510(k) · Cleared
Recalled · Class II · 2018Somatom Emotion 16
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Solution, Inc. USA · JAK System, X-Ray, Tomography, Computed
Cleared on Mar 1, 2005 as substantially equivalent to K023687 (Siemens Medical Systems, Inc., 2002) after a 22-day review. Its family tree reaches back 1 generation to 2002. 34 later devices descend from it. It has been recalled 8 times.
- Decided
- Mar 1, 2005
- Review
- 22 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 8
- Predicates
- 1
Its longest line of descent runs back 1 generation to K023687 (2002). 34 later devices descend from it, 4 citing it directly.
- Root · 1 generation backK023687 Somatom Emotion 6 CT SystemsSiemens Medical Systems, Inc. · 2002 · Recalled
- This deviceK050297 Somatom Emotion 16Siemens Medical Solution, Inc. USA · 2005
- Descendant · cites it directly · 30 of its ownK113287 Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2012 · Recalled
- Descendant · cites it directly · 29 of its ownK140912 Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc. · 2014 · Recalled
- Descendant · cites it directly · 29 of its ownK133590 Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2014 · Recalled
- Descendant · 2 generations on · 27 of its ownK142955 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge…Siemens Medical Solutions USA, Inc. · 2015 · Recalled
- Descendant · 3 generations on · 24 of its ownK161196 SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · Recalled
Showing 5 of 34 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 34 · See all- K113287Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2012 · cites it directly30descendants
- K140912Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc. · 2014 · cites it directly29descendants
- K133590Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2014 · cites it directly29descendants
- K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition…Siemens Medical Solutions USA, Inc. · 2015 · 2 generations on27descendants
- K161196SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · 3 generations on24descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K023687 | Somatom Emotion 6 CT Systems | Siemens Medical Systems, Inc. | 2002 | 6 |
| 0 (this device) | K050297 | Somatom Emotion 16 | Siemens Medical Solution, Inc. USA | 2005 | — |
| +1 | K113287 | Somatom Perspective | Siemens Medical Solutions USA, Inc. | 2012 | 2 |
| +1 | K140912 | Somatom Scope/Somatom Scope Power | Siemens Medical Solutions USA, Inc. | 2014 | 2 |
| +1 | K133590 | Somatom Perspective | Siemens Medical Solutions USA, Inc. | 2014 | 1 |
| +2 | K142955 | SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM… | Siemens Medical Solutions USA, Inc. | 2015 | 16 |
| +3 | K161196 | SOMATOM Drive | Siemens Medical Solutions USA, Inc. | 2016 | 4 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2009 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 2 | 0 |
| 2015 | 1 | 0 |
| 2016 | 4 | 0 |
| 2017 | 3 | 0 |
| 2018 | 2 | 0 |
| 2019 | 3 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 3 | 0 |
| 2024 | 3 | 0 |
| 2025 | 4 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K023687 | Somatom Emotion 6 CT SystemsSummary page 1…in commercial distribution: Predicate Device Name FDA Clearance Number FDA Clearance Date Siemens SOMATOM Emotion 6 K023687 11/22/2002 Siemens Medical Solutions, Inc. USA Page 7 Spec… | 2002 |
What descends from it? (34)
See all 344 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 15, 2018 | Class II | Z-1454-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1459-2018 | Software design | To inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1452-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2018 | Class II | Z-1453-2018 | Software design | to inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull… |
| Feb 15, 2016 | Class II | Z-1301-2016 | Device Design | Siemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio… |
| Mar 21, 2014 | Class II | Z-1721-2014 | Component design/selection | The rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury. |
| Aug 5, 2008 | Not classified | Z-2453-2008 | Nonconforming Material/Component | Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to… |
| Mar 9, 2006 | Not classified | Z-0793-06 | Other | When using the LungCare feature in the SOMATOM CT Systems some marks made on CT images may not be saved. |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
The Siemens SOMATOM Emotion 16 system is intended to produce cross-sectional images of
the body by computer reconstruction of x-ray transmission data from either the same axial plane
taken at different angles or spiral planes* taken at different angles.
(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube,
and the simultaneous translation of the patient.)
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K050297 · printed Nov 6, 2026
Cite this page
510(k) Report. “K050297: Somatom Emotion 16.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050297
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026