Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety Software

K111275510(k) TraditionalCleared Feb 1, 2013637-day reviewClass II21 CFR 880.5725Recalls 6 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smiths Medical Asd, Inc. · FRN Pump, Infusion

Cleared on Feb 1, 2013 as substantially equivalent to K072144 (Smiths Medical MD, Inc., 2008) and 2 other predicates after a 637-day review. Its family tree reaches back 1 generation to 1996. 6 of the 11 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled 6 times, including 5 Class I recalls.

Decided
Feb 1, 2013
Review
637 days
Class
Class II
Product code
FRN
Recalls
6 (5 Class I)
Predicates
3

Its longest line of descent runs back 1 generation to K960826 (1996). 3 ancestors in all. 7 later devices descend from it, 4 citing it directly. 6 of the 11 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 2 other ancestors, 1 with a Class I recall12016: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared, 1 with a Class I recall12014: 1 descendant cleared, 1 with a Class I recall12024: 2 descendants cleared2017: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant clearedK960826 · Cadd-Prizm Model 6100 Pump · Sims Deltec, Inc. · 1996 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 4 descendants, 3 with a Class I recall32 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 1 descendant K960826 · Cadd-Prizm Model 6100 Pump · Sims Deltec, Inc. · 1996 · recalled0
  1. Root · 1 generation backK960826 Cadd-Prizm Model 6100 PumpSims Deltec, Inc. · 1996 · Recalled
  2. This deviceK111275 Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication…Smiths Medical Asd, Inc. · 2013
  3. Descendant · cites it directly · 4 of its ownK130394 Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc. · 2013 · Recalled (Class I)
  4. Descendant · 2 generations on · 3 of its ownK170982 CADD®- Solis Ambulatory Infusion Pump with Wireless Communication…Smiths Medical Ads, Inc. · 2017 · Recalled (Class I)
  5. Descendant · cites it directly · 1 of its ownK232793 Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES)…Zevex, Inc. · 2024
  6. Descendant · cites it directly · 1 of its ownK140783 Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · Recalled (Class I)
  7. Descendant · 2 generations on · 0 of its ownK242660 Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES)…Zevex, Inc. · 2024

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2013: 12014: 12016: 12017: 12023: 12024: 2
2013high 2 (2024) · low 0 (2015)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 7 · See all
  • K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc. · 2013 · cites it directly4descendants
  • K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion…Smiths Medical Ads, Inc. · 2017 · 2 generations on3descendants
  • K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin…Zevex, Inc. · 2024 · cites it directly1descendant
  • K140783Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · cites it directly1descendant
  • K242660Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin…Zevex, Inc. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K960826Cadd-Prizm Model 6100 PumpSims Deltec, Inc.19961
0 (this device)K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety…Smiths Medical Asd, Inc.2013—
+1K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc.2013Class I
+2K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis…Smiths Medical Ads, Inc.2017Class I
+1K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software…Zevex, Inc.2024—
+1K140783Halo Ambulatory Infusion SystemZyno Medical, LLC2014Class I
+2K242660Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software…Zevex, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200821

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201311
201411
201611
201711
202310
202420
FIG. 1 — Lineage of K111275, 1996–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072144Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software -…
Summary page 1
…ystem Proprietary Name: CADD®-Solis VIP Ambulatory Infusion Pump CADDTM-Solis Medication Safety Software Predicate Devices: K072 144, CADDO-Solis Ambulatory Infusion Pump K960826, CADD-Prizm® VIP Ambulatory Infusion Pump…
Smiths Medical MD, Inc.2008Class I
*K960826Cadd-Prizm Model 6100 Pump
Summary page 1
…CADDTM-Solis Medication Safety Software Predicate Devices: K072 144, CADDO-Solis Ambulatory Infusion Pump K960826, CADD-Prizm® VIP Ambulatory Infusion Pump K082783, CADD TM -Solis Medication Safety Software-Point of…
Sims Deltec, Inc.19961
*K082783Cadd -Solis Medication Safety Software: Administrator and Point-of-Care
Summary page 1
…dicate Devices: K072 144, CADDO-Solis Ambulatory Infusion Pump K960826, CADD-Prizm® VIP Ambulatory Infusion Pump K082783, CADD TM -Solis Medication Safety Software-Point of Care K072 144, CADD TM -Solis Medication Safety So…
Smiths Medical MD, Inc.2008—

Also cleared against K072144: 1 other device →

What descends from it? (7)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0 Class ISmiths Medical Asd, Inc.20134
K232793Curlin 8000 Ambulatory Infusion System; Enterprise Solution (ES) Software; Curlin…Zevex, Inc.20241
K140783Halo Ambulatory Infusion System Class IZyno Medical, LLC20141
K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion… Class IZyno Medical, LLC20160

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 10, 2025Class IZ-1682-2025Under Investigation by firmThere is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.
Apr 10, 2025Class IZ-1668-2025Labeling designAffected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Jul 16, 2024Class IZ-0222-2025Component design/selectionDamage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this…
Jun 26, 2024Class IIZ-0466-2025Device DesignThe adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.
Feb 27, 2024Class IZ-2421-2024Software Manufacturing/Software DeploymentThe CADD Solis VIP Ambulatory Infusion Pump is indicated for the following uses: " For intravenous, intraarterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.
Mar 24, 2020Class IZ-0107-2022Component change controlSpecific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K072144Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - AdministratorSmiths Medical MD, Inc. · 2008 · Direct predicate · ◆ in the 2008 ancestors bar of FIG. 12 Class I recalls
  • K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion System, EVAA…Zyno Medical, LLC · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 13 Class I recalls
  • K130394Cadd -Solis Ambulatory Infusion Pump, Version 3.0Smiths Medical Asd, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K140783Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K170982CADD®- Solis Ambulatory Infusion Pump with Wireless Communication; CADD®-Solis Infusion Pump, Model 2110…Smiths Medical Ads, Inc. · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 15 Class I recalls

What does this lineage say?

  • 2 of the 3 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K130394 (Smiths Medical Asd, Inc., 2013) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K072144.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The CADD-Solis VIP Ambulatory Infusion Pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion. PCA (patient-controlled analgesia) delivery is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patient-controlled analgesia. Continuous delivery allows the infusion of drug/fluid at a constant…

Read the full statement (213 words)

The CADD-Solis VIP Ambulatory Infusion Pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, perineural, surgical site, epidural space, or subarachnoid space infusion.

PCA (patient-controlled analgesia) delivery is used for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both, such as patient-controlled analgesia.

Continuous delivery allows the infusion of drug/fluid at a constant, programmed rate.

Intermittent delivery allows the infusion of a specific volume of drug/fluid at a regular, programmed interval.

Step delivery allows an incremental increase in infusion rate to a specified maximum infusion rate for a specified total infusion volume.

Taper delivery allows a plateau rate of infusion with the option of tapering at the beginning and/or end and has a programmable KVO rate at the end of the infusion.

The CADD-Solis Medication Safety Software - Administrator allows use of a computer to create therapy-based protocol libraries to be used with the CADD-Solis VIP Ambulatory Infusion Pump, CADD-Solis Ambulatory Infusion Pump or CADD-Prizm PCS II Ambulatory Infusion Pump (software revision H or higher).

The CADD-Solis Medication Safety Software - Point of Care allows use of a computer to send therapy-based protocols developed by the CADD-Solis Medication Safety Software - Administrator to the CADD-Solis Ambulatory Infusion Pump and CADD-Prizm PCS II Ambulatory Infusion Pump (software revision H or higher).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K111275: Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111275

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026