Gambro Prismaflex HF1000 and HF1400 Sets

K042938510(k) SpecialCleared Jan 6, 200573-day reviewClass II21 CFR 876.5860Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gambro Renal Products · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Jan 6, 2005 as substantially equivalent to K011221 (Gambro Renal Products, 2001) and 1 other predicate after a 73-day review. Its family tree reaches back 6 generations to 1991. 1 of the 29 devices in its family, counting itself, has a Class I recall. 1 later device descends from it. It has been recalled 7 times.

Decided
Jan 6, 2005
Review
73 days
Class
Class II
Product code
KDI
Recalls
7
Predicates
2

Its longest line of descent runs back 6 generations to K910215 (1991). 27 ancestors in all, the oldest K850569 (1985). 1 later device descends from it, 1 citing it directly. 1 of the 29 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2003: 1 other ancestor2002: 1 other ancestor1998: 4 other ancestors, 1 with a Class I recall11999: 1 other ancestor1997: 3 other ancestors1991: 2 other ancestors1988: 1 other ancestor1985: 2 other ancestors1994: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor1987: 1 other ancestor2022: 1 descendant clearedK910215 · Drake Willock System 1000 Dialysate Delivery Syst · Cd Medical, Inc. · 1991 · recalledK954987 · Altratouch 1000 Hemodialysis Delivery System · Althin Medical AB an Affiliate of Baxter Intl · 1996 · recalledK970446 · Drake Willock System & Altra Touch 1000 Dialysis Delivery System · Althin Medical AB an Affiliate of Baxter Intl · 1997 · recalledK993064 · Prisma CFM (Continuous Fluid Management System) · Gambro Renal Care Products · 1999 · recalledK021615 · Baxter Accura System, Model 5M5660 · Baxter Healthcare Corp · 2002 · recalledK041005 · Gambro Prismaflex and Gambro Prismaflex M60 & M100 Sets · Gambro Renal Products · 2004 · recalled1990200020102020
  1. Root · 6 generations backK910215 Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc. · 1991 · Recalled
  2. 5 generations backK954987 Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1996 · Recalled
  3. 4 generations backK970446 Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl · 1997 · Recalled
  4. 3 generations backK993064 Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products · 1999 · Recalled
  5. 2 generations backK021615 Baxter Accura System, Model 5M5660Baxter Healthcare Corp · 2002 · Recalled
  6. 1 generation backK041005 Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products · 2004 · Recalled
  7. This deviceK042938 Gambro Prismaflex HF1000 and HF1400 SetsGambro Renal Products · 2005
  8. Descendant · cites it directly · 0 of its ownK220281 multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC · 2022

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K220281multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K910215Drake Willock System 1000 Dialysate Delivery SystCd Medical, Inc.19915
−5K954987Altratouch 1000 Hemodialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl19961
−4K970446Drake Willock System & Altra Touch 1000 Dialysis Delivery SystemAlthin Medical AB an Affiliate of Baxter Intl19975
−3K993064Prisma CFM (Continuous Fluid Management System)Gambro Renal Care Products19991
−2K021615Baxter Accura System, Model 5M5660Baxter Healthcare Corp20023
−1K041005Gambro Prismaflex and Gambro Prismaflex M60 & M100 SetsGambro Renal Products20047
0 (this device)K042938Gambro Prismaflex HF1000 and HF1400 SetsGambro Renal Products2005—
+1K220281multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198520
198710
198810
199120
199310
199410
199510
199610
199730
199841
199910
200110
200210
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K042938, 1991–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011221Prisma HF 1000 Set, Prisma HF 1000 Pre-Set
Summary page 1
…enterology - Urology CLASSIFICATION: Class II per 21 CFR 876.5860 PREDICATE DEVICES: Gambro Prisma HF1000 Set K011221 Gambro Prismaflex M60/M100 Set K041005 SUBSTANTIAL EQUIVALENCE: The proposed…
Gambro Renal Products20012
*K041005Gambro Prismaflex and Gambro Prismaflex M60 & M100 Sets
Summary page 1
…EDICATE DEVICES: Gambro Prisma HF1000 Set K011221 Gambro Prismaflex M60/M100 Set K041005 SUBSTANTIAL EQUIVALENCE: The proposed Prismaflex HF 1000 and HF 1400 sets are substantially equivalent to the Prisma HF 1000 se…
Gambro Renal Products20047

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220281multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC20220

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 6, 2026Class IIZ-1229-2026Component design/selectionThere is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Jan 6, 2026Class IIZ-1230-2026Component design/selectionThere is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Sep 14, 2022Class IIZ-0083-2023Error in labelingWithin the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Sep 14, 2022Class IIZ-0085-2023Error in labelingWithin the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.
Jul 22, 2021Class IIZ-2379-2021Vendor change controlProducts labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming
Jul 18, 2014Class IIZ-2204-2014Device DesignGambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.
Jul 18, 2014Class IIZ-2205-2014Device DesignGambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K963440Dialog Dialysis MachineB. Braun Medical AG · 1998 · 2 generations back · ◆ in the 1998 ancestors bar of FIG. 16 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 5 companies; a new generation was cleared every 2.3 years on median.
  • 13 of the 27 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K220281 (Fresenius Medical Care Renal Therapies Group, LLC, 2022) has gone furthest, with 0 descendants of its own.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The Prismaflex Set is indicated for use only with the Prismaflex Control Unit in providing continuous fluid management and renal replacement therapies. The system is intended for patients who have acute renal failure, fluid overload, or both.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042938: Gambro Prismaflex HF1000 and HF1400 Sets.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042938

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026