510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Somatom Emotion 16

K050297510(k) SpecialCleared Mar 1, 200522-day reviewClass II21 CFR 892.1750Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solution, Inc. USA · JAK System, X-Ray, Tomography, Computed

Cleared on Mar 1, 2005 as substantially equivalent to K023687 (Siemens Medical Systems, Inc., 2002) after a 22-day review. Its family tree reaches back 1 generation to 2002. 34 later devices descend from it. It has been recalled 8 times.

Decided
Mar 1, 2005
Review
22 days
Class
Class II
Product code
JAK
Recalls
8
Predicates
1

Its longest line of descent runs back 1 generation to K023687 (2002). 34 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2014: 2 descendants cleared2009: 1 descendant cleared2012: 1 descendant cleared2015: 1 descendant cleared2016: 4 descendants cleared2018: 2 descendants cleared2017: 3 descendants cleared2019: 3 descendants cleared2021: 2 descendants cleared2023: 3 descendants cleared2020: 2 descendants cleared2022: 1 descendant cleared2024: 3 descendants cleared2025: 4 descendants cleared2026: 2 descendants clearedK023687 · Somatom Emotion 6 CT Systems · Siemens Medical Systems, Inc. · 2002 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 3 descendants3 generations forward: 5 descendants4 generations forward: 3 descendants5 generations forward: 4 descendants6 generations forward: 4 descendants7 generations forward: 3 descendants8 generations forward: 4 descendants9 generations forward: 3 descendants10 generations forward: 1 descendant K023687 · Somatom Emotion 6 CT Systems · Siemens Medical Systems, Inc. · 2002 · recalled0+5+10
  1. Root · 1 generation backK023687 Somatom Emotion 6 CT SystemsSiemens Medical Systems, Inc. · 2002 · Recalled
  2. This deviceK050297 Somatom Emotion 16Siemens Medical Solution, Inc. USA · 2005
  3. Descendant · cites it directly · 30 of its ownK113287 Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2012 · Recalled
  4. Descendant · cites it directly · 29 of its ownK140912 Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc. · 2014 · Recalled
  5. Descendant · cites it directly · 29 of its ownK133590 Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2014 · Recalled
  6. Descendant · 2 generations on · 27 of its ownK142955 SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge…Siemens Medical Solutions USA, Inc. · 2015 · Recalled
  7. Descendant · 3 generations on · 24 of its ownK161196 SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · Recalled

Showing 5 of 34 descendants. See all descendants

Descendants cleared per year

2009: 12012: 12014: 22015: 12016: 42017: 32018: 22019: 32020: 22021: 22022: 12023: 32024: 32025: 42026: 2
2009high 4 (2016) · low 0 (2010)2026

Most influential descendants

Showing 5 of 34 · See all
  • K113287Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2012 · cites it directly30descendants
  • K140912Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc. · 2014 · cites it directly29descendants
  • K133590Somatom PerspectiveSiemens Medical Solutions USA, Inc. · 2014 · cites it directly29descendants
  • K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM Definition…Siemens Medical Solutions USA, Inc. · 2015 · 2 generations on27descendants
  • K161196SOMATOM DriveSiemens Medical Solutions USA, Inc. · 2016 · 3 generations on24descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K023687Somatom Emotion 6 CT SystemsSiemens Medical Systems, Inc.20026
0 (this device)K050297Somatom Emotion 16Siemens Medical Solution, Inc. USA2005—
+1K113287Somatom PerspectiveSiemens Medical Solutions USA, Inc.20122
+1K140912Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc.20142
+1K133590Somatom PerspectiveSiemens Medical Solutions USA, Inc.20141
+2K142955SOMATOM Force, SOMATOM Definition Flash, SOMATOM Definition Edge, SOMATOM…Siemens Medical Solutions USA, Inc.201516
+3K161196SOMATOM DriveSiemens Medical Solutions USA, Inc.20164

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201210
201420
201510
201640
201730
201820
201930
202020
202120
202210
202330
202430
202540
202620
FIG. 1 — Lineage of K050297, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K023687Somatom Emotion 6 CT Systems
Summary page 1
…in commercial distribution: Predicate Device Name FDA Clearance Number FDA Clearance Date Siemens SOMATOM Emotion 6 K023687 11/22/2002 Siemens Medical Solutions, Inc. USA Page 7 Spec…
Siemens Medical Systems, Inc.20026

Also cleared against K023687: 5 other devices →

What descends from it? (34)

See all 34

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K113287Somatom PerspectiveSiemens Medical Solutions USA, Inc.201230
K140912Somatom Scope/Somatom Scope PowerSiemens Medical Solutions USA, Inc.201429
K133590Somatom PerspectiveSiemens Medical Solutions USA, Inc.201429
K083852Biograph 16 True Point SystemsSiemens Medical Solutions USA, Inc.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 15, 2018Class IIZ-1454-2018Software designto inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull…
Feb 15, 2018Class IIZ-1459-2018Software designTo inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull…
Feb 15, 2018Class IIZ-1452-2018Software designto inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull…
Feb 15, 2018Class IIZ-1453-2018Software designto inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull…
Feb 15, 2016Class IIZ-1301-2016Device DesignSiemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio…
Mar 21, 2014Class IIZ-1721-2014Component design/selectionThe rubber damper may degrade and break, causing loosening of the motor mounting plates, which could lead to a series of events that could place the operator and patient at risk of injury.
Aug 5, 2008Not classifiedZ-2453-2008Nonconforming Material/ComponentPotential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to…
Mar 9, 2006Not classifiedZ-0793-06OtherWhen using the LungCare feature in the SOMATOM CT Systems some marks made on CT images may not be saved.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K113287 (Siemens Medical Solutions USA, Inc., 2012) has gone furthest, with 30 descendants of its own.
  • 5 other devices also cite its predicate K023687.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Siemens SOMATOM Emotion 16 system is intended to produce cross-sectional images of

the body by computer reconstruction of x-ray transmission data from either the same axial plane

taken at different angles or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube,

and the simultaneous translation of the patient.)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K050297: Somatom Emotion 16.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K050297

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026