Philips Medical Systems Nederland B.V. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Nov 22, 2017 as substantially equivalent to K163715 (Philips Medical Systems Nederland B.V., 2017) after a 65-day review. Its family tree reaches back 14 generations to 1980. 4 of the 62 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 53 times, including 9 Class I recalls.

Decided
Nov 22, 2017
Review
65 days
Class
Class II
Product code
OWB
Recalls
53 (9 Class I)
Predicates
1

Its longest line of descent runs back 14 generations to K801623 (1980). 55 ancestors in all. 6 later devices descend from it, 2 citing it directly. 4 of the 62 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2003: 2 other ancestors, 1 with a Class I recall12002: 5 other ancestors2000: 4 other ancestors2001: 1 other ancestor1995: 10 other ancestors1993: 2 other ancestors1989: 1 other ancestor1999: 2 other ancestors1997: 2 other ancestors1996: 3 other ancestors1994: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2018: 1 descendant cleared2020: 1 descendant cleared, 1 with a Class I recall12025: 2 descendants cleared2026: 2 descendants clearedReference device: K141979 · Allura Xper FD Series / Allura Xper or Table Series · Philips Medical Systems Netherland BV · 2014 · recalledK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalled×2: 2 generations cleared close together, K133292 (2014), K141979 (2014) ×2: 2 generations cleared close together, K162148 (2016), K163715 (2017) 19801990200020102020
Other ancestors, by generationDescendants, by generation 6 generations back: 4 other ancestors, 1 with a Class I recall17 generations back: 3 other ancestors8 generations back: 5 other ancestors9 generations back: 4 other ancestors10 generations back: 12 other ancestors11 generations back: 9 other ancestors12 generations back: 4 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 2 descendants3 generations forward: 2 descendants K163715 · Allura Xper R9 · Philips Medical Systems Nederland B.V. · 2017 · Class I recallK162148 · Allura Xper R9 · Philips Medical Systems Nethrlands BV · 2016 · recalledK141979 · Allura Xper FD Series / Allura Xper or Table Series · Philips Medical Systems Netherland BV · 2014 · recalledK133292 · Allura Xper FD; or Table · Philips Medical Systems Nederland B.V. · 2014 · recalledK130638 · Alluraclarity Xper FD Series X-Ray System · Philips Medical Systems Nederland B.V. · 2013 · recalledK102005 · Allura Xper or Table Series · Philips Medical Systems North America, Inc. · 2010 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−10−50
  1. Root · 14 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 10 generations between
  3. 3 generations backK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
  4. 2 generations backK162148 Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · Recalled
  5. 1 generation backK163715 Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · Recalled (Class I)
  6. This deviceK172822 Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017
  7. Descendant · cites it directly · 5 of its ownK181830 Azurion R2.0Philips Medical Systems Nederland B.V. · 2018 · Recalled
  8. Descendant · cites it directly · 4 of its ownK200917 Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · Recalled (Class I)
  9. Descendant · 2 generations on · 1 of its ownK251827 Azurion R3.1Philips Medical Systems B.V. · 2025
  10. Descendant · 2 generations on · 1 of its ownK243376 uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd. · 2025
  11. Descendant · 3 generations on · 0 of its ownK253726 uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd. · 2026
Show all 14 generations
  1. Root · 14 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 13 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 12 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 11 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 10 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 9 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 8 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 7 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 6 generations backK102005 Allura Xper or Table SeriesPhilips Medical Systems North America, Inc. · 2010 · Recalled
  10. 5 generations backK130638 Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V. · 2013 · Recalled
  11. 4 generations backK133292 Allura Xper FD; or TablePhilips Medical Systems Nederland B.V. · 2014 · Recalled
  12. 3 generations backK141979 Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV · 2014 · Recalled
  13. 2 generations backK162148 Allura Xper R9Philips Medical Systems Nethrlands BV · 2016 · Recalled
  14. 1 generation backK163715 Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · Recalled (Class I)
  15. This deviceK172822 Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2018: 12020: 12025: 22026: 2
2018high 2 (2025) · low 0 (2019)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K181830Azurion R2.0Philips Medical Systems Nederland B.V. · 2018 · cites it directly5descendants
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · cites it directly4descendants
  • K251827Azurion R3.1Philips Medical Systems B.V. · 2025 · 2 generations on1descendant
  • K243376uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd. · 2025 · 2 generations on1descendant
  • K253726uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−14K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−13K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−12K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−11K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−10K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−9K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−8K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−7K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−6K102005Allura Xper or Table SeriesPhilips Medical Systems North America, Inc.201039
−5K130638Alluraclarity Xper FD Series X-Ray SystemPhilips Medical Systems Nederland B.V.201339
−4K133292Allura Xper FD; or TablePhilips Medical Systems Nederland B.V.201445
−3K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV201481
−2K162148Allura Xper R9Philips Medical Systems Nethrlands BV20163
−1K163715Allura Xper R9Philips Medical Systems Nederland B.V.2017Class I
0 (this device)K172822Azurion series R1.2Philips Medical Systems Nederland B.V.2017—
+1K181830Azurion R2.0Philips Medical Systems Nederland B.V.201842
+1K200917Azurion R2.1Philips Medical Systems Nederland B.V.2020Class I
+2K251827Azurion R3.1Philips Medical Systems B.V.2025—
+2K243376uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd.2025—
+3K253726uAngio AVIVA CXShanghai United Imaging Healthcare Co., Ltd.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
198910
199130
199210
199320
199410
1995100
199630
199720
199920
200040
200110
200250
200321
200430

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202011
202520
202620
FIG. 1 — Lineage of K172822, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K163715Allura Xper R9
Summary page 4
…Allura Xper R9 Manufacturer: Philips Medical Systems Nederland B.V. 510(k) Clearance: K163715 (January 26, 2016) Classification Name: Image-intensified fluoroscopic x-ray system…
Philips Medical Systems Nederland B.V.2017Class I
†K141979Allura Xper FD Series / Allura Xper or Table SeriesPhilips Medical Systems Netherland BV201481

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181830Azurion R2.0Philips Medical Systems Nederland B.V.20185
K200917Azurion R2.1 Class IPhilips Medical Systems Nederland B.V.20204

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 6, 2026Not classifiedZ-0292-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0299-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0297-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0293-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0294-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0295-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0298-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0300-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Oct 6, 2026Not classifiedZ-0296-2027Equipment maintenanceSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of…
Jul 16, 2026Class IIZ-2931-2026Software designTwo software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements…
Jun 17, 2026Class IIZ-2892-2026Under Investigation by firmPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
Apr 28, 2026Class IIZ-2225-2026Device DesignPotential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
Mar 31, 2026Class IIZ-2170-2026Device DesignPotential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
Jan 26, 2026Class IIZ-1346-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1345-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1347-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1350-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1348-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1349-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1344-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Jan 26, 2026Class IIZ-1351-2026Device DesignUnder certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Dec 12, 2025Class IIZ-1091-2026Software designSix (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment…
Aug 1, 2025Class IIZ-2502-2025Under Investigation by firmPhilips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
Jun 4, 2025Class IIZ-2134-2025Device DesignPotential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
May 28, 2025Class IIZ-1969-2025Software designA software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations…
Jan 13, 2025Class IZ-1160-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1093-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IZ-1159-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IIZ-1092-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IIZ-1094-2025Device DesignPotential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the…
Jan 13, 2025Class IZ-1154-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1157-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1142-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1153-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1156-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1158-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 13, 2025Class IZ-1144-2025Under Investigation by firmPossibility of the patient falling from the table related to the mattress used on the patient table.
Jan 2, 2024Class IIZ-1145-2024Process designFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may…
Jan 2, 2024Class IIZ-1152-2024Process designDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Dec 7, 2023Class IIZ-2286-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2291-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2284-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2290-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2285-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2289-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2288-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Dec 7, 2023Class IIZ-2287-2026Software designDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray…
Jul 19, 2023Class IIZ-2403-2023Radiation Control for Health and Safety ActA foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation
Feb 2, 2022Class IIZ-0874-2022Under Investigation by firmDue to a software defect, the Philips StentBoost Live R2.0 might not process X-ray images of the current run and instead show an image processed in a previous run. The image shown could be from the same patient or from a different patient.
May 26, 2021Class IIZ-0886-2022Software designIn the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · Direct predicate · ◆ on the line of descent in FIG. 17 Class I recalls
  • K033737Allura Xper FD20Philips Medical Systems North America, Inc. · 2003 · 6 generations back · ◆ in the 2003 ancestors bar of FIG. 113 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · Cites it directly · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 37 years and passes through 9 companies; a new generation was cleared every 1.5 years on median.
  • 21 of the 55 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K181830 (Philips Medical Systems Nederland B.V., 2018) has gone furthest, with 5 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Azurion series (within the limits of the used Operation Room table) are intended for use to perform: * Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures. * Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures. Additionally: * The Azurion series can be…

Read the full statement (103 words)

The Azurion series (within the limits of the used Operation Room table) are intended for use to perform:

* Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.

* Cardiac imaging applications including diagnostics, interventional and minimally invasive surgery procedures.

Additionally:

* The Azurion series can be used in a hybrid Operation Room.

* The Azurion series contain a number of features to support a flexible and patient centric procedural workflow.

Patient Population

All human patients of all ages. Patient weight is limited to the specification of the patient table.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K172822: Azurion series R1.2.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172822

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026