GE Datex-Ohmeda Aisys

K073707510(k) SpecialCleared Jan 30, 200830-day reviewClass II21 CFR 868.5160Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datex-Ohmeda, Inc. · BSZ Gas-Machine, Anesthesia

Cleared on Jan 30, 2008 after a 30-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 11 of the 32 devices in its family, counting itself, have a Class I recall. 31 later devices descend from it. It has been recalled 2 times.

Decided
Jan 30, 2008
Review
30 days
Class
Class II
Product code
BSZ
Recalls
2
Predicates
0

31 later devices descend from it, 3 citing it directly. 11 of the 32 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2009: 2 descendants cleared2008: 1 descendant cleared2011: 4 descendants cleared, 3 with a Class I recall32010: 1 descendant cleared2012: 2 descendants cleared, 2 with a Class I recall22013: 5 descendants cleared, 1 with a Class I recall12015: 3 descendants cleared2017: 3 descendants cleared, 1 with a Class I recall12014: 2 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared2016: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2019: 1 descendant cleared2026: 1 descendant cleared20052010201520202025
  1. This deviceK073707 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2008
  2. Descendant · cites it directly · 30 of its ownK081844 GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · Recalled
  3. Descendant · cites it directly · 19 of its ownK090233 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · Recalled
  4. Descendant · 2 generations on · 13 of its ownK112722 GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · Recalled (Class I)
  5. Descendant · 3 generations on · 11 of its ownK123125 GE Datex-Ohmeda Avance CS2Datex-Ohmeda, Inc. · 2013 · Recalled (Class I)
  6. Descendant · 2 generations on · 9 of its ownK110213 GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · Recalled (Class I)

Showing 5 of 31 descendants. See all descendants

Descendants cleared per year

2008: 12009: 22010: 12011: 42012: 22013: 52014: 22015: 32016: 12017: 32018: 22019: 12020: 12021: 12023: 12026: 1
2008high 5 (2013) · low 0 (2022)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 31 · See all
  • K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2008 · cites it directly30descendants
  • K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2009 · cites it directly19descendants
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 2 generations on13descendants
  • K123125GE Datex-Ohmeda Avance CS2Datex-Ohmeda, Inc. · 2013 · 3 generations on11descendants
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 2 generations on9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K073707GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.2008—
+1K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc.20084
+1K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.20092
+2K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc.2011Class I
+3K123125GE Datex-Ohmeda Avance CS2Datex-Ohmeda, Inc.2013Class I
+2K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.2011Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200920
201010
201143
201222
201351
201421
201530
201611
201731
201820
201910
202011
202110
202311
202610
FIG. 1 — Lineage of K073707, 2008–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

11 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it? (31)

See all 31

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081844GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc.200830
K090233GE Datex-Ohmeda AisysDatex-Ohmeda, Inc.200919
K090892GE Datex-Ohmeda S/5 Adu CarestationGE Healthcare Finland Oy20090

Has anything in its family been recalled? (11 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 5, 2009Not classifiedZ-1425-2009Component design/selectionUnintended shut-down: The anesthesia machine may start shutting down without human intervention due to a faulty On/Standby Switch. Prior to shutdown, the system will issue an alarm and play a message that shutdown will occur in 8 seconds.…
Aug 28, 2008Not classifiedZ-0020-2009Manufacturing material removalThe electronic vaporization system in the GE Aisys Anesthesia machine contains the following components: Backpressure Valve, Inflow Check Valve and Cassette Interface Board, that can contribute to independent failures.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K102471AS3700 Anesthsia Delivery SystemMindray DS USA, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K112722GE Datex-Ohmeda AvanceDatex-Ohmeda, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K112114Maquet Flow-I C20, Maquet Flow-I C30, Maquet Flow-I C40Maquet Critical Care AB · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K110213GE Datex-Ohmeda AisysDatex-Ohmeda, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K122445Aespire 7900, Aespire ViewDatex-Ohmeda · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K172045Aestiva 7900, Aestiva MRI, Aestiva 7100, Aestiva 7100 Compact, Aespire 100, Aespire 7100, Aespire 7900…Datex-Ohmeda, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 110 Class I recalls
  • K133958Flow-I Anesthesia System C20, C30 & C40Maquet Critical Care AB · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K123125GE Datex-Ohmeda Avance CS2Datex-Ohmeda, Inc. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K151570Carestation 620/650/650CDatex-Ohmeda, Inc. · 2016 · 5 generations on · ◆ in the 2016 descendants bar of FIG. 19 Class I recalls
  • K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB · 2020 · 5 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls
  • K213867Carestation 750/750cDatex-Ohmeda, Inc. · 2023 · 6 generations on · ◆ in the 2023 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • Of its descendants, K081844 (Datex-Ohmeda, Inc., 2008) has gone furthest, with 30 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—
Cite this page

510(k) Report. “K073707: GE Datex-Ohmeda Aisys.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K073707

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026