510(k) · Cleared
Recalled · Class II · 2018Somatom Spirit
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Systems, Inc. · JAK System, X-Ray, Tomography, Computed
Cleared on Sep 14, 2004 as substantially equivalent to K003014 (Siemens Medical Solutions USA, Inc., 2000) and 1 other predicate after an 18-day review. Its family tree reaches back 2 generations to 1994. 30 later devices descend from it. It has been recalled 2 times.
- Decided
- Sep 14, 2004
- Review
- 18 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 2
- Predicates
- 2
Its longest line of descent runs back 2 generations to K940606 (1994). 5 ancestors in all. 30 later devices descend from it, 1 citing it directly.
- Root · 2 generations backK940606 Hispeed Advantage IIGE Medical Systems · 1994
- 1 generation backK991716 Hispeed NX/I CT Scanner SystemGe Medical Systems, Inc. · 1999 · Recalled
- This deviceK042328 Somatom SpiritSiemens Medical Systems, Inc. · 2004
- Descendant · cites it directly · 29 of its ownK140912 Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc. · 2014 · Recalled
- Descendant · 2 generations on · 27 of its ownK142955 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge…Siemens Medical Solutions USA, Inc. · 2015 · Recalled
- Descendant · 3 generations on · 24 of its ownK161196 SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · Recalled
- Descendant · 3 generations on · 21 of its ownK162302 Somatom Confidence ® RT ProSiemens Medical Solutions USA, Inc. · 2016 · Recalled
- Descendant · 3 generations on · 20 of its ownK173630 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM…Siemens Medical Solutions USA, Inc. · 2018 · Recalled
Showing 5 of 30 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 30 · See all- K140912Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc. · 2014 · cites it directly29descendants
- K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition…Siemens Medical Solutions USA, Inc. · 2015 · 2 generations on27descendants
- K161196SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · 3 generations on24descendants
- K162302Somatom Confidence ® RT ProSiemens Medical Solutions USA, Inc. · 2016 · 3 generations on21descendants
- K173630SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc. · 2018 · 3 generations on20descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K940606 | Hispeed Advantage II | GE Medical Systems | 1994 | — |
| −1 | K991716 | Hispeed NX/I CT Scanner System | Ge Medical Systems, Inc. | 1999 | 2 |
| 0 (this device) | K042328 | Somatom Spirit | Siemens Medical Systems, Inc. | 2004 | — |
| +1 | K140912 | Somatom Scope/Somatom Scope Power | Siemens Medical Solutions USA, Inc. | 2014 | 2 |
| +2 | K142955 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM… | Siemens Medical Solutions USA, Inc. | 2015 | 16 |
| +3 | K161196 | SOMATOM Drive | Siemens Medical Solutions USA, Inc. | 2016 | 4 |
| +3 | K162302 | Somatom Confidence ® RT Pro | Siemens Medical Solutions USA, Inc. | 2016 | 1 |
| +3 | K173630 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition… | Siemens Medical Solutions USA, Inc. | 2018 | 4 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 2 | 0 |
| 2000 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 3 | 0 |
| 2017 | 3 | 0 |
| 2018 | 2 | 0 |
| 2019 | 3 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 3 | 0 |
| 2024 | 3 | 0 |
| 2025 | 4 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K003014 | Somatom Emotion MSSummary page 2…equivalent to the following systems: Product Name 510(k) Clearance Date K003014 10/12/00 [_ Siemens Emotion Duo K991716… | 2000 |
| * | K991716 | Hispeed NX/I CT Scanner SystemSummary page 2…learance Date K003014 10/12/00 [_ Siemens Emotion Duo K991716 06/18/99 General Electric s HiSpeed NX/i… | 1999 |
What descends from it? (30)
See all 301 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K140912 | Somatom Scope/Somatom Scope Power | 2014 | 29 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 15, 2018 | Class II | Z-1451-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Mar 21, 2014 | Class II | Z-1721-2014 | Component design/selection | The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 5.2 years on median.
- 3 of the 5 ancestors have a recall of any class.
- Of its descendants, K140912 (Siemens Medical Solutions USA, Inc., 2014) has gone furthest, with 29 descendants of its own.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
The SOMATOM Spirit system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes* taken at different angles.
(*Spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K042328 · printed Nov 5, 2026
Cite this page
510(k) Report. “K042328: Somatom Spirit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042328
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026