Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - Administrator

K072144510(k) TraditionalCleared Mar 7, 2008218-day reviewClass II21 CFR 880.5725Recalls 7 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smiths Medical MD, Inc. · MEA Pump, Infusion, Pca

Cleared on Mar 7, 2008 after a 218-day review. The FDA summary names its predicate without a number. 6 of the 9 devices in its family, counting itself, have a Class I recall. 8 later devices descend from it. It has been recalled 7 times, including 2 Class I recalls.

Decided
Mar 7, 2008
Review
218 days
Class
Class II
Product code
MEA
Recalls
7 (2 Class I)
Predicates
0

8 later devices descend from it, 2 citing it directly. 6 of the 9 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2013: 2 descendants cleared, 2 with a Class I recall22016: 1 descendant cleared, 1 with a Class I recall12014: 1 descendant cleared, 1 with a Class I recall12024: 2 descendants cleared2017: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared20052010201520202025
  1. This deviceK072144 Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication…Smiths Medical MD, Inc. · 2008
  2. Descendant · cites it directly · 7 of its ownK111275 Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication…Smiths Medical Asd, Inc. · 2013 · Recalled (Class I)
  3. Descendant · cites it directly · 4 of its ownK130394 Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc. · 2013 · Recalled (Class I)
  4. Descendant · 2 generations on · 3 of its ownK170982 CADD®- Solis Ambulatory Infusion Pump with Wireless Communication…Smiths Medical Ads, Inc. · 2017 · Recalled (Class I)
  5. Descendant · 2 generations on · 1 of its ownK232793 Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES)…Zevex, Inc. · 2024
  6. Descendant · 2 generations on · 1 of its ownK140783 Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · Recalled (Class I)

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2013: 22014: 12016: 12017: 12023: 12024: 2
2013high 2 (2013) · low 0 (2015)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety SoftwareSmiths Medical Asd, Inc. · 2013 · cites it directly7descendants
  • K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc. · 2013 · cites it directly4descendants
  • K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion…Smiths Medical Ads, Inc. · 2017 · 2 generations on3descendants
  • K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin…Zevex, Inc. · 2024 · 2 generations on1descendant
  • K140783Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K072144Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety…Smiths Medical MD, Inc.2008—
+1K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety…Smiths Medical Asd, Inc.2013Class I
+1K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc.2013Class I
+2K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis…Smiths Medical Ads, Inc.2017Class I
+2K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software…Zevex, Inc.2024—
+2K140783Halo Ambulatory Infusion SystemZyno Medical, LLC2014Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201322
201411
201611
201711
202310
202420
FIG. 1 — Lineage of K072144, 2008–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—CADD-Prizm® PCS II Model 6101 Ambulatory Infusion System
Named without a submission number
—
*—CADD-Sentry Pro® Medication Safety
Named without a submission number
—

What descends from it? (8)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety Software Class ISmiths Medical Asd, Inc.20137
K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0 Class ISmiths Medical Asd, Inc.20134

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 10, 2025Class IZ-1667-2025Labeling designAffected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Feb 27, 2024Class IZ-2419-2024Software Manufacturing/Software DeploymentCADD-Solis Ambulatory Infusion Pumps, with software versions before v4.3, may have the following issues:1. Upstream Occlusion, 2. Stop and Power Keys Unresponsive, 3. Manual Mode Air Detector, 4. Single Bubble Air Detection, 5. Error…
May 26, 2015Class IIZ-1813-2015Nonconforming Material/ComponentSmiths Medical has become aware of an issue with an intermittent occurrence of binding of the locking assembly on some CADD¿-Solis pumps. Binding of the Cassette/ Keypad Lock can occur after latching the CADD¿-Solis Medication Cassette…
Sep 29, 2014Class IIZ-0166-2015Process controlSmiths Medical has identified an issue with a single batch (Lot Number 2752712) of CADD"-Solis Medication Safety Software Administrator 3.1 CDs. CADD"-Solis Medication Safety Software Administrator 3.1 CDs were sent with the CADD"-Solis…
May 10, 2011Not classifiedZ-0004-2012Nonconforming Material/ComponentReports of non-delivery/ significant under-delivery of fluids when the CADD Medication Cassette Reservoir is not correctly latched to the pump. EXPANDED: On 7/14/11, Smith Medical expanded this recall to include all consignees with…
Jan 13, 2011Not classifiedZ-1033-2011Nonconforming Material/ComponentReports of the Cassette Lock mechanism on certain CADD-Solis Pumps becoming stuck in the locked position, making it nearly impossible to remove the medication cassette reservoirs (or administration set). The purpose of the Cassette Lock…
Jan 27, 2010Not classifiedZ-1082-2010Nonconforming Material/ComponentPotential for the power cord's prongs to crack and fail at/or inside the plug. Other medical device manufacturers have reported incidents of sparking, charring, and fires from the affected power cords used with their devices. Smiths…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety SoftwareSmiths Medical Asd, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 15 Class I recalls
  • K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion System, EVAA…Zyno Medical, LLC · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 13 Class I recalls
  • K140783Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion Pump, Model 2110…Smiths Medical Ads, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 15 Class I recalls

What does this lineage say?

  • Of its descendants, K111275 (Smiths Medical Asd, Inc., 2013) has gone furthest, with 7 descendants of its own.

How does its category compare?

All MEA devices

MEA Pump, Infusion, Pca: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · MEA

2in 2017

2008: 12009: 12013: 12016: 12017: 2
2007high 2 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MEA

Too few MEA 510(k)s in any year since 2007 to give a median.

Recalls per year · MEA

1in 2024

2008: 12010: 42011: 22014: 12017: 22022: 12024: 1
2007high 4 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MEA

3,278in 2025

2007: 812008: 2372009: 1532010: 2192011: 2452012: 4462013: 4532014: 3822015: 992016: 572017: 442018: 2562019: 1322020: 2,3312021: 3,7762022: 6,4432023: 3,9842024: 4,5592025: 3,278
2007high 6,443 (2022) · low 44 (2017)2025

MAUDE reports naming product code MEA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—081
20081—1237
20091—0153
20100—4219
20110—2245
20120—0446
20131—0453
20140—1382
20150—099
20161—057
20172—244
20180—0256
20190—0132
20200—02,331
20210—03,776
20220—16,443
20230—03,984
20240—14,559
20250—03,278
20260—0—

What is it indicated for?

The CADD®-Solis ambulatory infusion pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia). The CADD®-Solis Medication…

Read the full statement (122 words)

The CADD®-Solis ambulatory infusion pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, in close proximity to nerves, into an intraoperative site (soft tissue, body cavity/surgical wound site), epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia).

The CADD®-Solis Medication Safety Software consists of the Administrator and the Point of Care software applications. This software allows you to create a set of standard pump Protocols to be used with the CADD®-Solis Ambulatory Infusion Pump and CADD-Prizm® PCS II Ambulatory Infusion Pump. The Point of Care software application is used for the CADD-Prizm® PCS II Ambulatory Infusion pump (with software revision H or higher) only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072144: Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - Administrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072144

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026