510(k) Report

510(k) · Cleared

Recalled · Class I · 2024

Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion System, EVAA Ambulatory Infusion System)

K153193510(k) AbbreviatedCleared Aug 30, 2016301-day reviewClass II21 CFR 880.5725Recalls 5 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zyno Medical, LLC · FRN Pump, Infusion

Cleared on Aug 30, 2016 as substantially equivalent to K140783 (Zyno Medical, LLC, 2014) and 2 other predicates after a 301-day review. Its family tree reaches back 4 generations to 1983. 4 of the 14 devices in its family, counting itself, have a Class I recall. It has been recalled 5 times, including 3 Class I recalls.

Decided
Aug 30, 2016
Review
301 days
Class
Class II
Product code
FRN
Recalls
5 (3 Class I)
Predicates
3

Its longest line of descent runs back 4 generations to K832284 (1983). 13 ancestors in all. No later device cites it as a predicate yet. 4 of the 14 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2013: 1 other ancestor, 1 with a Class I recall12008: 2 other ancestors, 1 with a Class I recall12006: 1 other ancestor2003: 1 other ancestor1998: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor2000: 1 other ancestorK832284 · Solution Admin. Set 2C9096 W/10 PVC · Travenol Laboratories, S.A. · 1983 · recalledK981792 · Solution Administration Set · Baxter Healthcare Corp · 1998 · recalledK120685 · Zyno Medical Administration Set · Zyno Medical, LLC · 2012K140783 · Halo Ambulatory Infusion System · Zyno Medical, LLC · 2014 · Class I recall1980199020002010
Other ancestors, by generation 1 generation back: 1 other ancestor, 1 with a Class I recall12 generations back: 7 other ancestors, 1 with a Class I recall13 generations back: 1 other ancestor K140783 · Halo Ambulatory Infusion System · Zyno Medical, LLC · 2014 · Class I recallK120685 · Zyno Medical Administration Set · Zyno Medical, LLC · 2012K981792 · Solution Administration Set · Baxter Healthcare Corp · 1998 · recalledK832284 · Solution Admin. Set 2C9096 W/10 PVC · Travenol Laboratories, S.A. · 1983 · recalled0
  1. Root · 4 generations backK832284 Solution Admin. Set 2C9096 W/10 PVCTravenol Laboratories, S.A. · 1983 · Recalled
  2. 3 generations backK981792 Solution Administration SetBaxter Healthcare Corp · 1998 · Recalled
  3. 2 generations backK120685 Zyno Medical Administration SetZyno Medical, LLC · 2012
  4. 1 generation backK140783 Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · Recalled (Class I)
  5. This deviceK153193 Halo II Ambulatory Infusion System (also with trade names Nimbus II…Zyno Medical, LLC · 2016

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K832284Solution Admin. Set 2C9096 W/10 PVCTravenol Laboratories, S.A.19832
−3K981792Solution Administration SetBaxter Healthcare Corp19985
−2K120685Zyno Medical Administration SetZyno Medical, LLC2012—
−1K140783Halo Ambulatory Infusion SystemZyno Medical, LLC2014Class I
0 (this device)K153193Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory…Zyno Medical, LLC2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199710
199810
200010
200310
200610
200821
201311
FIG. 1 — Lineage of K153193, 1983–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140783Halo Ambulatory Infusion System
Summary page 4
…II Panel: 80 General Hospital Product codes: FRN, MEA, FPA Predicate Device: Halo Ambulatory Infusion System (K140783) CADD-Solis VIP Ambulatory Infusion Pump (K111275) Zyno Medical Administration Set (K…
Zyno Medical, LLC2014Class I
*K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety Software
Summary page 4
…ital Product codes: FRN, MEA, FPA Predicate Device: Halo Ambulatory Infusion System (K140783) CADD-Solis VIP Ambulatory Infusion Pump (K111275) Zyno Medical Administration Set (K120685) I. INDICATIONS FOR USE The Halo II ambulatory infusion s…
Smiths Medical Asd, Inc.2013Class I
*K120685Zyno Medical Administration Set
Summary page 4
…Halo Ambulatory Infusion System (K140783) CADD-Solis VIP Ambulatory Infusion Pump (K111275) Zyno Medical Administration Set (K120685) I. INDICATIONS FOR USE The Halo II ambulatory infusion system is intended to deliver medications and/or fluids to a pati…
Zyno Medical, LLC2012—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 18, 2024Class IZ-1811-2024Device DesignNimbus II Plus, Infusion Pump Systems, manufactured by InfuTronix, are being removed the from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or…
Feb 21, 2024Class IZ-1285-2024Device DesignInfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump…
Feb 21, 2024Class IZ-1286-2024Device DesignInfuTronix is removing the Nimbus Infusion Pump Systems from the market due to multiple potential failure modes, including battery failure, upstream occlusion, system errors, drug product leakage, high or low flow rate, or damaged pump…
Oct 19, 2021Class IIZ-0832-2022Employee errorAn unreleased Version 4 Protocol Library was installed on Nimbus PainPro (H) infusion pumps in error that may impact patient safety if used on patients.
May 28, 2019Class IIZ-2055-2019Employee errorPotential for tube leakage

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K140783Halo Ambulatory Infusion SystemZyno Medical, LLC · 2014 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K111275Cadd Solis Vip Ambulatory Infusion Pump and Cadd Solis Medication Safety SoftwareSmiths Medical Asd, Inc. · 2013 · Direct predicate · ◆ in the 2013 ancestors bar of FIG. 15 Class I recalls
  • K072144Cadd -Solis Ambulatory Infusion Pump and Cadd -Solis Medication Safety Software - AdministratorSmiths Medical MD, Inc. · 2008 · 2 generations back · ◆ in the 2008 ancestors bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 33 years and passes through 3 companies; a new generation was cleared every 14.1 years on median.
  • 8 of the 13 ancestors have a recall of any class; 3 have a Class I recall.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Halo II Ambulatory Infusion System is intended to deliver medications and/or fluids to a patient under the direction

or supervision of physician or other certified healthcare professional in clinical or nonclinical environments, such as

homes. The device is intended for subcutaneous, percutaneous, perineural, epidural and intravenous infusion, including

but not limited to patient controlled analgesia (PCA) delivery.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K153193: Halo II Ambulatory Infusion System (also with trade names Nimbus II Ambulatory Infusion System, EVAA Ambulatory Infusion System).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K153193

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026