510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Philips IU22 Ultrasound System

K042540510(k) TraditionalCleared Oct 4, 200414-day reviewClass II21 CFR 892.1560Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Ultrasound, Inc. · IYO System, Imaging, Pulsed Echo, Ultrasonic

Cleared on Oct 4, 2004 as substantially equivalent to K040227 (Philips Ultrasound, 2004) and 2 other predicates after a 14-day review. Its family tree reaches back 3 generations to 1995. 6 of the 890 devices in its family, counting itself, have a Class I recall. 882 later devices descend from it. It has been recalled 4 times.

Decided
Oct 4, 2004
Review
14 days
Class
Class II
Product code
IYO
Recalls
4
Predicates
3

Its longest line of descent runs back 3 generations to K934139 (1995). 7 ancestors in all. 882 later devices descend from it, 12 citing it directly. 6 of the 890 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2002: 1 other ancestor2000: 2 other ancestors2006: 5 descendants cleared2007: 4 descendants cleared2013: 36 descendants cleared2010: 17 descendants cleared2012: 23 descendants cleared2008: 13 descendants cleared2005: 2 descendants cleared2009: 9 descendants cleared2018: 64 descendants cleared2015: 49 descendants cleared2011: 21 descendants cleared2014: 38 descendants cleared2017: 73 descendants cleared, 2 with a Class I recall22016: 57 descendants cleared, 1 with a Class I recall12020: 73 descendants cleared, 1 with a Class I recall12019: 48 descendants cleared2026: 54 descendants cleared2021: 56 descendants cleared2022: 58 descendants cleared2023: 51 descendants cleared2024: 57 descendants cleared2025: 74 descendants cleared, 2 with a Class I recall2K934139 · Tomtec Echo-View · Tomtec Tomographic Technologies GmbH · 1995K993398 · Echo View, Easy View, Omni View · Tomtec Imaging Systems GmbH · 1999K022824 · Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X & Surgical View 1.X · Tomtec Imaging Systems GmbH · 20021995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors 1 generation forward: 12 descendants2 generations forward: 50 descendants3 generations forward: 102 descendants4 generations forward: 125 descendants5 generations forward: 146 descendants6 generations forward: 128 descendants, 2 with a Class I recall27 generations forward: 112 descendants, 2 with a Class I recall28 generations forward: 77 descendants, 1 with a Class I recall19 generations forward: 52 descendants, 1 with a Class I recall110 generations forward: 37 descendants11 generations forward: 29 descendants12 generations forward: 9 descendants13 generations forward: 2 descendants14 generations forward: 1 descendant K022824 · Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X & Surgical View 1.X · Tomtec Imaging Systems GmbH · 2002K993398 · Echo View, Easy View, Omni View · Tomtec Imaging Systems GmbH · 1999K934139 · Tomtec Echo-View · Tomtec Tomographic Technologies GmbH · 19950+5+10
  1. Root · 3 generations backK934139 Tomtec Echo-ViewTomtec Tomographic Technologies GmbH · 1995
  2. 2 generations backK993398 Echo View, Easy View, Omni ViewTomtec Imaging Systems GmbH · 1999
  3. 1 generation backK022824 Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV…Tomtec Imaging Systems GmbH · 2002
  4. This deviceK042540 Philips IU22 Ultrasound SystemPhilips Ultrasound, Inc. · 2004
  5. Descendant · cites it directly · 588 of its ownK093563 IU22 Ultrasound System and TransducersPhilips Ultrasound, Inc. · 2010 · Recalled
  6. Descendant · 2 generations on · 527 of its ownK102041 AixplorerSupersonic Imagine · 2010
  7. Descendant · 3 generations on · 525 of its ownK112255 AixplorerSupersonic Imagine · 2012
  8. Descendant · 2 generations on · 405 of its ownK121329 AixplorerSupersonic Imagine · 2012
  9. Descendant · 3 generations on · 312 of its ownK132274 AixplorerSupersonic Imagine · 2013

Showing 5 of 882 descendants. See all descendants

Descendants cleared per year

2005: 22006: 52007: 42008: 132009: 92010: 172011: 212012: 232013: 362014: 382015: 492016: 572017: 732018: 642019: 482020: 732021: 562022: 582023: 512024: 572025: 742026: 54
2005high 74 (2025) · low 2 (2005)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 882 · See all
  • K093563IU22 Ultrasound System and TransducersPhilips Ultrasound, Inc. · 2010 · cites it directly588descendants
  • K102041AixplorerSupersonic Imagine · 2010 · 2 generations on527descendants
  • K112255AixplorerSupersonic Imagine · 2012 · 3 generations on525descendants
  • K121329AixplorerSupersonic Imagine · 2012 · 2 generations on405descendants
  • K132274AixplorerSupersonic Imagine · 2013 · 3 generations on312descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K934139Tomtec Echo-ViewTomtec Tomographic Technologies GmbH1995—
−2K993398Echo View, Easy View, Omni ViewTomtec Imaging Systems GmbH1999—
−1K022824Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X…Tomtec Imaging Systems GmbH2002—
0 (this device)K042540Philips IU22 Ultrasound SystemPhilips Ultrasound, Inc.2004—
+1K093563IU22 Ultrasound System and TransducersPhilips Ultrasound, Inc.20102
+2K102041AixplorerSupersonic Imagine2010—
+3K112255AixplorerSupersonic Imagine2012—
+2K121329AixplorerSupersonic Imagine2012—
+3K132274AixplorerSupersonic Imagine2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200210
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200520
200650
200740
2008130
200990
2010170
2011210
2012230
2013360
2014380
2015490
2016571
2017732
2018640
2019480
2020731
2021560
2022580
2023510
2024570
2025742
2026540
FIG. 1 — Lineage of K042540, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYO device

  1. #1 K243937 Accuro 3Sscore 20.5 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040227Qlab Quantification Software
Summary page 1
…lly Equivalent Devices Philips Ultrasound believes that the iU22 system and transducers are substantially equivalent to the following currently marketed devices: QLAB (K040227;) LVAnalysis (K022824;) and M2424 (K022303) 4) Device Description & Technical Comparison to Predicate Devices Th…
Philips Ultrasound2004—
*K022824Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X & Surgical…
Summary page 1
…Philips Ultrasound believes that the iU22 system and transducers are substantially equivalent to the following currently marketed devices: QLAB (K040227;) LVAnalysis (K022824;) and M2424 (K022303) 4) Device Description & Technical Comparison to Predicate Devices The iU22 system and tran…
Tomtec Imaging Systems GmbH2002—
*K022303Philips M2424 Diagnostic Ultrasound System &Philips 21315 Diagnostic Ultrasound Transducer
Summary page 1
…believes that the iU22 system and transducers are substantially equivalent to the following currently marketed devices: QLAB (K040227;) LVAnalysis (K022824;) and M2424 (K022303) 4) Device Description & Technical Comparison to Predicate Devices The iU22 system and transducer(s) function in…
Philips Ultrasound, Inc.2002—

Also cleared against K040227: 4 other devices →

What descends from it? (882)

See all 882

12 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093563IU22 Ultrasound System and TransducersPhilips Ultrasound, Inc.2010588
K132304Epiq Diagnostic Ultrasound SystemPhilips Ultrasound, Inc.2013211
K060949DP-6600 Digital Ultrasonic Diagnostic Imaging SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006191
K063500DC-6 Diagnostic Ultrasound SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006189
K072164DC-6 Diagnostic Ultrasound SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.2007188
K083001M5 Diagnostic Ultrasound SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.2008179
K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.2006154
K082326Zonare Ultra Ultrasound SystemZonare Medical Systems, Inc.2008133
K130159Qlab Quantification Software Heart ModelPhilips Health Care201392
K120525Qlab Quantification SoftwarePhilips Healthcare Group201292

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 5, 2025Class IIIZ-1627-2026Device DesignTo provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Jul 3, 2025Class IIIZ-2407-2025Device DesignTo provide clarification and labelling to define the useful life of ultrasound transducers in the field.
May 28, 2025Class IIZ-2058-2025Process controlUltrasound transducer devices were refurbished beyond their useful life.
Jun 30, 2023Class IIZ-2517-2023Process controlTransducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K242670Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 6 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K242519Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 6 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 7 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 8 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 9 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 2.9 years on median.
  • None of the 7 ancestors has a recall.
  • Of its descendants, K093563 (Philips Ultrasound, Inc., 2010) has gone furthest, with 588 descendants of its own.
  • 4 other devices also cite its predicate K040227.

How does its category compare?

All IYO devices

IYO System, Imaging, Pulsed Echo, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYO

7in 2025

2007: 202008: 142009: 132010: 152011: 112012: 152013: 172014: 142015: 192016: 102017: 192018: 142019: 132020: 112021: 152022: 52023: 112024: 102025: 72026 (partial year): 10
2007high 20 (2007) · low 5 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYO

171in 2025

2007: 17 days median2008: 36 days median2009: 79 days median2010: 113 days median2011: 69 days median2012: 68 days median2013: 50 days median2014: 63 days median2015: 46 days median2016: 36 days median2017: 72 days median2018: 56 days median2019: 59 days median2020: 57 days median2021: 64 days median2022: 190 days median2023: 177 days median2024: 220 days median2025: 171 days median2026 (partial year): 139 days median
2007high 220 (2024) · low 17 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYO

1in 2022

2007: 42008: 52009: 92010: 82012: 22013: 32014: 22015: 12018: 12019: 12021: 112022: 12026 (partial year): 1
2007high 11 (2021) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYO

485in 2025

2007: 392008: 202009: 272010: 202011: 582012: 782013: 4362014: 3042015: 2682016: 2542017: 2382018: 3682019: 5372020: 4982021: 6172022: 5112023: 5272024: 6402025: 485
2007high 640 (2024) · low 20 (2008)2025

MAUDE reports naming product code IYO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072017439
20081436520
20091379927
201015113820
20111169058
20121568278
201317503436
201414632304
201519461268
201610360254
201719720238
201814561368
201913591537
202011570498
2021156411617
202251901511
2023111770527
2024102200640
202571710485
2026101391—

What is it indicated for?

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042540: Philips IU22 Ultrasound System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K042540

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026