510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Ziehm Vision RFD 3D

K142740510(k) TraditionalCleared Apr 6, 2015194-day reviewClass II21 CFR 892.1650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ziehm Imaging GmbH · OWB Interventional Fluoroscopic X-Ray System

Cleared on Apr 6, 2015 as substantially equivalent to K132904 (Ziehm Imaging GmbH, 2013) and 1 other predicate after a 194-day review. Its family tree reaches back 6 generations to 2000. 3 of the 24 devices in its family, counting itself, have a Class I recall. 10 later devices descend from it. It has been recalled once.

Decided
Apr 6, 2015
Review
194 days
Class
Class II
Product code
OWB
Recalls
1
Predicates
2

Its longest line of descent runs back 6 generations to K003266 (2000). 13 ancestors in all. 10 later devices descend from it, 3 citing it directly. 3 of the 24 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 2 other ancestors2001: 3 other ancestors, 1 with a Class I recall12003: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor, 1 with a Class I recall12020: 2 descendants cleared2019: 1 descendant cleared2021: 1 descendant cleared2023: 1 descendant cleared2025: 2 descendants cleared2022: 1 descendant cleared2026: 2 descendants clearedK003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 2000K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K061310 · Ziehm Vario 3D Digital Mobile Imaging System · Ziehm Imaging, Inc. · 2006K073346 · Ziehm Vision 2 · Ziehm Imaging, Inc. · 2008 · recalledK083545 · Ziehm Vision RFD · Ziehm Imaging, Inc. · 2009 · recalledK132904 · Ziehm Vision RFD · Ziehm Imaging GmbH · 2013 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 1 other ancestor4 generations back: 4 other ancestors, 3 with a Class I recall3 1 generation forward: 3 descendants2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants K132904 · Ziehm Vision RFD · Ziehm Imaging GmbH · 2013 · recalledK083545 · Ziehm Vision RFD · Ziehm Imaging, Inc. · 2009 · recalledK073346 · Ziehm Vision 2 · Ziehm Imaging, Inc. · 2008 · recalledK061310 · Ziehm Vario 3D Digital Mobile Imaging System · Ziehm Imaging, Inc. · 2006K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 2000−50
  1. Root · 6 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 5 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 4 generations backK061310 Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006
  4. 3 generations backK073346 Ziehm Vision 2Ziehm Imaging, Inc. · 2008 · Recalled
  5. 2 generations backK083545 Ziehm Vision RFDZiehm Imaging, Inc. · 2009 · Recalled
  6. 1 generation backK132904 Ziehm Vision RFDZiehm Imaging GmbH · 2013 · Recalled
  7. This deviceK142740 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2015
  8. Descendant · cites it directly · 4 of its ownK202360 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2020
  9. Descendant · 2 generations on · 3 of its ownK231701 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2023 · Recalled
  10. Descendant · 3 generations on · 2 of its ownK243303 Ziehm Vision RFD 3DZiehm Imaging GmbH · 2025
  11. Descendant · cites it directly · 2 of its ownK190064 nView system 1Nview Medical, Inc. · 2019
  12. Descendant · 2 generations on · 1 of its ownK211064 nView s1 with nav optionNview Medical · 2021

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2019: 12020: 22021: 12022: 12023: 12025: 22026: 2
2019high 2 (2020) · low 0 (2024)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 10 · See all
  • K202360Ziehm Vision RFD 3DZiehm Imaging GmbH · 2020 · cites it directly4descendants
  • K231701Ziehm Vision RFD 3DZiehm Imaging GmbH · 2023 · 2 generations on3descendants
  • K243303Ziehm Vision RFD 3DZiehm Imaging GmbH · 2025 · 3 generations on2descendants
  • K190064nView system 1Nview Medical, Inc. · 2019 · cites it directly2descendants
  • K211064nView s1 with nav optionNview Medical · 2021 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−5K032280Siremobil Iso-C 3DSiemens AG2003—
−4K061310Ziehm Vario 3D Digital Mobile Imaging SystemZiehm Imaging, Inc.2006—
−3K073346Ziehm Vision 2Ziehm Imaging, Inc.20085
−2K083545Ziehm Vision RFDZiehm Imaging, Inc.20093
−1K132904Ziehm Vision RFDZiehm Imaging GmbH20131
0 (this device)K142740Ziehm Vision RFD 3DZiehm Imaging GmbH2015—
+1K202360Ziehm Vision RFD 3DZiehm Imaging GmbH2020—
+2K231701Ziehm Vision RFD 3DZiehm Imaging GmbH20231
+3K243303Ziehm Vision RFD 3DZiehm Imaging GmbH2025—
+1K190064nView system 1Nview Medical, Inc.2019—
+2K211064nView s1 with nav optionNview Medical2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200131
200211
200311
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202020
202110
202210
202310
202520
202620
FIG. 1 — Lineage of K142740, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.2 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132904Ziehm Vision RFD
Summary page 5
…Device: System, X-ray, tomography, computed Classification Product JAK Code: Predicate Device: K132904 Ziehm Vision RFD Decision Date: 12/05/2013 Classification Product…
Ziehm Imaging GmbH20131
*K073346Ziehm Vision 2
Summary page 6
…Codes: Ziehm Vision RFD 3D - 510(k) Submission VOL_004_002_510(k) Summary 2/8 Predicate Device: K073346 Ziehm Vision2 FD Vario 3D Decision Date: 03/12/2008 Classification Product…
Ziehm Imaging, Inc.20085

Also cleared against K132904: 2 other devices →

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K202360Ziehm Vision RFD 3DZiehm Imaging GmbH20204
K190064nView system 1Nview Medical, Inc.20192
K203281Mobile ImagingRing System, IRm, Loop-X Mobile Imaging Robot, Loop-XMedphoton GmbH20201

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 15, 2016Class IIZ-2639-2016Device DesignMalfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 4 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 4 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 4 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 2.8 years on median.
  • 8 of the 13 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K202360 (Ziehm Imaging GmbH, 2020) has gone furthest, with 4 descendants of its own.
  • 2 other devices also cite its predicate K132904.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher…

Read the full statement (273 words)

The Ziehm Vision RFD 3D system is intended for use in providing both 2D and 3D medical imaging for all adult and pediatric populations, using pulsed and continuous fluoroscopic imaging.

The device provides 2D medical imaging for fluoroscopy, digital subtraction, and acquisition of cine loops during diagnostic interventional and surgical procedures where intra-operative imaging and visualization of complex anatomical structures of both lower and higher contrast density are required. Such procedures may include but are not limited to those of interventional cardiology, heart surgery, hybrid procedures, interventional radiology, interventional angiography, electrophysiology, pediatrics, endoscopic, urological, gastroenterology, orthopedic, maxillofacial surgery, neurology, neurosurgery, critical care, emergency room procedures, and those procedures visualizing structures of the cervical, thoracic, and lumbar regions of the spine and joint fractures of the upper and lower extremities, and where digital image data is required for Computer-Assisted Surgery procedures.

The device is also intended to provide 3D medical imaging of patients during orthopedic, neurological, intra-operative surgical procedures and where the clinician benefits from 3D visualization of complex anatomical structures, such as but not limited to those of high contrast objects, bones, joints, maxillofacial, cervical, thoracic, and lumbar regions of the spine, pelvis, acetabulum and joint fractures of the upper and lower extremities, and where digital image and C-arm positioning data is required for Computer-Assisted Surgery procedures.

The visualization of such anatomical structures assists the clinician in the clinical outcome. At the discretion of a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography. The system is not intended for use near MRI systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142740: Ziehm Vision RFD 3D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142740

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026