • K161748SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc. · 2016 · 2 generations on11descendants
  • K163528HITACHI Supria Whole-body X-ray CT System Phase 3Hitachi Medical Systems America, Inc. · 2017 · 3 generations on10descendants
  • K171738Supria True64Hitachi Healthcare Americas Corporation · 2017 · 4 generations on9descendants
  • K150595SCENARIA Phase 3 Whole-body X-ray CT SystemHitachi Medical Systems America, Inc. · 2015 · 2 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K031469Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Toshiba America Medical Systems, In.C20031
−3K033418TSX-101A/a and TSX-101A/C Aquilion Multislice CT ScannersToshiba America Medical Systems, In.C20032
−2K051833TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT ScannerToshiba America Medical Systems, In.C2005—
−1K080211Model TSX-101A/H/I Aquilion 32/64 SP CT SystemToshiba America Medical Systems, In.C20088
0 (this device)K101888Scenaria Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc.2011—
+1K123509Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc.2013—
+2K161748SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc.2016—
+3K163528HITACHI Supria Whole-body X-ray CT System Phase 3Hitachi Medical Systems America, Inc.2017—
+4K171738Supria True64Hitachi Healthcare Americas Corporation2017—
+2K150595SCENARIA Phase 3 Whole-body X-ray CT SystemHitachi Medical Systems America, Inc.20152

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201520
201610
201720
201920
202010
202210
202310
202410
202510
202620
FIG. 1 — Lineage of K101888, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080211Model TSX-101A/H/I Aquilion 32/64 SP CT System
Summary page 2
…892.1750 Regulatory Class Class 11 Predicate Device(s): TOSHIBA AQUILION 64 Slice System (K08021 1) Reason for Submission New Device Device Intended Use The SCENARIA system is indicated to acquire axial…
Toshiba America Medical Systems, In.C20088

Also cleared against K080211: 1 other device →

What descends from it? (15)

See all 15

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123509Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc.201314

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 16, 2019Class IIZ-0271-2020Nonconforming Material/ComponentThere is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
Jun 30, 2015Class IIZ-2392-2016Component design/selectionThe firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
May 4, 2012Class IIZ-2037-2012Software designHitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 2.7 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K123509 (Hitachi Medical Systems America, Inc., 2013) has gone furthest, with 14 descendants of its own.
  • 1 other device also cites its predicate K080211.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Scenadia system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes.

The volume datasets acquired by the Scenadia can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Scenaria can be transferred to external devices via a DICOM standard interface.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K101888: Scenaria Whole-Body X-Ray CT System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101888

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026