510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Griffin Spect/CT Imaging System

K041218510(k) TraditionalCleared May 24, 200414-day reviewClass II21 CFR 892.1200Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Adac Laboratories · KPS System, Tomography, Computed, Emission

Cleared on May 24, 2004 as substantially equivalent to K031705 (Adac Laboratories, 2003) and 1 other predicate after a 14-day review. Its family tree reaches back 4 generations to 1993. 1 later device descends from it. It has been recalled 11 times.

Decided
May 24, 2004
Review
14 days
Class
Class II
Product code
KPS
Recalls
11
Predicates
2

Its longest line of descent runs back 4 generations to K930090 (1993). 12 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2001: 2 other ancestors2000: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2005: 1 descendant clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 4 other ancestors4 generations back: 1 other ancestor 1 generation forward: 1 descendant K012009 · MX8000 Idt CT System · Marconi Medical Systems, Inc. · 2001 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 19930
  1. Root · 4 generations backK930090 HelicatElscint, Inc. · 1993
  2. 3 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 2 generations backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. 1 generation backK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001 · Recalled
  5. This deviceK041218 Griffin Spect/CT Imaging SystemAdac Laboratories · 2004
  6. Descendant · cites it directly · 0 of its ownK051351 Precedence Spect/CT Imaging SystemAdac Laboratories · 2005 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K051351Precedence Spect/CT Imaging SystemAdac Laboratories · 2005 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K930090HelicatElscint, Inc.1993—
−3K961464Select/SPElscint, Inc.19963
−2K982060Volumax CT Imaging SystemElscint, Inc.1998—
−1K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161
0 (this device)K041218Griffin Spect/CT Imaging SystemAdac Laboratories2004—
+1K051351Precedence Spect/CT Imaging SystemAdac Laboratories20051

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200010
200120
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
FIG. 1 — Lineage of K041218, 1993–2005Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPS device

  1. #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031705Skylight Imaging System
Summary page 1
…uct Code: 90 KPS and 90 JAK C. Date prepared: April 26, 2004 D. Predicate Device: Skylight Imaging System (K031705) MX8000 IDT CT System (KO 12009) E. Performance Standards o 21 CFR 102…
Adac Laboratories2003—
*K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.200161

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K051351Precedence Spect/CT Imaging SystemAdac Laboratories20050

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 15, 2017Class IIZ-0513-2018Under Investigation by firmDuring collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart…
Dec 15, 2017Class IIZ-0514-2018Under Investigation by firmDuring collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart…
Mar 6, 2015Class IIZ-1778-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1779-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Sep 27, 2011Not classifiedZ-0255-2012Device DesignA mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
Sep 27, 2011Not classifiedZ-0257-2012Device DesignA mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
Sep 27, 2011Not classifiedZ-0256-2012Device DesignA mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
Sep 27, 2011Not classifiedZ-0258-2012Device DesignA mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
Dec 28, 2010Not classifiedZ-1549-2011OtherPhilips Healthcare Nuclear Medicine determined that the detector arm assembly could be compromised.
Dec 28, 2010Not classifiedZ-1550-2011OtherPhilips Healthcare Nuclear Medicine determined that the detector arm assembly could be compromised.
Sep 28, 2007Not classifiedZ-0676-2008Software designIncorrect Measurement: The firm discovered a software defect in Cardiac and Pulmonary Viewer software application options used in the Precedence CT Sub-System. The defect causes incorrect measurements in derived images which are zoomed…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 3.0 years on median.
  • 4 of the 12 ancestors have a recall of any class.
  • Of its descendants, K051351 (Adac Laboratories, 2005) has gone furthest, with 0 descendants of its own.

How does its category compare?

All KPS devices

KPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPS

12in 2025

2007: 82008: 172009: 92010: 32011: 72012: 62013: 112014: 112015: 32016: 112017: 112018: 132019: 32020: 122021: 152022: 52023: 122024: 112025: 122026 (partial year): 6
2007high 17 (2008) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPS

65in 2025

2007: 38 days median2008: 25 days median2009: 42 days median2010: 91 days median2011: 28 days median2012: 94 days median2013: 84 days median2014: 92 days median2015: 112 days median2016: 65 days median2017: 104 days median2018: 59 days median2019: 60 days median2020: 72 days median2021: 66 days median2022: 60 days median2023: 63 days median2024: 59 days median2025: 65 days median2026 (partial year): 67 days median
2007high 112 (2015) · low 25 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPS

16in 2025

2007: 222008: 162009: 142010: 42011: 342012: 132013: 132014: 182015: 52016: 162017: 112018: 12, 3 Class I2019: 5, 5 Class I2020: 82021: 42022: 8, 6 Class I2023: 6, 3 Class I2024: 152025: 162026 (partial year): 6
2007high 34 (2011) · low 4 (2010)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · KPS

36in 2025

2007: 382008: 112009: 152010: 442011: 562012: 342013: 2402014: 332015: 242016: 102017: 312018: 262019: 1052020: 222021: 102022: 152023: 222024: 232025: 36
2007high 240 (2013) · low 10 (2016)2025

MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078382238
200817251611
20099421415
2010391444
20117283456
20126941334
2013118413240
201411921833
20153112524
201611651610
2017111041131
2018135912 (3)26
20193605 (5)105
20201272822
20211566410
20225608 (6)15
202312636 (3)22
202411591523
202512651636
20266676—

What is it indicated for?

Griffin is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Griffin may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT…

Read the full statement (152 words)

Griffin is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Griffin may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data for anatomical localization of the nuclear medicine data. Griffin may be used either as a separate single photon system, a separate CT system or as a combined CT and single photon system. The nuclear medicine and CT images may be transferred to other systems such as a radiation therapy planning system. The Griffin Imaging System should only be used by trained healthcare professionals.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K041218: Griffin Spect/CT Imaging System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K041218

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026