510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

Halo Ambulatory Infusion System

K140783510(k) AbbreviatedCleared Oct 17, 2014200-day reviewClass II21 CFR 880.5725Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zyno Medical, LLC · FRN Pump, Infusion

Cleared on Oct 17, 2014 as substantially equivalent to K982839 (Sims Deltec, Inc., 1998) and 3 other predicates after a 200-day review. Its family tree reaches back 3 generations to 1983. 4 of the 14 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 2 times, including 2 Class I recalls.

Decided
Oct 17, 2014
Review
200 days
Class
Class II
Product code
FRN
Recalls
2 (2 Class I)
Predicates
4

Its longest line of descent runs back 3 generations to K832284 (1983). 12 ancestors in all. 1 later device descends from it, 1 citing it directly. 4 of the 14 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor1998: 1 other ancestor2008: 2 other ancestors, 1 with a Class I recall12006: 1 other ancestor2000: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor2016: 1 descendant cleared, 1 with a Class I recall1K832284 · Solution Admin. Set 2C9096 W/10 PVC · Travenol Laboratories, S.A. · 1983 · recalledK981792 · Solution Administration Set · Baxter Healthcare Corp · 1998 · recalledK120685 · Zyno Medical Administration Set · Zyno Medical, LLC · 201219801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors, 1 with a Class I recall12 generations back: 6 other ancestors, 1 with a Class I recall1 1 generation forward: 1 descendant, 1 with a Class I recall1 K120685 · Zyno Medical Administration Set · Zyno Medical, LLC · 2012K981792 · Solution Administration Set · Baxter Healthcare Corp · 1998 · recalledK832284 · Solution Admin. Set 2C9096 W/10 PVC · Travenol Laboratories, S.A. · 1983 · recalled0
  1. Root · 3 generations backK832284 Solution Admin. Set 2C9096 W/10 PVCTravenol Laboratories, S.A. · 1983 · Recalled
  2. 2 generations backK981792 Solution Administration SetBaxter Healthcare Corp · 1998 · Recalled
  3. 1 generation backK120685 Zyno Medical Administration SetZyno Medical, LLC · 2012
  4. This deviceK140783 Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014
  5. Descendant · cites it directly · 0 of its ownK153193 Halo II Ambulatory Infusion System (also with trade names Nimbus II…Zyno Medical, LLC · 2016 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion…Zyno Medical, LLC · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K832284Solution Admin. Set 2C9096 W/10 PVCTravenol Laboratories, S.A.19832
−2K981792Solution Administration SetBaxter Healthcare Corp19985
−1K120685Zyno Medical Administration SetZyno Medical, LLC2012—
0 (this device)K140783Halo Ambulatory Infusion SystemZyno Medical, LLC2014—
+1K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory…Zyno Medical, LLC2016Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199710
199810
200010
200310
200610
200821
201311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201611
FIG. 1 — Lineage of K140783, 1983–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982839Cadd-Legacy Pca Ambulatory Infusion System, Model 6300
Summary page 4
…mp, product code MEA (21 CFR 880.5725); and Administration Set, product code FPA (21 CFR 880.5440) Predicate Device: CADD-Legacy PCA Model 6300 (K982839) ambIT Continuous Ambulatory Infusion Pump (K033325) CADD-Solis VIP Ambulatory Infusion…
Sims Deltec, Inc.1998—
*K033325Ambit Continuous Ambulatory Infusion Pump
Summary page 4
…ration Set, product code FPA (21 CFR 880.5440) Predicate Device: CADD-Legacy PCA Model 6300 (K982839) ambIT Continuous Ambulatory Infusion Pump (K033325) CADD-Solis VIP Ambulatory Infusion Pump (K111275) Zyno Medical Administration Set (K12…
Sorenson Medical, Inc.20034
*K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety Software
Summary page 4
…DD-Legacy PCA Model 6300 (K982839) ambIT Continuous Ambulatory Infusion Pump (K033325) CADD-Solis VIP Ambulatory Infusion Pump (K111275) Zyno Medical Administration Set (K120685) I. INTENDED USE The Halo Ambulatory Infusion System is in…
Smiths Medical Asd, Inc.2013Class I
*K120685Zyno Medical Administration Set
Summary page 4
…ontinuous Ambulatory Infusion Pump (K033325) CADD-Solis VIP Ambulatory Infusion Pump (K111275) Zyno Medical Administration Set (K120685) I. INTENDED USE The Halo Ambulatory Infusion System is intended to deliver medications and/or fluids to a patient under…
Zyno Medical, LLC2012—

Also cleared against K982839: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion… Class IZyno Medical, LLC20160

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 21, 2024Class IZ-1285-2024Device DesignInfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump…
Feb 21, 2024Class IZ-1286-2024Device DesignInfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety SoftwareSmiths Medical Asd, Inc. · 2013 · Direct predicate · ◆ in the 2013 ancestors bar of FIG. 15 Class I recalls
  • K072144Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - AdministratorSmiths Medical MD, Inc. · 2008 · 2 generations back · ◆ in the 2008 ancestors bar of FIG. 12 Class I recalls
  • K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion System, EVAA…Zyno Medical, LLC · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 31 years and passes through 3 companies; a new generation was cleared every 14.1 years on median.
  • 7 of the 12 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K153193 (Zyno Medical, LLC, 2016) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K982839.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Halo Ambulatory Infusion System is intended to deliver medications and/or fluids to a patient under the direction or supervision

of physician or other certified healthcare professional. The device is intended for subcutaneous, perineural, epidural and intravenous

infusion. The device is intended for 240 hours or1500mL of infusion, whichever limit is reached first.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140783: Halo Ambulatory Infusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140783

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026