510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

MX8000 Idt CT System

K012009510(k) TraditionalCleared Sep 13, 200178-day reviewClass II21 CFR 892.1750Recalls 61

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Marconi Medical Systems, Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Sep 13, 2001 as substantially equivalent to K982060 (Elscint, Inc., 1998) and 1 other predicate after a 78-day review. Its family tree reaches back 3 generations to 1993. 263 later devices descend from it. It has been recalled 61 times.

Decided
Sep 13, 2001
Review
78 days
Class
Class II
Product code
JAK
Recalls
61
Predicates
2

Its longest line of descent runs back 3 generations to K930090 (1993). 8 ancestors in all. 263 later devices descend from it, 11 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1997: 1 other ancestor1995: 1 other ancestor1994: 2 other ancestors2004: 7 descendants cleared2008: 7 descendants cleared2003: 3 descendants cleared2009: 4 descendants cleared2017: 12 descendants cleared2011: 4 descendants cleared2007: 3 descendants cleared2005: 3 descendants cleared2016: 5 descendants cleared2006: 3 descendants cleared2015: 3 descendants cleared2018: 14 descendants cleared2019: 9 descendants cleared2014: 9 descendants cleared2013: 6 descendants cleared2010: 2 descendants cleared2012: 6 descendants cleared2020: 22 descendants cleared2024: 19 descendants cleared2023: 20 descendants cleared2025: 35 descendants cleared2022: 22 descendants cleared2021: 19 descendants cleared2026: 26 descendants clearedK930090 · Helicat · Elscint, Inc. · 1993K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK982060 · Volumax CT Imaging System · Elscint, Inc. · 19981990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 11 descendants2 generations forward: 24 descendants3 generations forward: 43 descendants4 generations forward: 43 descendants5 generations forward: 29 descendants6 generations forward: 30 descendants7 generations forward: 26 descendants8 generations forward: 17 descendants9 generations forward: 13 descendants10 generations forward: 8 descendants11 generations forward: 10 descendants12 generations forward: 6 descendants13 generations forward: 2 descendants14 generations forward: 1 descendant K982060 · Volumax CT Imaging System · Elscint, Inc. · 1998K961464 · Select/SP · Elscint, Inc. · 1996 · recalledK930090 · Helicat · Elscint, Inc. · 19930+5+10
  1. Root · 3 generations backK930090 HelicatElscint, Inc. · 1993
  2. 2 generations backK961464 Select/SPElscint, Inc. · 1996 · Recalled
  3. 1 generation backK982060 Volumax CT Imaging SystemElscint, Inc. · 1998
  4. This deviceK012009 MX8000 Idt CT SystemMarconi Medical Systems, Inc. · 2001
  5. Descendant · cites it directly · 214 of its ownK033326 Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · Recalled
  6. Descendant · cites it directly · 147 of its ownK042293 Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc. · 2004 · Recalled
  7. Descendant · 2 generations on · 92 of its ownK130159 Qlab Quantification Software Heart ModelPhilips Health Care · 2013
  8. Descendant · 3 generations on · 91 of its ownK132165 Qlab Quantification SoftwarePhilips Ultrasound, Inc. · 2013 · Recalled
  9. Descendant · cites it directly · 87 of its ownK032036 Gemini 16Philips Medical Systems (Cleveland), Inc. · 2003 · Recalled

Showing 5 of 263 descendants. See all descendants

Descendants cleared per year

2003: 32004: 72005: 32006: 32007: 32008: 72009: 42010: 22011: 42012: 62013: 62014: 92015: 32016: 52017: 122018: 142019: 92020: 222021: 192022: 222023: 202024: 192025: 352026: 26
2003high 35 (2025) · low 2 (2010)2026

Most influential descendants

Showing 5 of 263 · See all
  • K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc. · 2003 · cites it directly214descendants
  • K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc. · 2004 · cites it directly147descendants
  • K130159Qlab Quantification Software Heart ModelPhilips Health Care · 2013 · 2 generations on92descendants
  • K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc. · 2013 · 3 generations on91descendants
  • K032036Gemini 16Philips Medical Systems (Cleveland), Inc. · 2003 · cites it directly87descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K930090HelicatElscint, Inc.1993—
−2K961464Select/SPElscint, Inc.19963
−1K982060Volumax CT Imaging SystemElscint, Inc.1998—
0 (this device)K012009MX8000 Idt CT SystemMarconi Medical Systems, Inc.2001—
+1K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003131
+1K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc.20041
+2K130159Qlab Quantification Software Heart ModelPhilips Health Care2013—
+3K132165Qlab Quantification SoftwarePhilips Ultrasound, Inc.20131
+1K032036Gemini 16Philips Medical Systems (Cleveland), Inc.20032

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199420
199510
199710
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200330
200470
200530
200630
200730
200870
200940
201020
201140
201260
201360
201490
201530
201650
2017120
2018140
201990
2020220
2021190
2022220
2023200
2024190
2025350
2026260
FIG. 1 — Lineage of K012009, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982060Volumax CT Imaging System
Summary page 2
…f comparable type and substantially equivalent to the legally marketed devices currently in commercial distribution, namely: e Mx8000 (initially identified as Volumax), K982060 e Mx8000 v5.0, K010817 ' 6. Safety and Effectiveness Considerations The safety of the device is assured by adherence to GMP pr…
Elscint, Inc.1998—
*K010817MX8000 V5.0 CT System
Summary page 2
…bstantially equivalent to the legally marketed devices currently in commercial distribution, namely: e Mx8000 (initially identified as Volumax), K982060 e Mx8000 v5.0, K010817 ' 6. Safety and Effectiveness Considerations The safety of the device is assured by adherence to GMP practices and to Internati…
Philips Medical Systems (Cleveland), Inc.200112

What descends from it? (263)

See all 263

11 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K033326Philips Plus Computed Tomography SystemPhilips Medical Systems, Inc.2003214
K042293Brilliance CT, Private Practice CV Configuration, CT ScannerPhilips Medical Systems (Cleveland), Inc.2004147
K032036Gemini 16Philips Medical Systems (Cleveland), Inc.200387
K041955Gemini Pet/CT Imaging SystemPhilips Medical Systems (Cleveland), Inc.200465
K033357Acqsim-Multislice-CT, Model 453567042271Philips Medical Systems (Cleveland), Inc.200410
K023785Lung Nodule Assessment and Comparison OptionPhilips Medical Systems (Cleveland), Inc.200310
K033677Brain Perfusion OptionPhilips Medical Systems (Cleveland), Inc.20043
K091195Philips MX 16 Slice CT System PHASE11, Model 9896 056 56511Philips Medical Systems (Cleveland), Inc.20091
K041218Griffin Spect/CT Imaging SystemAdac Laboratories20041
K090173Neuviz 16 Multi-Slice CT Scanner SystemNeusoft Medical Systems Co., Ltd.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 13, 2025Class IIZ-2596-2025Process controlThe patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
May 1, 2021Class IIZ-1981-2021Device DesignSignificant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining…
Oct 18, 2018Class IIZ-0548-2019Process controlThere is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Oct 18, 2018Class IIZ-0545-2019Process controlThere is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Oct 18, 2018Class IIZ-0547-2019Process controlThere is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Oct 18, 2018Class IIZ-0546-2019Process controlThere is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
Jul 3, 2018Class IIZ-1142-2019Device DesignPhilips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator…
Mar 30, 2018Class IIZ-2175-2018Device DesignThe patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent…
Mar 30, 2018Class IIZ-2176-2018Device DesignThe patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent…
Feb 8, 2018Class IIZ-1412-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Feb 8, 2018Class IIZ-1415-2018Software designClosing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
Jul 20, 2016Class IIZ-2656-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Jul 20, 2016Class IIZ-2655-2016Software Design ChangeAfter upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at…
Mar 24, 2016Class IIZ-2370-2016Software change controlSoftware defect causing intermittently slow response of Host.
Mar 24, 2016Class IIZ-2369-2016Software change controlSoftware defect causing intermittently slow response of Host.
Mar 4, 2016Class IIZ-1506-2016Software designThe AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data…
Sep 14, 2015Class IIZ-0173-2016Software designThe firm discovered Ring/Dot artifact due to X-ray measurement error.
Mar 16, 2015Class IIZ-1651-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 16, 2015Class IIZ-1652-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 16, 2015Class IIZ-1653-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 16, 2015Class IIZ-1654-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 16, 2015Class IIZ-1655-2015Employee errorThe firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
Mar 6, 2015Class IIZ-1766-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1765-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Mar 6, 2015Class IIZ-1793-2015Nonconforming Material/ComponentThe firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
Dec 8, 2014Class IIZ-0979-2015Process controlImproper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
Jul 27, 2014Class IIZ-2237-2014Software designThe following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction…
Apr 1, 2014Class IIZ-1722-2014Component design/selectionA customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
Mar 27, 2014Class IIZ-1461-2014Nonconforming Material/ComponentBlock assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can…
Mar 11, 2014Class IIZ-1678-2014Software designDuring SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in…
Aug 30, 2013Class IIZ-0013-2014Software designThere are artifacts appearing on the Tracker images during threshold-triggered bolus scans with protocols using a Rotation Time of 0.4 seconds.
Jul 5, 2013Class IIZ-2210-2013Component design/selectionThe patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
Mar 12, 2013Class IIZ-1338-2013Use errorPhilips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system…
Dec 20, 2012Class IIZ-0945-2013Component design/selectionPhilips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
Oct 24, 2012Class IIZ-0881-2013Software designSoftware anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was…
Aug 28, 2012Class IIIZ-2404-2012Software change controlPhilips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button.
Aug 21, 2012Class IIZ-2451-2012Nonconforming Material/ComponentA bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
Aug 3, 2012Class IIZ-2245-2012Device DesignThe patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.
Jul 2, 2012Class IIZ-2035-2012Process controlPhilips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
May 23, 2012Class IIZ-1988-2012Software designPhilips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a…
May 23, 2012Class IIZ-1987-2012Software designThe Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.
May 22, 2012Class IIZ-2000-2012Software designIn the Multi-Modality Tumor Tracking application, when changing the contours of the lesion using the editing mode "edit contour" tool, the measurements are not recalculated causing discrepancies with the lesion measurements.
Mar 26, 2012Not classifiedZ-1693-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Mar 26, 2012Not classifiedZ-1692-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Mar 26, 2012Not classifiedZ-1691-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Mar 26, 2012Not classifiedZ-1690-2012Nonconforming Material/ComponentComponent defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems…
Sep 15, 2011Not classifiedZ-0182-2012Software designA problem related to a software error has been detected in the Philips Multiple Gated Radionuclide Angiocardiography (MUGA) application. A false negative interpretation due to a falsely elevated ejection fraction or a false positive…
Dec 10, 2010Not classifiedZ-1669-2011OtherPhilips Medical Systems has decided to recall the patient support vertical brake hub in all its Brilliance systems due to the collapse in the vertical direction. The short key connecting the vertical brake hub to the motor shaft can…
Jul 27, 2010Not classifiedZ-0133-2011Software change controlPhilips Healthcare has decided to recall the Computed Tomography X-Ray System MX 8000 IDT due to the fact that the software upgrade for the MX 8000 IDT 16 v4.2.1 may not provide the optional collimation of 4x4.5 (18mm). If the customer…
Oct 3, 2009Not classifiedZ-1819-2011OtherDuring an FDA inspection on 10/1/2010 the firm was notified by FDA that their field action they initiated in Oct 2009, constituted a recall. The firm determined that the functional analysis feature in the Comprehensive Cardiac Analysis…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 2 companies; a new generation was cleared every 3.0 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • Of its descendants, K033326 (Philips Medical Systems, Inc., 2003) has gone furthest, with 214 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Mx8000 IDT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K012009: MX8000 Idt CT System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K012009

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026