Allura Xper FD series and Allura Xper OR Table series

K162859510(k) TraditionalCleared Dec 2, 201651-day reviewClass II21 CFR 892.1650Recalls 36 (12 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Dec 2, 2016 as substantially equivalent to K161563 (Philips Medical Systems Nederland B.V., 2016) after a 51-day review. Its family tree reaches back 13 generations to 1980. 2 of the 56 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 36 times, including 12 Class I recalls.

Decided
Dec 2, 2016
Review
51 days
Class
Class II
Product code
OWB
Recalls
36 (12 Class I)
Predicates
1

Its longest line of descent runs back 13 generations to K801623 (1980). 54 ancestors in all. 1 later device descends from it, 1 citing it directly. 2 of the 56 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2003: 2 other ancestors, 1 with a Class I recall12002: 5 other ancestors2000: 4 other ancestors2001: 1 other ancestor1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 2 other ancestors1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2022: 1 descendant clearedReference device: K130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalledK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalled×2: 2 generations cleared close together, K133292 (2014), K141979 (2014) K161563 · Allura Xper FD series and Allura Xper OR Table series · Philips Medical Systems Nederland B.V. · 2016 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 5 generations back: 4 other ancestors, 1 with a Class I recall16 generations back: 3 other ancestors7 generations back: 5 other ancestors8 generations back: 4 other ancestors9 generations back: 12 other ancestors10 generations back: 9 other ancestors11 generations back: 4 other ancestors 1 generation forward: 1 descendant K161563 · Allura Xper FD series and Allura Xper OR Table series · Philips Medical Systems Nederland B.V. · 2016 · recalledK141979 · Allura Xper FD Series / Allura Xper or Table Series · Philips Medical Systems Netherland BV · 2014 · recalledK133292 · Allura Xper FD; or Table · Philips Medical Systems Nederland B.V. · 2014 · recalledK130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−10−50
  1. Root · 13 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 9 generations between
  3. 3 generations backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  4. 2 generations backK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
  5. 1 generation backK161563 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · Recalled
  6. This deviceK162859 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016
  7. Descendant · cites it directly · 0 of its ownK220311 Cardiovascular Angiography SystemAllengers Medical Systems Limited · 2022
Show all 13 generations
  1. Root · 13 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 12 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 11 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 10 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 9 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 8 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 7 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 6 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 5 generations backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  10. 4 generations backK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
  11. 3 generations backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  12. 2 generations backK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
  13. 1 generation backK161563 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · Recalled
  14. This deviceK162859 Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K220311Cardiovascular Angiography SystemAllengers Medical Systems Limited · 2022 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−12K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−11K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−10K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−9K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−8K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−7K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−6K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−5K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
−4K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201339
−3K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V.201445
−2K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV201481
−1K161563Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V.201645
0 (this device)K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V.2016—
+1K220311Cardiovascular Angiography SystemAllengers Medical Systems Limited2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199920
200040
200110
200250
200321
200430

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
FIG. 1 — Lineage of K162859, 1980–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.2 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161563Allura Xper FD series and Allura Xper OR Table series
Summary page 4
…series Manufacturer: Philips Medical Systems Nederland B.V. 510(k) Clearance: K161563 (July 29, 2016) Classification Name: Image-intensified fluoroscopic x-ray system…
Philips Medical Systems Nederland B.V.201645
†K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201339

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K220311Cardiovascular Angiography SystemAllengers Medical Systems Limited20220

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 29, 2026Class IIZ-2928-2026Software designFirmware issue potentially leading to loss of geometry movements.
Aug 1, 2025Class IIZ-2503-2025Under Investigation by firmPhilips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Jun 30, 2025Class IIZ-2189-2025Software designPotential for temporary loss of imaging (X-ray) functionality due to software issue.
Jan 13, 2025Class IZ-1152-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1081-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IZ-1147-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1086-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1078-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1082-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1076-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1080-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1084-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1083-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IZ-1148-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1079-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IZ-1161-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1146-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1077-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1085-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IZ-1143-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1142-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1149-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1151-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1150-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1145-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1144-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
May 16, 2024Class IIZ-1896-2024Under Investigation by firmWhen Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
May 16, 2024Class IIZ-1895-2024Under Investigation by firmWhen Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
May 16, 2024Class IIZ-1897-2024Under Investigation by firmWhen Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
Jan 2, 2024Class IIZ-1156-2024Process designDual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Jan 2, 2024Class IIZ-1144-2024Process designFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may…
Jan 2, 2024Class IIZ-1151-2024Process designDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Nov 3, 2023Class IIZ-0558-2024Device DesignCeiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility…
Aug 15, 2023Class IIZ-2587-2023Device DesignA damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.
Jul 19, 2023Class IIZ-2402-2023Radiation Control for Health and Safety ActA foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
Jun 12, 2023Class IIZ-2283-2023Component design/selectionThe bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K033737Allura Xper FD20Philips Medical Systems North America, Inc. · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 113 Class I recalls

What does this lineage say?

  • The line of descent spans 36 years and passes through 8 companies; a new generation was cleared every 1.5 years on median.
  • 20 of the 54 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K220311 (Allengers Medical Systems Limited, 2022) has gone furthest, with 0 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform: • Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. • Cardiac imaging…

Read the full statement (168 words)

The Allura Xper series and the Allura Xper OR Table series (within the limits of the used OR table) are intended for use on human patients to perform:

• Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis.

• Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implantations, and electrophysiology (EP).

• Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.

Additionally:

• The Allura Xper and Allura Xper OR Table series is compatible with a hybrid Operating Room.

• Allura Xper FD10 is compatible with specified magnetic navigation systems.

• Combined with a qualified, compatible OR table, the Allura Xper OR Table series can be used for imaging in the Hybrid OR within the applications domains Neuro, Vascular, Non Vascular and Cardiac. The OR table can also be used standalone for surgical use in the OR.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K162859: Allura Xper FD series and Allura Xper OR Table series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162859

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026